Skip to content

BD BBL™ Taxo™ A 231552Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Taxo™ A Differentiation Test for Group A Streptococci BACITRACIN 0.04 IU/IE/UI · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 30382902315521), received between Apr 6, 2022 and Nov 22, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING AN UNSPECIFIED NUMBER OF BD BBL¿ TAXO¿ A, DISCS, THERE IS NO ZONE OF INHIBITION WHEN TESTING WITH STREP PYOGENES SAMPLES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20763423 ↗
  • MalfunctionApr 6, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT WHILE TESTING WITH THE BD BBL¿ TAXO¿ A DISCS FOR DIFFERENTIATION OF GROUP A STREPTOCOCCI, THERE WAS NO ZONE OF INHIBITION. CONFIRMATORY TESTING WAS PERFORMED WITH A REMEL PATHODX AGGLUTINATION TEST, AND THE RESULT WAS POSITIVE FOR STREP GROUP A. RESULT WAS REPORTED AS STREP GROUP A, AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER PROBLEM: CUSTOMER IS REPORTING NO ZONE OF INHIBITION WITH 231552 DISC IF MEDIA PERFORMANCE ISSUE PROVIDE ORGANISM ATCC STRAIN INFORMATION (IF APPLICABLE): N/A STEPS TAKEN WITH CUSTOMER/TROUBLESHOOTING: CUSTOMER IS REPORTING THEY ARE NOT GETTING A ZONE OF INHIBITION WITH 231552 DISC FROM LOT 0343425. THE CUSTOMER STATES THEY DID A WET PREP ON THE ORGANISM AND IT WAS COCCI IN CHAINS AND THERE WAS CLEAR HEMOLYSIS ON THE BLOOD AGAR PLATE. THE SUBBED TO BLOOD AGAR PLATE AND DROPPED A 231552 DISC AND ALLOWED THE PLATE TO GROW IN A 36 DEGREE INCUBATOR OVERNIGHT. THE DISC HAD NO ZONE OF INHIBITION AROUND IT. THE CUSTOMER DECIDED TO RUN A REMEL PATHODX AGGLUTINATION TEST AND IT CAME BACK POSITIVE FOR STREP GROUP A. QC FOR THE DISCS IS RUN WEEKLY AND THE CUSTOMER REPORTS THEY HA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14025913 ↗

Other DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION devices

All DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page