DAR 444SP02012 (UDI 20884522035191)Recalls & reported problems

Covidien LP · Closed Suction System, Replaceable Catheter,Endotracheal Length,DETACH. DSE · Class 1

Quick summary

This is a Covidien closed suction catheter used for tracheal suctioning, with a detachable endotracheal-length design. The 50 reports mostly describe problems found before or during setup: air or gas leaks, torn or cracked parts, tubing that would not connect properly, and suction control valves that could not be pressed. None of the reports involved a patient, and there are no deaths or injuries. No recalls name this device. Reports in the past 12 months fell to 0, compared with 26 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 20884522035191), received between May 23, 2024 and Jun 5, 2025 — classified as 50 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (23 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 7 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 5 reportsOct 2024: 10 reportsNov 2024: 6 reportsDec 2024: 4 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 7 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem23
  • Gas/Air Leak19
  • Material Split, Cut or Torn11
  • Connection Problem8
  • Device Damaged Prior to Use2
  • Incomplete or Inadequate Connection2
  • Crack1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions50

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2025Use of Device Problem; Material Split, Cut or Torn; Use of Device Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND CONNECTED THE NEGATIVE PRESSURE SUCTION TUBE, THE SUCTION CONTROL VALVE COULD NOT BE PRESSED. THERE WAS NO PATIENT INVOLVED. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND A COMPONENT DAMAGED/TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22154349 ↗
  • MalfunctionJun 5, 2025Use of Device Problem; Material Split, Cut or Torn; Use of Device Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND CONNECTED THE NEGATIVE PRESSURE SUCTION TUBE, THE SUCTION CONTROL VALVE COULD NOT BE PRESSED. THERE WAS NO PATIENT INVOLVED. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND A COMPONENT DAMAGED/TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22154078 ↗
  • MalfunctionMay 14, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND CONNECTED THE NEGATIVE PRESSURE SUCTION LINE, THE SUCTION CONTROL VALVE COULD NOT BE PRESSED. THIS ISSUE OCCURRED AT A HOSPITAL AND HAD NOT BEEN OBSERVED DURING PREVIOUS USE. THERE WAS NO PATIENT INVOLVED. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND A COMPONENT DAMAGE/TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22021758 ↗
  • MalfunctionMay 14, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND CONNECTED THE NEGATIVE PRESSURE SUCTION LINE, THE SUCTION CONTROL VALVE COULD NOT BE PRESSED. THIS ISSUE OCCURRED AT A HOSPITAL AND HAD NOT BEEN OBSERVED DURING PREVIOUS USE. THERE WAS NO PATIENT INVOLVED. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND A COMPONENT DAMAGE/TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22021739 ↗
  • MalfunctionMay 14, 2025Use of Device Problem; Material Split, Cut or Torn; Use of Device Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND THE NEGATIVE PRESSURE SUCTION LINE WAS CONNECTED TO THE PRODUCT, THE SUCTION CONTROL VALVE COULD NOT BE PRESSED. THIS ISSUE OCCURRED AT A HOSPITAL AND HAD NOT BEEN OBSERVED DURING PREVIOUS USE. THERE WAS NO PATIENT INVOLVED. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND A COMPONENT DAMAGE/TORN AND THE SUCTION VALVE WITH THE CAP THAT COULD NOT ROTATE PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22020632 ↗
  • MalfunctionApr 25, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21914851 ↗
  • MalfunctionMar 19, 2025Connection Problem; Gas/Air Leak; Connection Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN TESTING WITH A TEST LUNG BEFORE STARTING THE VENTILATOR, THERE WAS A LEAK NOTED ON THE SIX DEVICES. THERE WAS ALSO A CRACK IN THE Y-PIECE REGION (ON THE VENTILATOR SIDE). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21642909 ↗
  • MalfunctionMar 13, 2025Connection Problem; Gas/Air Leak; Connection Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN TESTING WITH A TEST LUNG BEFORE STARTING THE VENTILATOR, THERE WAS A LEAK NOTED ON THE SIX DEVICES. THERE WAS ALSO A CRACK IN THE Y-PIECE REGION (ON THE VENTILATOR SIDE). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21601227 ↗
  • MalfunctionMar 13, 2025Connection Problem; Gas/Air Leak; Connection Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN TESTING WITH A TEST LUNG BEFORE STARTING THE VENTILATOR, THERE WAS A LEAK NOTED ON THE SEVEN DEVICES. THERE WAS ALSO A CRACK IN THE Y-PIECE REGION (ON THE VENTILATOR SIDE). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21601038 ↗
  • MalfunctionMar 13, 2025Connection Problem; Gas/Air Leak; Connection Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN TESTING WITH A TEST LUNG BEFORE STARTING THE VENTILATOR, THERE WAS A LEAK NOTED ON THE SIX DEVICES. THERE WAS ALSO A CRACK IN THE Y-PIECE REGION (ON THE VENTILATOR SIDE). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21591657 ↗
  • MalfunctionMar 13, 2025Connection Problem; Gas/Air Leak; Connection Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN TESTING WITH A TEST LUNG BEFORE STARTING THE VENTILATOR, THERE WAS A LEAK NOTED ON THE SIX DEVICES. THERE WAS ALSO A CRACK IN THE Y-PIECE REGION (ON THE VENTILATOR SIDE). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21591627 ↗
  • MalfunctionMar 12, 2025Connection Problem; Gas/Air Leak; Connection Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN TESTING WITH A TEST LUNG BEFORE STARTING THE VENTILATOR, THERE WAS A LEAK NOTED ON THE SIX DEVICES. THERE WAS ALSO A CRACK IN THE Y-PIECE REGION (ON THE VENTILATOR SIDE). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21590662 ↗
  • MalfunctionMar 12, 2025Connection Problem; Gas/Air Leak; Connection Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN TESTING WITH A TEST LUNG BEFORE STARTING THE VENTILATOR, THERE WAS A LEAK NOTED ON THE SEVEN DEVICES. THERE WAS ALSO A CRACK IN THE Y-PIECE REGION (ON THE VENTILATOR SIDE). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21590660 ↗
  • MalfunctionJan 31, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND CONNECTED THE NEGATIVE PRESSURE SUCTION LINE, THE NEGATIVE PRESSURE SUCTION BUTTON COULD NOT BE PRESSED. THIS PROBLEM HAD NOT OCCURRED DURING PREVIOUS USE. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE THREE SUCTION VALVES WITH THE CAP COULD NOT ROTATE PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21283722 ↗
  • MalfunctionJan 31, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND CONNECTED THE NEGATIVE PRESSURE SUCTION LINE, THE NEGATIVE PRESSURE SUCTION BUTTON COULD NOT BE PRESSED. THIS PROBLEM HAD NOT OCCURRED DURING PREVIOUS USE. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE THREE SUCTION VALVES WITH THE CAP COULD NOT ROTATE PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21283691 ↗
  • MalfunctionJan 31, 2025Nonstandard Device; Material Split, Cut or Torn; Nonstandard Device; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER THE CUSTOMER OPENED THE PACKAGE AND CONNECTED THE NEGATIVE PRESSURE SUCTION LINE, THE NEGATIVE PRESSURE SUCTION BUTTON COULD NOT BE PRESSED. THIS PROBLEM HAD NOT OCCURRED DURING PREVIOUS USE. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE THREE SUCTION VALVES WITH THE CAP COULD NOT ROTATE PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21283663 ↗
  • MalfunctionDec 11, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USAGE, THE BREATHING SYSTEM SUCTION HAD FAILED THE PERFORMED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20901776 ↗
  • MalfunctionDec 11, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USAGE, THE BREATHING SYSTEM SUCTION HAD FAILED THE PERFORMED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20901745 ↗
  • MalfunctionDec 10, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USAGE, THE BREATHING SYSTEM SUCTION HAD FAILED THE PERFORMED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20891678 ↗
  • MalfunctionDec 10, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE HAD FAIL LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20889903 ↗
  • MalfunctionNov 29, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE SUCTION CATHETER FAILED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20810888 ↗
  • MalfunctionNov 29, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE SUCTION CATHETER FAILED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20810832 ↗
  • MalfunctionNov 28, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USAGE, THE BREATHING SYSTEM SUCTION HAD FAILED THE PERFORMED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20805992 ↗
  • MalfunctionNov 26, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USAGE, THE BREATHING SYSTEM SUCTION HAD FAILED THE PERFORMED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20788664 ↗
  • MalfunctionNov 8, 2024Device Damaged Prior to Use; Gas/Air Leak; Device Damaged Prior to Use; Gas/Air Leak

    ACCORDING TO THE REPORTER, DURING USE, BOTH BREATHING SYSTEM SUCTION HAD A LEAK. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20641859 ↗

Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices