Triage 97000HSRecalls & reported problems
Quidel Cardiovascular Inc. · Triage Cardiac Panel · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 30014613337911), received between Feb 28, 2020 and Sep 16, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K030286); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 25, 2023Open, Classified
Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.
Z-1974-2023 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Low Test Results4
- High Test Results3
- Incorrect, Inadequate or Imprecise Result or Readings2
- Low Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Chest Pain3
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Low Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50793170 ↗MalfunctionSep 16, 2026Low Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50793167 ↗MalfunctionSep 16, 2026Low Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50793163 ↗MalfunctionFeb 4, 2026Low Readings
CUSTOMER REPORTED DISCREPANT TNI RESULTS DURING CORRELATION OF TRIAGE CARDIAC WITH VITROS ON 5 SAMPLES. NO LOT NUMBER, PATIENT SYMPTOMS, OR DIAGNOSIS PROVIDED. NO PATIENT DEATH OR AE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24263520 ↗MalfunctionJan 23, 2026High Test Results
CUSTOMER REPORTED DISCREPANT HIGH TNI RESULTS ON TRIAGE CARDIAC VS VITROS FOR 1 PATIENT. PATIENT 1: SYMPTOMS: NONE REPORTED, PATIENT JUST HAD A BABY. DIAGNOSIS: UNKNOWN. TREATMENT: N/A.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24172237 ↗MalfunctionDec 5, 2025High Test Results; High Test Results
CUSTOMER REPORTED DISCREPANT HIGH TNI RESULTS ON TRIAGE CARDIAC VS UNKNOWN LAB METHOD FOR 1 PATIENT. PATIENT SYMPTOMS: CHEST PAIN, CHEST TIGHTNESS FOR 4 DAYS DIAGNOSIS: CHONDROCOSTAL JUNCTION SYNDROM. TREATMENT: PROTONIX, GI COCKTAIL OF MYLANTA, BENADRYL AND LIDOCAINE, TORADOL FOR PAIN. TREATMENT WAS NOT BASED ON TRIAGE RESULTS. PATIENT DISCHARGED AFTER NEGATIVE CONFIRMATORY TESTING FROM EXTERNAL LAB.
Patient problems: Chest Pain
MAUDE report 23732474 ↗MalfunctionOct 16, 2025Low Test Results; Low Test Results
CUSTOMER REPORTED DISCREPANT LOW TNI RESULTS FOR ONE PATIENT ON TRIAGE VS. ATELLICA HS-TNI METHOD. PATIENT SYMPTOMS: MILD CHEST DISCOMFORT, ACHING SENSATION RADIATING TO BOTH ARMS PATIENT DIAGNOSIS AND TREATMENT: PATIENT WAS INITIALLY DISCHARGED, THEN CALLED BACK TO THE ER. HEPARIN (100 UNITS/ML), IV INFUSION (1000 UNITS/HR) (NITROGLYCERIN 0.4 MG, HEPARIN INJECTION 5000 UNITS). PATIENT WAS TRANSFERRED TO THE HOSPITAL.
Patient problems: Chest Pain
MAUDE report 23317929 ↗MalfunctionNov 6, 2024High Test Results; High Test Results
CUSTOMER REPORTED TRIAGE WAS POSITIVE AND LABORATORY METHOD WAS NEGATIVE FOR ONE PATIENT. CUSTOMER FOUND THAT SEVERAL SAMPLES COLLECTED THAT DAY WERE NOT FULLY FILLED IN THE 4.0 TUBES. UNKNOWN WHETHER THE SAMPLE FOR THIS PATIENT WAS AFFECTED. NO PATIENT DATA AVAILABLE, PATIENT HAD GENERAL CHEST PAINS. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Chest Pain
MAUDE report 20616203 ↗MalfunctionFeb 28, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED THE FOLLOWING OCCURRING FOR ONE PATIENT. ON (B)(6) 2020, A MALE PATIENT PRESENTED WITH LIGHT HEADEDNESS, NAUSEA AND RINGING EARS. PATIENT WAS GIVEN ZOFRAN AND FLUIDS. PATIENT WAS TESTED ON THE TRIAGE PLATFORM FOUR TIMES. 17:16: TRIAGE METER SERIAL NUMBER (B)(6) : CKMB: 3.5; MYO: 147; TNI: 0.17. 17:21: TRIAGE METER SERIAL NUMBER (B)(6) : CKMB: 4.1; MYO: 144; TNI: <0.05. 17:25: TRIAGE METER SERIAL NUMBER (B)(6) : CKMB: 3.7; MYO: 141; TNI: <0.05. 17:34: TRIAGE METER SERIAL NUMBER (B)(6) : CKMB: 3.8; MYO: 142; TNI: <0.05 . FACILITY TRIAGE CUT-OFF FOR TNI: 0.05. PATIENT WAS DISCHARGED ON (B)(6) 2020 WITH NO MEDS. CUSTOMER PROVIDED AN UPDATE ON 3/2/2020. PATIENT "FINAL DX OF MYOCARDITIS".
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9772076 ↗MalfunctionFeb 28, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED THE FOLLOWING OCCURRING FOR ONE PATIENT. ON (B)(6) 2020, A FEMALE PATIENT PRESENTED TO THE CUSTOMERS FACILITY WITH PAIN IN CHEST. PATIENT WAS GIVEN NITRO PATCH, ASPIRIN AND ATIVAN. PATIENT WAS TESTED ON THE TRIAGE PLATFORM THREE TIMES. 16:15: METER SERIAL NUMBER (B)(6): CKMB: <1.0; MYO: 37.8; TNI: 0.22. 16:21: METER SERIAL NUMBER (B)(6): CKMB: <1.0; MYO: 48.0; TNI: <0.05. 16:40: METER SERIAL NUMBER (B)(6): CKMB: <1.0; MYO: 48.7; TNI: <0.05. FACILITY TRIAGE CUT-OFF FOR TNI: 0.05 CUSTOMER PROVIDED AN UPDATE ON 3/2/2020. PATIENT "FINAL DX OF MYOCARDITIS."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9771726 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K030286. One submission often covers many models, so check the product and lot details in each record.
- Z-1929-2012 · May 22, 2012 · TerminatedAlere Triage Cardiac Panel PN 97000HS Used for the quantitative determination of creatine kinase MD (CK--MB), myoglobin and troponin I in EDTA anticoagulated whole blood or plasma
- Z-1391-2009 · Apr 27, 2009 · TerminatedTriage Cardiac Panel, Biosite Catalog Number 97000HS, 25 individually pouched devices in a labeled kit box. Fluorescence immunoassay kit to be used with the Triage Meter for the
Other IMMUNOASSAY METHOD, TROPONIN SUBUNIT devices
- i-STAT 03P90-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT474 reports
- Atellica IM TNIH 10997841Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT463 reports
- Access hsTnI B52669Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit2 recalls170 reports
- i-STAT 06P23-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT195 reports
- VITROS 6802301Ortho Clinical Diagnostics · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall139 reports
- Atellica IM TNIH 10997840Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT166 reports
- Dimension EXL LOCI Troponin I High Sensitive 10471068Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall76 reports
- Access AccuTnI+3 A98143Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit1 death106 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.