Skip to content

Atrieve™ Vascular Snare Kit 382006020Recalls & reported problems

Argon Medical Devices, Inc. · Atrieve™ Vascular Snare Kit 6F · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20886333004007), received between Oct 2, 2023 and Oct 2, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE ATRIEVE WAS OPENED TO USE FOR EVAR PROCEDURE. THE STENT GRAFT GOING TO USE FOR EVAR WAS ENDURANT BY MEDTRONIC. A 7F/100CMGUIDING CATHETER PARENT FROM MEDIKIT WAS INSERTED TO THE PATIENT FROM BRACHIAL ARTERY. TO PULL THOUGH, THE ATRIEVE DELIVERY CATHETER WAS INSERTED TO THE PARENT GUIDING CATHETER AND PLACED IN THE TARGET AREA. AND THEN, THE ATRIEVE SNARE WAS INSERTED TO THE GUIDING CATHETER. AT THAT TIME, THE PHYSICIAN NOTICED THAT THE RADIOPAQUE BAND ON THE TIP OF THE DELIVERY CATHETER HAD COME OFF IN THE BLOOD VESSEL. IT BECAME UNABLE TO CONFIRM THE PLACEMENT OF THE TIP OF THE DELIVERY CATHETER WITHOUT THE RADIOPAQUE BAND, THEREFORE, THE DELIVERY CATHETER AND THE SNARE WAS REMOVED. AFTER THE PROCEDURE, IT WAS CONFIRMED THE DETACHED RADIOPAQUE BAND REMAINED AND STAYED AT TH 9 INSIDE OF THE PATIENT. THE PATIENT IS STABLE AND ADDITIONAL TREATMENT FOR THE REMAINING RADIOPAQUE BAND IN THE PATIENT HAS NOT BEEN TAKEN.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 17858266 ↗

FDA clearance and classification

Cleared through 510(k) K021606 on May 31, 2002 (decision: Substantially Equivalent). Submitted by Medical Device Technologies, Inc. as “En-Snare Endovascular Snare And Catheter”. The FDA review took 15 days. FDA record ↗

The FDA regulates product code MMX (device, percutaneous retrieval) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Atrieve™ Vascular Snare Kit 382006020

Has the Atrieve™ Vascular Snare Kit 382006020 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20886333004007), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Atrieve™ Vascular Snare Kit 382006020?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Oct 2, 2023. That report lists device problem “Detachment of Device or Device Component” (1); patient problem “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Atrieve™ Vascular Snare Kit 382006020?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Atrieve™ Vascular Snare Kit 382006020 FDA approved?

It was cleared through 510(k) K021606 on May 31, 2002. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Atrieve™ Vascular Snare Kit 382006020, and is it still sold?

The FDA's device database lists Argon Medical Devices, Inc. as the company responsible for it (brand Atrieve™ Vascular Snare Kit, model 382006020). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Percutaneous Retrieval devices

All Device, Percutaneous Retrieval devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page