Medtronic Xomed, Inc. · BLADE 1884016HR RAD60 5PK M4 4MM ROTATE · Class 1
All models of RAD (29)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 20885074080646), received between Dec 15, 2017 and Jul 22, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Vibration3
- Material Fragmentation1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient2
- Insufficient Information1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25902469 ↗MalfunctionDec 28, 2025Vibration; Vibration
IT WAS REPORTED THAT DURING THE ESS PROCEDURE, THERE WAS AN UNUSUAL DISCOMFORT (A SENSATION THAT THE BLADE TIP WAS FLOATING). RESETTING DID NOT CHANGE THE SITUATION, SO IT WAS DETERMINED TO BE A MALFUNCTION. A SPARE PRODUCT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23904067 ↗MalfunctionDec 4, 2025Vibration; Vibration
IT WAS REPORTED THAT WHEN TESTED THE BLADE ON THE HANDPIECE MADE A STRANGE NOISE AND DIDN'T FIT PROPERLY. THEY TRIED ON ANOTHER HANDPIECE AND THE SAME PROBLEM OCCURRED. THEY FOUND BLADE TO BE DAMAGED WITHOUT ANY FRAGMENTS. PROCEDURE WAS COMPLETED WITH BACK UP PRODUCT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23724091 ↗MalfunctionJun 25, 2025Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT IN ESS PROCEDURE, THE TIP OF THE BLADE DID NOT ROTATE IN 2 OF THE SAME LOT DURING USE, THE TIP WAS DAMAGED WHICH WAS NON-SPECIFIC BROKEN, PROCEDURE WAS EXTENDED FOR A FEW MINUTES WHICH WAS NEEDED FOR REPLACEMENT OF THE DEVICE. A SPARE PRODUCT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22329293 ↗MalfunctionApr 25, 2025Vibration; Vibration
IT WAS REPORTED THAT WHILE DOING AN ENT FUSION CASE, A 60 ROTATABLE BLADE WAS USED. IT WAS NOTED THAT THE BLADE WAS NOT ROTATING, IT WAS JUST MOVING BACK AND FORTH. A SECOND BLADE WAS OPENED, AND IT WAS ALSO NOT ROTATING, UPON FURTHER REVIEW, THE LOT NUMBERS WERE THE SAME. A 3RD BLADE WAS OPENED, WITH A DIFFERENT LOT NUMBER, AND THAT BLADE HAD NO ISSUES. THERE WAS NO KNOWN PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 21908586 ↗MalfunctionJun 1, 2022Break; Break
A HEALTH CARE PROVIDER (HCP) REPORTED THAT THE BLADE DID NOT WORK CORRECTLY DURING SINUS SURGERY AND HAD THE STAFF SAVE IT. HE PROCEEDED WITH A DIFFERENT BLADE, BUT THE STAFF DID NOT KNOW THE DETAILS OF WHAT MALFUNCTIONED. ON ADDITIONAL INFORMATION, IT INDICATED THE INTERIOR OF THE BLADE BECAME DISCONNECTED. THERE WAS NO CASE DELAY NO PATIENT IMPACT. NO BROKEN PIECES OF THE REPORTED PRODUCT REMAIN INSIDE THE PATIENT'S BODY. ADDITIONAL INFORMATION RECEIVED AND IT CLARIFIED THAT CONNECTION ISSUE WAS NOT AN ELECTRICAL CONNECTION ISSUE, IT WAS LIKE THE BLADES 2 PIECES CAME APART AND WOULD NOT OSCILLATE CORRECTLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14566903 ↗MalfunctionApr 25, 2022Break; Break
IT WAS REPORTED THAT THE SHAFT OF THE BLADE BROKE, NO PATIENT IMPACT. OTHER BLADE OF SAME LOT # USED AND WAS WORKING FINE. ON FOLLOW-UP IT WAS REPORTED THAT NO PIECES FELL INTO PATIENT OR REPORTED. THE INCIDENT OCCURRED DURING SURGERY. AFTER SURGERY CUSTOMER DID A SCAN TO MAKE SURE NO PIECES WERE IN THE PATIENT, IT WAS ALL CLEAR. THE PROCEDURE WAS A FESS. THE PRODUCT WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14193081 ↗MalfunctionMay 13, 2021Material Fragmentation; Material Fragmentation
ADDITIONAL INFORMATION RECEIVED MENTIONED THAT THE BLADE BROKE IN HALF WHILE USING ON A PATIENT. THE FRAGMENTS CAME OFF OR GOT DETACHED FROM THE DEVICE. ANOTHER BLADE WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11819444 ↗MalfunctionSep 23, 2019Break; Break
A HEALTH CARE PROVIDER (HCP) REPORTED VIA A MANUFACTURER REPRESENTATIVE THAT THE TIP OF THE BLADE WAS BROKEN DURING AN ESS (ENDOSCOPIC SINUS SURGERY) PROCEDURE. IT WAS NOTED THAT THERE WAS NO BROKEN PIECE(S) OF THE DEVICE THAT WAS LEFT INSIDE THE PATIENT'S BODY. IT WAS THE INITIAL USE OF THE DEVICE. THERE WAS NO DELAY IN THE PROCEDURE AS A RESULT OF THIS EVENT. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS NO PATIENT IMPACT OR INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9104276 ↗MalfunctionDec 15, 2017Break; Break
THE CUSTOMER REPORTED THAT THE RAD 60 BLADE TIP DETACHED AS SOON AS THE OPERATOR BEGAN USING THE BLADE DURING AN ENDOSCOPIC SINUS SURGERY PROCEDURE. NO FRAGMENT RESULTED FROM THE BREAKAGE. THERE WAS NO PATIENT IMPACT REPORTED AS A RESULT OF THIS EVENT. THE PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7123122 ↗
Other BUR, EAR, NOSE AND THROAT devices
- Quadcut® 1884380HRMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall7 reports
- Silver Bullet® 1882905HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall3 reports
- Diego Elite MBUR6060FRCVGyrus ACMI, LLC · BUR, EAR, NOSE AND THROAT1 recall2 reports
- Quadcut® 1884380HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall1 report
- RAD® 1882916Medtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall1 report
- N/A 1884032Medtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
- DIEGO ELITE MBUR5060FRCVGyrus ACMI, LLC · BUR, EAR, NOSE AND THROAT1 recall0 reports
- Tricut® 1884033HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page