Medi-Trace Cadence 22770PC (UDI 20884527022394)Recalls & reported problems
Cardinal Health, Inc. · Adult Multi-Function Defibrillation Electrodes, Pre-connect · Class 2
All models of Medi-Trace Cadence (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 20884527022394), received between Dec 22, 2017 and Dec 26, 2025 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Arcing of Electrodes” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Arcing of Electrodes3
- Failure to Deliver Shock/Stimulation3
- Detachment of Device or Device Component1
- Gel Leak1
- Material Separation1
- No Pacing1
- Peeled/Delaminated1
Patient problems
- Insufficient Information3
- No Consequences Or Impact To Patient3
- Burn(s)2
- Bradycardia1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 26, 2025Arcing of Electrodes; Arcing of Electrodes
THE CUSTOMER REPORTED THERE WAS A PATIENT IN VFIB REQUIRING DEFIBRILLATION, PADS PUT ON PATIENT AND SHOCKED C 200J. AUDIBLE POPPING AND NOISE HEARD, SMELL OF SOMETHING BURNING/SMOKING. PATIENT SUCCESSFULLY DEFIBRILLATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23896100 ↗MalfunctionNov 22, 2023Arcing of Electrodes; Arcing of Electrodes
THE CUSTOMER REPORTED THE DEFIB PADS ARCED, AND THE PATIENT RECEIVED A MINOR BURN. THIS HAS BEEN REPORTED IN MEDWATCH. THE PATIENT DID NOT REQUIRE ANY TREATMENT.
Patient problems: Burn(s)
MAUDE report 18192205 ↗MalfunctionNov 22, 2023Arcing of Electrodes; Arcing of Electrodes
THE CUSTOMER REPORTED THE DEFIB PADS ARCED, AND THE PATIENT RECEIVED A MINOR BURN. THIS HAS BEEN REPORTED IN MEDWATCH. THE PATIENT WAS TREATED WITH POLYSPORIN AND 2X2.
Patient problems: Burn(s)
MAUDE report 18192031 ↗InjurySep 19, 2023No Pacing; No Pacing
CUSTOMER REPORTS: THE CLIENT NEEDED TO USE THE MULTIFUNCTION ELECTRODES TO DO EXTERNAL PACING TO A PATIENT AND IT DID NOT WORK. THE ELECTRODE GENERATED NO PULSE. PER ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER ON (B)(6) 2023, DESPITE THE FREQUENCY OF ZOLL SET AT 60 BPM, THE PATIENT REMAINED BRADYCARDIUM AT 30 BPM (FEMORAL PULSE). THE ELECTRODE SEEMED TO RECEIVE A SIGNAL BECAUSE WE SAW SPICULE AT THE MONITOR. WE INSTALLED THE ELECTRODES TO DO EXTERNAL PACING AND NOT DEFIBRILLATION. THE SIGNAL WAS INACCURATE BECAUSE THE PULSE REQUESTED AT THE ZOLL WAS 60 BPM, BUT THE PATIENT REMAINED WITH A FEMORAL PULSE OF 30 BPM. THERE WAS A DELAY TO TREATMENT, BUT NOT IN DEFIBRILLATION BUT RATHER IN RESTORING THE PATIENT'S PULSE. WE HAD TO INSTALL A TEMPORARY ENDOVENOUS PACEMAKER FOR THE PATIENT AND WE HAD TO PUT ISUPREL AND DOPAMINE. THE PATIENT WAS DISCHARGED.
Patient problems: Bradycardia
MAUDE report 17777709 ↗MalfunctionFeb 23, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER REPORTED THAT THE PRE-CONNECT DEFIBRILLATION ELECTRODE FAILED TO DISCHARGE A CHARGE TO A PATIENT DURING CODE BLUE AND A SMART DEFIBRILLATOR WAS USED DURING THE INCIDENT. THE INTERNAL ONSITE HOSPITAL DEPARTMENT BIOMED (EQUIPMENT MANAGEMENT TEAM, INCLUDING QUALITY ASSURANCE (QA), MATERIAL) DID TESTING ON 22770PC (LOT # 121672X) AND THE SAMPLES WERE DISCHARGED PROPERLY.
Patient problems: Insufficient Information
MAUDE report 13586825 ↗MalfunctionFeb 23, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER REPORTED THAT THE DEFIBRILLATION CHARGE NOT HAPPENED TO A PATIENT DURING CODE BLUE AND A SMART DEFIBRILLATOR WAS USED DURING THE INCIDENT. THE INTERNAL ONSITE HOSPITAL DEPARTMENT BIOMED (EQUIPMENT MANAGEMENT TEAM, INCLUDING QUALITY ASSURANCE (QA), MATERIAL) DID TESTING ON (B)(4) (LOT # 226434X) AND THE SAMPLES WERE DISCHARGED PROPERLY.
Patient problems: Insufficient Information
MAUDE report 13586736 ↗MalfunctionFeb 23, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER REPORTED THAT THE DEFIBRILLATION ELECTRODE FAILED TO DISCHARGE A CHARGE TO PATIENT DURING CODE BLUE AND A SMART DEFIBRILLATOR USED DURING THE INCIDENT. ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE PATIENT EXPIRED. SEVERAL ATTEMPTS TO GATHER INFORMATION FROM THE CUSTOMER WERE MADE. TO DATE, NO RESPONSE HAS BEEN RECEIVED. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, THE REPORT WILL BE UPDATED.
Patient problems: Insufficient Information
MAUDE report 13586720 ↗MalfunctionJun 17, 2020Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED THAT THE DEFIBRILLATION ELECTRODES ADHERED VERY POORLY. IT WAS NOT THE FIRST TIME THAT THE ADHESIVE SURFACE OR THE CONTACT GEL OF THE ELECTRODES CAME OFF WITH THE REMOVAL OF THE PROTECTIVE FOIL BUT THIS TIME IT WAS SIMPLY A BIT MORE EXTENSIVE. NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE AT THIS TIME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10165432 ↗MalfunctionSep 30, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THAT THE DEFIB PAD WIRE UNATTACHED FROM THE PAD WHILE ATTACHED TO A PATIENT THAT WAS GIVEN DEFIBRILLATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9136287 ↗MalfunctionSep 25, 2019Material Separation; Material Separation
THE CUSTOMER REPORTED THAT THE DEFIBRILLATOR WIRE RIPPED OFF OF THE PAD WHILE APPLYING THE PRODUCT TO THE PATIENT. THE PRODUCT WAS NOT USED. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9120678 ↗MalfunctionDec 22, 2017Gel Leak; Gel Leak
THE CUSTOMER REPORTED THAT THE GEL FROM THE DEFIB PAD IS PEELING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7143531 ↗
Other Pacemaker, Cardiac, External Transcutaneous (Non-Invasive) devices
- Tempus LS-MANUAL 3.940589Schiller AG · Pacemaker, Cardiac, External Transcutaneous (Non-Invasive)4 recalls34 deaths391 reports
- M SERIES BIPHASIC 40010011100123010Zoll Medical Corporation · PACEMAKER, CARDIAC, EXTERNAL TRANSCUTANEOUS (NON-INVASIVE)1 death247 reports
- M SERIES BIPHASIC 40010511100123010Zoll Medical Corporation · PACEMAKER, CARDIAC, EXTERNAL TRANSCUTANEOUS (NON-INVASIVE)77 reports
- M SERIES BIPHASIC 40010011100103011Zoll Medical Corporation · PACEMAKER, CARDIAC, EXTERNAL TRANSCUTANEOUS (NON-INVASIVE)48 reports
- Medi-Trace Cadence 22550RCardinal Health, Inc. · Pacemaker, Cardiac, External Transcutaneous (Non-Invasive)1 death20 reports
- M SERIES BIPHASIC 40010711100123010Zoll Medical Corporation · PACEMAKER, CARDIAC, EXTERNAL TRANSCUTANEOUS (NON-INVASIVE)14 reports
- M SERIES BIPHASIC 60010011100123010Zoll Medical Corporation · PACEMAKER, CARDIAC, EXTERNAL TRANSCUTANEOUS (NON-INVASIVE)9 reports
- Medi-Trace Cadence 22550PCCardinal Health, Inc. · Pacemaker, Cardiac, External Transcutaneous (Non-Invasive)8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page