Skip to content

NA 74013480Recalls & reported problems

Smith & Nephew, Inc. · NON RIMMED SPEED PIN 65MM STERILE · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 20885556070929), received between Aug 20, 2018 and Nov 10, 2025 — classified as 6 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Break4
  • Flaked2
  • Material Twisted/Bent2
  • Device Damaged Prior to Use1
  • Illegible Information1
  • Improper or Incorrect Procedure or Method1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient4
  • No Consequences Or Impact To Patient3
  • Device Embedded In Tissue or Plaque1
  • Perforation1
  • Rupture1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY THE TIP OF ONE (1) NONRIM SPEED PIN 65 STERILE BROKE OFF DURING REMOVAL. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR SN BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23518543 ↗
  • MalfunctionSep 5, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY, ONE (1) NONRIM SPEED PIN 65 STERILE BROKE OFF IN ONE (1) MIS PIN DRIVER, ALL PIECES ACCOUNTED FOR, IT OCCURRED WHILE DRILLING THE PIN THROUGH A CUT GUIDE AND GOT BOUND UP, NO PIECES TO REMOVE FROM PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22985161 ↗
  • InjuryMar 26, 2024Fracture; Improper or Incorrect Procedure or Method; Fracture; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT, A PATIENT WHO HAD A PREVIOUS ACL REPAIR UNDERWENT A CORI ASSISTED TKA PROCEDURE, AND THEREFORE ALREADY HAD A SCREW IN THEIR TIBIA. DURING SURGERY, AN ADJUSTABLE TIBIA CUT GUIDE WAS ADJUSTED WITH A NONRIM SPEED PIN 65 STERILE AND THE SPEED PIN SNAPPED IN HALF WHEN IT HIT THE ACL SCREW THAT WAS IN THE TIBIA (THE PROXIMAL AND MEDIAL PIN HOLE ON THE FIXATION BLOCK). ANOTHER PIN WAS ADDED TO THE DISTAL PIN HOLE ON THE FIXATION BLOCK. THE SURGEON LATER USED A VICE GRIP AND OSTEOTOME TO REMOVE THE THREADS OF THE BROKEN PIN AND THE ACL SCREW FROM THE TIBIA, REQUIRING ADDITIONAL CUTS. THE PROCEDURE WAS COMPLETED ROBOTICALLY, AFTER A SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: Perforation; Rupture

    MAUDE report 18980038 ↗
  • MalfunctionFeb 7, 2024Fracture; Separation Failure; Fracture; Separation Failure

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) NONRIM SPEED PIN 65 STERILE BENT AND GOT STUCK INSIDE OF A TIBIAL CUTTING GUIDE THREE DEG RIGHT. ADDITIONALLY, THE TIP OF THE NONRIM SPEED PIN 65 STERILE BROKE OFF INSIDE OF A MIS PIN DRIVER. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18663569 ↗
  • MalfunctionJul 27, 2023Manufacturing, Packaging or Shipping Problem; Material Twisted/Bent; Manufacturing, Packaging or Shipping Problem; Material Twisted/Bent

    IT WAS REPORTED THAT, DURING SET UP OR INSPECTION FOR A TKA, IT WAS FOUND THAT ONE (1) NONRIM SPEED PIN 65 STERILE WAS BENT OUT OF PACKAGE. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. SINCE INCIDENT OCCURRED BEFORE PROCEDURE; PATIENT WAS NOT YET INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17410965 ↗
  • MalfunctionJul 27, 2023Device Damaged Prior to Use; Material Twisted/Bent; Device Damaged Prior to Use; Material Twisted/Bent

    IT WAS REPORTED THAT DURING SET UP FOR A TKA, A PIN IN A NONRIM SPEED PIN 65 STERILE WAS BENT WHEN REMOVED FROM PACKAGING. BOX WAS NOT DAMAGED BEFORE OPENING. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. SINCE INCIDENT OCCURRED BEFORE PROCEDURE; PATIENT WAS NOT YET INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17410963 ↗
  • InjuryJun 15, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING TKA SURGERY, 2 TO 3MM FROM THE TIP OF A NONRIM SPEED PIN 65 STERILE BROKE OFF. ON A POST-OPERATIVE X-RAY, IT WAS FOUND THAT THERE WAS A FOREIGN BODY IN THE FEMUR, THE BROKEN PIECE WAS NOT PLANNED TO BE REMOVED BECAUSE THE PIECE WAS LOCATED DEEP IN THE BONE.. THE PROCEDURE HAD BEEN FINISHED UNEVENTFULLY. CURRENT PATIENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 17137108 ↗
  • InjuryMay 11, 2022Break; Break

    IT WAS REPORTED THAT THE TIP OF NONRIM SPEED PIN 65 STERILE BROKE OFF AND LODGED INTO SOFT TISSUE OF THE KNEE. THIS WAS NOTICED AFTER TKA SURGERY. THE CONDITION OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 14366116 ↗
  • InjuryOct 11, 2021Break; Break

    IT WAS REPORTED DURING A TKA PROCEDURE, THAT 2 NONRIM SPEED PIN 65 STERILE BROKE OFF IN A PATIENTS KNEE WHEN THE SURGEON HIT THE SCREW IN BOTH THE FEMUR AND TIBIA. THE PIECE REMAINS INSIDE THE PATIENT, SURGEON WAS NOT ABLE TO RETRIEVE IT. PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE, NO DELAY REPORTED. AN INJURY IS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 12611777 ↗
  • InjuryOct 11, 2021Break; Break

    IT WAS REPORTED DURING A TKA PROCEDURE, THAT (2) NONRIM SPEED PIN 65 STERILE BROKE OFF IN A PATIENTS KNEE WHEN THE SURGEON HIT THE SCREW IN BOTH THE FEMUR AND TIBIA. THE PIECE REMAINS INSIDE THE PATIENT, SURGEON WAS NOT ABLE TO RETRIEVE IT. PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE, NO DELAY REPORTED. AN INJURY IS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 12611758 ↗
  • MalfunctionNov 19, 2019Flaked; Flaked

    IT WAS REPORTED THAT DURING SURGERY, AFTER THE DISTAL FEMORAL CUTTING, THERE WERE THE SHAVING SUBSTANCES. THE MARKING OF THE PIN WAS DAMAGED AND VANISHED, SO THE SHAVINGS WERE CONSIDERED THEY WERE FROM THE PIN. THIS ISSUE WAS FOUND AFTER SCREWING THE PIN AND DID NOT AFFECT THE SURGERY. EVERYTHING WAS REMOVED FROM PATIENT. DELAY LESS THAN 30 MINUTES AND NO INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9340507 ↗
  • MalfunctionNov 19, 2019Flaked; Flaked

    IT WAS REPORTED THAT DURING SURGERY, AFTER THE DISTAL FEMORAL CUTTING, THERE WERE THE SHAVING SUBSTANCES. THE MARKING OF THE PIN WAS DAMAGED AND VANISHED, SO THE SHAVINGS WERE CONSIDERED THEY WERE FROM THE PIN. THIS ISSUE WAS FOUND AFTER SCREWING THE PIN AND DID NOT AFFECT THE SURGERY. EVERYTHING WAS REMOVED FROM PATIENT. DELAY LESS THAN 30 MINUTES AND NO INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9340368 ↗
  • MalfunctionApr 1, 2019Illegible Information; Illegible Information

    IT WAS REPORTED THAT DURING SURGERY, AFTER THE DISTAL FEMORAL CUTTING, THERE WERE THE SHAVING SUBSTANCES. THE MARKING OF THE PIN WAS DAMAGED AND VANISHED, SO THE SHAVINGS WERE CONSIDERED THEY WERE FROM THE PIN. THIS ISSUE WAS FOUND AFTER SCREWING THE PIN AND DID NOT AFFECT THE SURGERY. EVERYTHING WAS REMOVED FROM PATIENT. DELAY LESS THAN 30 MINUTES AND NO INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8470815 ↗
  • InjuryAug 20, 2018Break

    IT WAS REPORTED THAT TIP OF SPEED PIN (5MM FROM TIP) BROKE AND REMAINED IN FEMORAL. IT WAS FOUND DURING X-RAY CHECK IMMEDIATE POST-OPERATION. THE SURGEON DECIDED THAT BROKEN TIP LEFT IN PATIENT BECAUSE IT WAS LOCATED DIFFICULT REGION.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7799518 ↗

FDA clearance and classification

Cleared through 510(k) K113038 on Jan 5, 2012 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew Inc. Unicompartmental And Biocompartmental Knee System Instruments”. The FDA review took 85 days. FDA record ↗

The FDA regulates product code HSX (prosthesis, knee, femorotibial, non-constrained, cemented, metal/polymer) under 21 CFR 888.3520 as a Class II device. 21 CFR 888.3520 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 74013480

Has the NA 74013480 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20885556070929), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 74013480?

The FDA's MAUDE database lists 14 adverse event reports for this device, received between Aug 20, 2018 and Nov 10, 2025. In those reports, the most common specific device problems are “Fracture” (5), “Break” (4) and “Flaked” (2); the most common patient problems are “Foreign Body In Patient” (4), “Device Embedded In Tissue or Plaque” (1) and “Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 74013480?

None of the reports are classified as deaths, 6 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 74013480 FDA approved?

It was cleared through 510(k) K113038 on Jan 5, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA 74013480, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 74013480). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page