Skip to content

Microstream Advance Neonatal-Infant Intubated CO2 Filter Line Neo/Inf Intub CO2 Line LTermRecalls & reported problems

Philips Medical Systems Hsg · Microstream Advance Neonatal-Infant Intubated CO2 Filter Line, Extended Duration · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20884838093137), received between Aug 25, 2023 and Aug 25, 2023 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Suction Failure” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Suction Failure5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information3

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2023Suction Failure; Suction Failure

    PHILIPS RECEIVED A COMPLAINT ON THE NEONATAL-INFANT INTUBATED CO2 FILTER LINE, EXTENDED DURATION INDICATING A POTENTIAL NEONATAL PATIENT SAFETY ISSUE DUE TO MICROSTREAM CO2 ADAPTOR & SMITHS ET TUBE STICKING TOGETHER RESULTING IN NOT BEING ABLE TO SUCTION A CRITICALLY ILL PATIENT. SOME INSTANCES RESULTED IN A PATIENT HAVING TO BE EXTUBATED SO THAT THEY CAN BE SUCTIONED AND THE AIRWAY HAVING TO BE REINSERTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17627732 ↗
  • MalfunctionAug 25, 2023Suction Failure; Suction Failure

    THE CUSTOMER STATED THAT THE PATIENT REQUIRED AN ADDITIONAL MEDICAL PROCEDURE (OUTSIDE THE SCOPE OF NORMAL TREATMENT). THE REPORTED POTENTIAL HARM IS CLASSIFIED AS SERIOUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17625374 ↗
  • MalfunctionAug 25, 2023Suction Failure; Suction Failure

    THE CUSTOMER STATED THAT THE PATIENT REQUIRED AN ADDITIONAL MEDICAL PROCEDURE (OUTSIDE THE SCOPE OF NORMAL TREATMENT). THE REPORTED POTENTIAL HARM IS CLASSIFIED AS SERIOUS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17625139 ↗
  • MalfunctionAug 25, 2023Suction Failure; Suction Failure

    PHILIPS RECEIVED A COMPLAINT ON THE NEONATAL-INFANT INTUBATED CO2 FILTER LINE, EXTENDED DURATION INDICATING A POTENTIAL NEONATAL PATIENT SAFETY ISSUE DUE TO MICROSTREAM CO2 ADAPTOR & SMITHS ET TUBE STICKING TOGETHER RESULTING IN NOT BEING ABLE TO SUCTION A CRITICALLY ILL PATIENT. SOME INSTANCES RESULTED IN A PATIENT HAVING TO BE EXTUBATED SO THAT THEY CAN BE SUCTIONED AND THE AIRWAY HAVING TO BE REINSERTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17624869 ↗
  • MalfunctionAug 25, 2023Suction Failure; Suction Failure

    PHILIPS RECEIVED A COMPLAINT ON THE NEONATAL-INFANT INTUBATED CO2 FILTER LINE, EXTENDED DURATION INDICATING A POTENTIAL NEONATAL PATIENT SAFETY ISSUE DUE TO MICROSTREAM CO2 ADAPTOR & SMITHS ET TUBE STICKING TOGETHER RESULTING IN NOT BEING ABLE TO SUCTION A CRITICALLY ILL PATIENT. SOME INSTANCES RESULTED IN A PATIENT HAVING TO BE EXTUBATED SO THAT THEY CAN BE SUCTIONED AND THE AIRWAY HAVING TO BE REINSERTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17624431 ↗

Other ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices

All ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page