Microstream Advance Neonatal-Infant Intubated CO2 Filter Line Neo/Inf Intub CO2 Line LTermRecalls & reported problems
Philips Medical Systems Hsg · Microstream Advance Neonatal-Infant Intubated CO2 Filter Line, Extended Duration · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20884838093137), received between Aug 25, 2023 and Aug 25, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Suction Failure” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Suction Failure5
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information3
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 25, 2023Suction Failure; Suction Failure
PHILIPS RECEIVED A COMPLAINT ON THE NEONATAL-INFANT INTUBATED CO2 FILTER LINE, EXTENDED DURATION INDICATING A POTENTIAL NEONATAL PATIENT SAFETY ISSUE DUE TO MICROSTREAM CO2 ADAPTOR & SMITHS ET TUBE STICKING TOGETHER RESULTING IN NOT BEING ABLE TO SUCTION A CRITICALLY ILL PATIENT. SOME INSTANCES RESULTED IN A PATIENT HAVING TO BE EXTUBATED SO THAT THEY CAN BE SUCTIONED AND THE AIRWAY HAVING TO BE REINSERTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17627732 ↗MalfunctionAug 25, 2023Suction Failure; Suction Failure
THE CUSTOMER STATED THAT THE PATIENT REQUIRED AN ADDITIONAL MEDICAL PROCEDURE (OUTSIDE THE SCOPE OF NORMAL TREATMENT). THE REPORTED POTENTIAL HARM IS CLASSIFIED AS SERIOUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17625374 ↗MalfunctionAug 25, 2023Suction Failure; Suction Failure
THE CUSTOMER STATED THAT THE PATIENT REQUIRED AN ADDITIONAL MEDICAL PROCEDURE (OUTSIDE THE SCOPE OF NORMAL TREATMENT). THE REPORTED POTENTIAL HARM IS CLASSIFIED AS SERIOUS.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17625139 ↗MalfunctionAug 25, 2023Suction Failure; Suction Failure
PHILIPS RECEIVED A COMPLAINT ON THE NEONATAL-INFANT INTUBATED CO2 FILTER LINE, EXTENDED DURATION INDICATING A POTENTIAL NEONATAL PATIENT SAFETY ISSUE DUE TO MICROSTREAM CO2 ADAPTOR & SMITHS ET TUBE STICKING TOGETHER RESULTING IN NOT BEING ABLE TO SUCTION A CRITICALLY ILL PATIENT. SOME INSTANCES RESULTED IN A PATIENT HAVING TO BE EXTUBATED SO THAT THEY CAN BE SUCTIONED AND THE AIRWAY HAVING TO BE REINSERTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17624869 ↗MalfunctionAug 25, 2023Suction Failure; Suction Failure
PHILIPS RECEIVED A COMPLAINT ON THE NEONATAL-INFANT INTUBATED CO2 FILTER LINE, EXTENDED DURATION INDICATING A POTENTIAL NEONATAL PATIENT SAFETY ISSUE DUE TO MICROSTREAM CO2 ADAPTOR & SMITHS ET TUBE STICKING TOGETHER RESULTING IN NOT BEING ABLE TO SUCTION A CRITICALLY ILL PATIENT. SOME INSTANCES RESULTED IN A PATIENT HAVING TO BE EXTUBATED SO THAT THEY CAN BE SUCTIONED AND THE AIRWAY HAVING TO BE REINSERTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17624431 ↗
Other ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page