Skip to content

BD BBL™ Trypticase™ Soy Broth 221093Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Trypticase™ Soy Broth · Class 1

All models of BD BBL Trypticase Soy Broth (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 30382902210932), received between Aug 1, 2019 and Oct 24, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings4
  • Contamination of Device Ingredient or Reagent2
  • Contamination1
  • Inaccurate Information1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH THAT GRAM-NEGATIVE BACILLI WERE OBSERVED IN A GRAM STAIN OF A CSF CULTURE AFTER 5 DAYS INCUBATION AT 35°C. THE END USER SUSPECTED CONTAMINATION AND PERFORMED A GRAM STAIN ON UNUSED SOY BROTH. GRAM-NEGATIVE BACILLI WERE OBSERVED IN AN UNSPECIFIED NUMBER OF TUBES. SUBCULTURE OF THE BROTH RESULTED IN NO GROWTH. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20530051 ↗
  • MalfunctionOct 12, 2023Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    IT WAS REPORTED THAT WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH, THERE WAS CONTAMINATION OF AN UNSPECIFIED NUMBER OF TUBES. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING CONTAMINATED TUBES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17921493 ↗
  • MalfunctionAug 15, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT BD BBL¿ TRYPTICASE¿ SOY BROTH WAS FOUND CONTAMINATED. NO PATIENT IMPACT WAS REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS TSB 221093__3129898 CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17553715 ↗
  • MalfunctionAug 11, 2023Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    IT WAS REPORTED THAT PRIOR TO USING BD BBL¿ TRYPTICASE¿ SOY BROTH, THERE WAS BIOLOGICAL CONTAMINATION OF 4 TUBES. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTING CONTAMINATION AND EVAPORATION OF PRODUCT 221093 LOT NO. 3096339. ONE LOOKS TO BE FUNGAL, AND THE OTHER CONTAMINATED TUBE LOOKS TO BE BACTERIAL."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17515898 ↗
  • MalfunctionMar 20, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT BD BBL¿ TRYPTICASE¿ SOY BROTH BOTTLES CONTAINED SUSPENDED PARTICLES THAT HAVE SIMILAR APPEARANCE TO FUNGAL INFECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THEY NOTICED THAT ALL THE BOTTLES CONTAIN SUSPENDED PARTICLES. IT IS QUITE DISTURBING FOR THEM BECAUSE THE FUNGAL INFECTION CAN PRODUCE A SIMILAR APPEARANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579926 ↗
  • MalfunctionJun 18, 2021Inaccurate Information; Inaccurate Information

    IT WAS REPORTED THAT WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH INCORRECT LABEL INFORMATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " CUSTOMER PROBLEM: CUSTOMER REPORTING LOT NUMBER ON TUBE LABELS DOESN¿T MATCH LOT NUMBER ON PACKAGING LABELS FOR PRODUCT 221093".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12025739 ↗
  • MalfunctionMar 27, 2021Contamination; Contamination

    IT WAS REPORTED THAT WHILE USING 3 BD BBL¿ TRYPTICASE¿ SOY BROTH FOREIGN MATTER WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " CUSTOMER REPORTED THAT THERE WERE SEDIMENTS AT THE BOTTOM OF THE TUBES 221093 LOT 0364881.3 BOXES WERE INSPECTED FROM AN ORDER OF 8 BOXES (100 TUBES EACH). CUSTOMER STATED THAT A MAJORITY OF THE TUBES WERE AFFECTED. THE TECH GRAM-STAINED THE TUBES AND GRAM POSITIVE ROD WAS FOUND. "

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11579719 ↗
  • MalfunctionAug 1, 2019Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    IT WAS REPORTED THAT WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH THE CUSTOMER ALLEGED THAT THE MEDIA WAS DEFECTIVE AND WAS CAUSING NON-SPECIFIC ISSUES WITH CULTURING AND PROCESSING TISSUES. THE CUSTOMER INDICATED THE USE OF THE MEDIA RESULTED IN DISCREPANT RESULTS AND LOSS OF DONOR TISSUE. NO INFORMATION WAS AVAILABLE ON HOW THE PATIENT TREATMENT WAS CHANGED OR IF TREATMENT WAS CHANGED. NO MISDIAGNOSIS OR WORSENING IN PATIENT CONDITION WAS INDICATED. THREE ATTEMPTS WERE MADE TO FOLLOW-UP FOR ADDITIONAL INFORMATION WITHOUT A RESPONSE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8851314 ↗

Other CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL devices

All CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL devices →