Atrieve™ Vascular Snare Kit 381003004Recalls & reported problems
Argon Medical Devices, Inc. · Atrieve™ Vascular Snare Kit 3.2F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20886333003970), received between Oct 11, 2021 and Jan 11, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Output Problem1
- Physical Resistance/Sticking1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 11, 2022Physical Resistance/Sticking
AFTER THE SNARE OF THE ATRIEVE CACHED THE TARGET FOREIGN OBJECT, THE SNARE AND THE FOREIGN OBJECT WERE NOT ABLE TO WITHDRAW INTO THE GUIDING CATHETER BECAUSE THE SHAFT DID NOT MOVE SMOOTHLY AND THERE WAS RESISTANCE DURING MANIPULATING THE SHAFT. THE UNIT WAS SWITCHED TO ANOTHER NEW ONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13226023 ↗MalfunctionOct 11, 2021Detachment of Device or Device Component; Output Problem
DETACHMENT OF THE PLATINUM MAKER LOOP WAS REPORTED. DETAILS: THE ATIREVE WAS USED TO REMOVE A GUIDE WIRE LEFT IN LMT AT PCI PROCEDURE. THE DOCTOR TRIED TO REMOVE THE GUIDE WIRE BY ATRIEVE WITH A 6F GUIDING CATHETER SEVERAL TIMES. THEN AFTER THE ATRIEVE AND THE DELIVERY CATHETER WAS TAKEN OUT FROM THE 6F GUIDING CATHETER AND WHEN ANOTHER MEDICAL DEVICE WAS INSERTED TO THE GUIDING CATHETER, IT STUCK IN THE CATHETER. THE GUIDING CATHETER WAS CHECKED UNDER X-RAY, AND A BLACK OBJECT WAS CONFIRMED IN THE CATHETER. THE GUIDING CATHETER WAS REMOVED FROM THE PATIENT AND FOUND THAT THE OBJECT STUCK IN THE CATHETER WAS THE PLATINUM MARKER LOOP OF THE ATRIEVE DELIVERY CATHETER. THE PLATINUM LOOP WAS REMOVED FROM THE GUIDING CATHETER THEN THE PROCEDURE WAS CARRIED ON AND COMPLETED. THE DETACHED PLATINUM LOOP WAS NOT RETURNED ONLY THE SNARE AND THE DELIVERY CATHETER WAS RETURNED, HOWEVER, A PHOTO WAS PROVIDED. THE CUSTOMER WANTS TO KNOW THE ROOT CAUSE OF THE DETACHMENT OF THE LOOP. ADDITIONAL INFORMATION: THE COMPLAINT LOT NUMBER WAS PURCHASED BY COSMOTEC FROM ARGON DIRECTLY.
Patient problems: Insufficient Information
MAUDE report 12603198 ↗
Other Device, Percutaneous Retrieval devices
- Gunther Tulip G13287Cook Incorporated · Device, percutaneous retrieval1 death98 reports
- CloverSnare G53008Cook Incorporated · Device, percutaneous retrieval32 reports
- Indy OTW G51836Cook Incorporated · Device, percutaneous retrieval17 reports
- Aveir™ LSCR111St. Jude Medical, Inc. · Device, percutaneous retrieval2 deaths15 reports
- Bard Snare OUS SRK35Heraeus Medical Components LLC · Device, Percutaneous Retrieval13 reports
- Indy OTW G51835Cook Incorporated · Device, percutaneous retrieval8 reports
- EN Snare® 00884450452462Merit Medical Systems, Inc. · Device, percutaneous retrieval7 reports
- Gunther Tulip G13287Cook Incorporated · Device, percutaneous retrieval6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page