NA 74013471Recalls & reported problems
Smith & Nephew, Inc. · RIMMED SPEED PIN 45MM STERILE · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20885556069985), received between Nov 13, 2020 and Nov 18, 2023 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Mechanical Jam1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2023Fracture; Mechanical Jam; Fracture; Mechanical Jam
IT WAS REPORTED THAT DURING A TKA, USING A SPEED PIN QK CONNECT ADAPTER, WHEN CONNECTING THE THREADED RIM SPEED PIN 45MM STERILE TO THE FEMORAL CUTTING GUIDE, IT BROKE AND BECAME STUCK IN THE PIN GUIDE. NO PIECES FELL INSIDE OF THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18166917 ↗InjuryJul 3, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, A TRIAL TIBIA TRAY WAS INSERTED AND PINNED IN PLACE TO CONFIRM SIZING AND ROTATION. THIS WAS NOTED TO JUST AVOID THE METAL TIBIAL INTERFERENCE SCREW, SO IT DID NOT HAVE TO BE EXTRACTED. THE ANTERIOR RIM SPEED PIN 45MM STERILE THAT WAS DRILLED CONTACTED THE INTERFERENCE SCREW AND SNAPPED, BUT THE RETAINED SEGMENT WAS THREADED AND BURIED IN BONE SO WAS LEFT IN PLACE SO AS NOT TO DISRUPT ANY OF THE PROXIMAL METAPHYSEAL BONE. IT IS UNKNOWN IF SURGERY WAS DELAYED AS CONSEQUENCE OF THE ISSUE. THE PATIENT IS DOING WELL AND HAS BEEN WORKING WITH PHYSICAL THERAPY TWICE A WEEK. HE CONTINUES TO HAVE STIFFNESS THAT IMPROVES AFTER EXERCISE; MOST LIKELY DUE TO HIS PREVIOUS KNEE SURGERY. X-RAYS OF THE RIGHT KNEE DEMONSTRATE THE TOTAL KNEE ARTHROPLASTY PROSTHESIS IN GOOD POSITION WITH NO EVIDENCE OF LOOSENING, OSTEOLYSIS, OR SUBSIDENCE NOTED. GOOD ALIGNMENT IS NOTED.
Patient problems: Foreign Body In Patient
MAUDE report 17249122 ↗MalfunctionAug 1, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, WHILE PLACING A RIM SPEED PIN 45MM STERILE INTO A GII MIS DCF AP CT BLK 8, THE HEAD OF THE PIN BROKE OFF IN THE SPEED PIN QK CONNECT ADAPTER. THE PROCEDURE WAS COMPLETED, WITH A NON-SIGNIFICANT DELAY, USING A SMITH AND NEPHEW BACK UP DEVICE. NO INJURY TO PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15141789 ↗InjuryNov 13, 2020Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DURING TOTAL KNEE ARTHROPLASTY THE SPEEDPIN HEAD WORN AWAY, MAKING IT IMPOSSIBLE TO REMOVE FROM THE CUTTING BLOCK WITH A DRILL. A PIN PULLER WAS USED TO TWIST IT OUT. THERE WAS A DELAY REPORTED OF LESS THAN OR EQUAL TO 30 MINUTES. THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACKUP DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10840403 ↗
Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- Scorpio 73-3710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls1 report
- Scorpio 73-3110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls0 reports
- Arthrex® AR-503-TTTEArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall14 reports
- Arthrex® AR-503-TTTJArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- Arthrex® AR-503-TTTGArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10011-300Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10003-400Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page