CODMAN 80-1399Recalls & reported problems
Cerenovus, Inc. · CODMAN Surgical Patties 1/4 Inch x 1/4 Inch (0.64cm x 0.64cm) · Class 2
All models of CODMAN (41)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20886704036477), received between May 17, 2023 and May 17, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Separation Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K880402); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Sep 20, 2024Class IIOpen, Classified
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
Z-0238-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 17, 2023Separation Problem; Separation Problem
A FACILITY REPORTED A SURGICAL PATTY (ID 801399) WAS PLACED IN SITU ON THE BRAIN. HOWEVER, WHEN SURGEON WENT TO REMOVE PATTY BY PULLING THE STRING, THE STRING CAME AWAY AND LEFT THE PATTY IN SITU. THE PATTY WAS REMOVED FROM BRAIN AFTER IT WAS NOTICED TO BE STILL IN SITU AFTER PULLING THE STRING OFF. WHEN THE PATTY WAS PLACED ON THE BRAIN IT WAS NOT MOISTENED AS IT WAS ALREADY IN SITU TO ABSORB BLOOD. THE PATTIES WERE MOISTENED WITH RINGERS SOLUTION WITHIN 5 MINUTES PRIOR TO USE. IT WAS NOT ALTERED OR CUT FROM THE ORIGINAL SIZE PRIOR TO USE, AND NO DELAY OF THE PROCEDURE WAS REPORTED. SURGICAL PROCEDURE: RETRO SIGMOID REPAIR OF RIGHT MASTOID ENCEPHALOCELE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16946524 ↗
FDA clearance and classification
Cleared through 510(k) K880402 on Mar 2, 1988 (decision: Substantially Equivalent). Submitted by Codman & Shurtleff, Inc. as “Codman Surgical Pattie”. The FDA review took 33 days. FDA record ↗
The FDA regulates product code HBA (neurosurgical paddie) under 21 CFR 882.4700 as a Class II device. 21 CFR 882.4700 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K880402. One submission often covers many models, so check the product and lot details in each record.
- Z-0093-2011 · Jul 16, 2010 · TerminatedCodman 5/16" Round Surgical Patties (20 pouches of 10 patties per box) Product code: 80-1396
- Z-0095-2011 · Jul 16, 2010 · TerminatedSurgical Pattie, 1/4" x 1/4" Non--Sterile (Each Unit of sales is 100 cards) of patties on each card) Product code: 24-5403
- Z-0094-2011 · Jul 16, 2010 · TerminatedCodman 1/4 x 1/4" Surgical Patties (20 pouches of 10 patties per box) Product code: 80-1399
Common questions about the CODMAN 80-1399
Has the CODMAN 80-1399 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0238-2025 (Class II), began on Sep 20, 2024: Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the CODMAN 80-1399?
The FDA's MAUDE database lists 1 adverse event report for this device, received on May 17, 2023. That report lists device problem “Separation Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CODMAN 80-1399?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CODMAN 80-1399 FDA approved?
It was cleared through 510(k) K880402 on Mar 2, 1988. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the CODMAN 80-1399, and is it still sold?
The FDA's device database lists Cerenovus, Inc. as the company responsible for it (brand CODMAN, model 80-1399). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other NEUROSURGICAL PADDIE devices
- CODMAN® 801400Integra Lifesciences Production Corporation · Neurosurgical paddie2 recalls9 reports
- CODMAN® 801407Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall6 reports
- CODMAN® 801451Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall5 reports
- MEDLINE NEUROSPNG05Medline Industries, Inc. · Neurosurgical paddie1 recall5 reports
- CODMAN® 801404Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall4 reports
- CODMAN® 801403Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall2 reports
- CODMAN® 801406Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall2 reports
- CODMAN 80-1407Cerenovus, Inc. · NEUROSURGICAL PADDIE1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page