Skip to content

Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS… recallZ-1974-2023 · Quidel Cardiovascular Inc · initiated May 25, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Quidel Cardiovascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 5/25/2023, correction notices were sent to customers and distributors who were advised the following: 1) If you have an alternate method, please discard all unused material. 2) If you do not have an alternate method, please follow these steps, as applicable, to minimize patient risk. a) Flag all negative results reported to clinicians as possibly inaccurate until lots of unaffected product are used. b) Use results from an alternate clinical laboratory analyzer when troponin results are below or close to the cutoff and myocardial infarction is suspected. c) Perform serial sampling. Keep patients until at least 3 negative troponin values have been obtained. d) Use all Triage troponin results in conjunction with the patient s risk factors, clinical presentation, EKG, and other imaging. e) Consider recommendations by the ACC, ESC guidelines and the Fourth Universal Definition of Myocardial Infarction for monitoring a patient for a rise or fall pattern of troponin. 3) If you are experiencing issues with Proficiency testing, contact your local Technical Solutions Center. 4) Complete and return the confirmation of receipt form. 5) forward this notification if the product was distributed outside of your facility. If you have further questions, please contact: - For North America, Canada, Asia-Pacific, and Latin America, please call 858.552.1100 - For Europe, Middle East, and Africa, please call +353 (91) 412 474 - For China, please call 0400 920 9366 or +86 021 3217 8300 - EU authorized representative: MDSS GmbH, Tel.: +49-511-62628630, [email protected] On 7/12/23, Update/Reminder Notices were sent informing customers to: Discontinue use of all affected devices. Use an unaffected lot number, obtain unaffected product, use an alternate method or send patients to an alternate testing site utilizing an alternate method. If an alternate method is not available, see above.

Recall details

Recall number
Z-1974-2023
Status
Open, Classified
Initiated by the company
May 25, 2023
Posted by the FDA
Jul 7, 2023
Recalling firm
Quidel Cardiovascular Inc
Quantity
16,194
Product code
MMI
FDA submissions
K030286
Distribution
US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD. OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL

Product description

Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

REF/UDI-DI/Lots(Expiration): 97000HS/30014613337911/T13666N(2023-06-26), T13667N(2023-07-03), T13669N(2023-07-08), T13706N(2023-07-21), T13826N(2023-09-03), T13944N(2023-10-14), T13948N(2023-10-20), T13949N(2023-10-22), T14019N(2023-11-05), T14020N(2023-11-05), T14023N(2023-11-11); 97021HS/30014613337935/T13665RN(2023-06-25), T13707RN(2023-06-28), T13709RN(2023-08-11), T13827RN(2023-09-04), T13829RN(2023-09-18), T13942RN(2023-10-09), T13950RN(2023-10-23); 97022HS/30014613337942/T13664RN(2023-06-20), T13708RN(2023-08-03), T13943RN(2023-10-10); 97000HSEU/30014613337928/T13668RBN(2023-07-05), T13705RN(2023-07-14), T13765RBN(2023-08-26), T13825RBN(2023-08-27), T13828RBN(2023-09-18), T13946RBN(2023-10-16); 97000HSEUJP/T13668RN(2023-07-05), T13765RN(2023-08-26), T13828RN(2023-09-18), T13946RN(2023-10-16); 97000HZ01/T13763(2023-08-19), T13764(2023-08-21), T13830(2023-09-24), T13831(2023-09-26), T13940(2023-10-01), T13941(2023-10-03), T13947(2023-10-21), T13951(2023-10-27), T14021(2023-11-06), T14022(2023-11-10), T14024(2023-11-15), T14025(2023-11-19), T14041(2023-12-13), T14042(2023-12-18), T14043(2023-12-19), T14044(2023-12-20), T14045(2023-12-24); 97000QIL/T13825RNQ(2023-08-27). 97000HS/97021HS/97022HS:Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24). 97000HS/97000HSEU/97000QIL/97021HS: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN, T13831NQ, T13831RBNQ, T13831RNQ(2023-09-26); T14021N, T14021RBN, T14021RN, T14021NQ, T14021RBNQ, T14021RNQ(2023-11-06); T14022N, T14022RBN, T14022RN, T14022NQ, T14022RBNQ, T14022RNQ(2023-11-10); T14024N, T14024RNB, T14024RN, T14024NQ, T14024RNBQ, T14024RNQ(2023-11-15); T14025N, T14025RBN, T14025RN, T14025NQ, T14025RBNQ, T14025RNQ(2023-11-19); T14041N, T14041RBN, T14041RN, T14041NQ, T14041RBNQ, T14041RNQ(2023-12-13); T14042N, T14042RBN, T14042RN, T14042NQ, T14042RBNQ, T14042RNQ(2023-12-18); T14043N, T14043RBN, T14043RN, T14043NQ, T14043RBNQ, T14043RNQ(2023-12-19); T14044N, T14044RBN, T14044RN, T14044NQ, T14044RBNQ, T14044RNQ(2023-12-20); T14045N, T14045RBN, T14045RN, T14045NQ, T14045RBNQ, T14045RNQ(2023-12-24). 97000HSEU/97000HSEUJP: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24).

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page