Skip to content

BD Micro-Fine 329420Recalls & reported problems

Becton, Dickinson and Company Inc · BD Insulin Syringes with the BD Micro-Fine IV Needle, U-100, 12.7mm, 28G, 1mL · Class 2

All models of BD Micro-Fine (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 30382903294207), received between Feb 5, 2018 and Sep 9, 2026 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 1, 2023Class IIOpen, Classified

    Nonsterile syringe potentially exposing the patient to pathogenic microorganisms and a risk of local infection.

    Z-1238-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break7
  • Leak/Splash2
  • Material Puncture/Hole2
  • Defective Device1
  • Mechanical Problem1
  • Physical Resistance/Sticking1
  • Positioning Problem1
  • Use of Device Problem1

Patient problems

  • No Known Impact Or Consequence To Patient5
  • Needle Stick/Puncture2
  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2026Material Puncture/Hole

    Patient problems: Needle Stick/Puncture

    MAUDE report 50697351 ↗
  • MalfunctionNov 5, 2024Material Puncture/Hole; Material Puncture/Hole

    MATERIAL # 329420 MATERIAL DESCRIPTION - SYRINGE 1.0ML 28GA 1/2IN BLS 500 SUS CA MATERIAL LOT# - 3107277 COMPLAINT DESCRIPTION ¿ AS PER ACCOUNT, THE CUSTOMER HAD 2 INCIDENTS OF NEEDLES POKING OUT. ONE OF THEIR NURSES GOT POKED BUT IS THANKFULLY OKAY. NOTES ¿ IMAGES ATTACHED FOR REFERENCE PROVIDED BY CUSTOMER.

    Patient problems: Needle Stick/Puncture

    MAUDE report 20614578 ↗
  • MalfunctionApr 28, 2021Break; Leak/Splash; Break; Leak/Splash

    IT WAS REPORTED THAT A SYRINGE 1.0ML 28GA 1/2IN BLS 500 SUS CA WAS FOUND TO BE DAMAGED DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT PLUNGER ROD IS DIFFICULT TO MOVE AND VEINS ARE BLOWING UP. THIS CAUSES PAIN TO THE PATIENT. VERBATIM: FROM PHONE CALL ON (B)(6) 2021 16:57:37: CONSUMER STATED, PLUNGER IS HARD TO PUSH ON WHEN TAKING INJECTIONS. STATED, IT GETS "STUCK", CAUSING HER TO PUSH HARDER AND AS RESULT, SHE EXPERIENCES PAIN IN HER "VEINS".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11737018 ↗
  • MalfunctionApr 28, 2021Mechanical Problem; Use of Device Problem; Positioning Problem; Physical Resistance/Sticking

    MATERIAL NO. 329420 BATCH NO. 0262090. IT WAS REPORTED THAT PLUNGER ROD IS DIFFICULT TO MOVE AND VEINS ARE BLOWING UP. THIS CAUSES PAIN TO THE PATIENT. VERBATIM: FROM PHONE CALL ON 2021-04-06 16:57:37: CONSUMER STATED, PLUNGER IS HARD TO PUSH ON WHEN TAKING INJECTIONS. STATED, IT GETS "STUCK", CAUSING HER TO PUSH HARDER AND AS RESULT, SHE EXPERIENCES PAIN IN HER "VEINS".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11735994 ↗
  • MalfunctionMay 7, 2020Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT SYRINGE 1.0ML 28GA 1/2IN BLS 500 SUS CA PLUNGER ROD IS DIFFICULT TO MOVE. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 329420 BATCH NO. UNKNOWN. IT WAS REPORTED THAT SYRINGES STICK AND ARE VERY HARD TO PRESS, LEAVING VEINS ALL BLOWN UP. ALSO REPORTED BUMPS BECAUSE SYRINGE IS STICKING IN SKIN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10033237 ↗
  • InjuryApr 14, 2020Defective Device; Defective Device

    IT WAS REPORTED THAT SYRINGE 1.0ML 28GA 1/2IN BLS 500 SUS CA NEEDLE SEPARATED FROM THE HUB AND PATIENT WENT TO THE ER AND RECEIVED X-RAYS TO DETERMINE IF THE NEEDLE REMAINED IN HIS ARM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT AFTER THE INJECTION THE PATIENT AND PERSON ADMINISTERING NOTED THAT THE NEEDLE WAS MISSING. AFTER GOING TO THE ER AND HAVING AN X-RAY PERFORMED, IT WAS CONFIRMED THAT NO NEEDLE WAS PRESENT IN THE PATIENT'S ARM. THE SYRINGE WAS X-RAYED AND IT INDICATED THAT THE NEEDLE WAS IN THE BARREL OF THE SYRINGE. "

    Patient problems: No Code Available

    MAUDE report 9960122 ↗
  • MalfunctionAug 23, 2019Break; Break

    IT WAS REPORTED THAT A NEEDLE BREAK OCCURRED DURING USE WITH A SYRINGE 1.0 ML 28GA 1/2 IN BLS 500 SUS CA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "CUSTOMER IS STATING THEY NO LONGER CAN USE AND ALSO THAT THE NEEDLES ARE BREAKING. I DON'T KNOW IF THEY REPORTED THE ISSUE BUT THEY WON'T USE AND WE DON'T HAVE ANY OTHER CUSTOMERS THAT USE THIS PRODUCT. 300 CASES. 105 CASES. ITEM #329420 LOT AND EXPIRATION: LOT #8337907 EXPIRATION IS 2-29-2024, LOT #7079902 EXPIRATION IS 4-30-2022, LOT #8183955 EXPIRATION IS 9-30-2023, LOT #8169595 EXPIRATION IS 9-30-2023."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8925456 ↗
  • MalfunctionAug 22, 2019Break; Break

    MATERIAL NO. 329420, BATCH NO. 8337907,7079902,8183955,8169595. IT WAS REPORTED THAT DURING USE OF THE BD MICRO-FINE¿ IV INSULIN SYRINGE THE NEEDLES ARE BREAKING. THE DISTRIBUTOR PROVIDED FOUR DIFFERENT LOT NUMBERS: 8337907, 7079902, 8183955, AND 8169595. CUSTOMER PROVIDED THREE DATES: THIS COMPLAINT CAPTURES (B)(6) 2019. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS STATING THEY NO LONGER CAN USE AND ALSO THAT THE NEEDLES ARE BREAKING. I DON'T KNOW IF THEY REPORTED THE ISSUE BUT THEY WON'T USE AND WE DON'T HAVE ANY OTHER CUSTOMERS THAT USE THIS PRODUCT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8921800 ↗
  • MalfunctionAug 5, 2019Break; Break

    MATERIAL NO. 329420 BATCH NO. UNKNOWN (PROVIDED 8337907,7079902,8183955,8169595). IT WAS REPORTED THAT DURING USE OF THE BD MICRO-FINE¿ IV INSULIN SYRINGE THE NEEDLES ARE BREAKING. THE DISTRIBUTOR PROVIDED FOUR DIFFERENT LOT NUMBERS, BUT THE ACTUAL LOT NUMBER IS UNKNOWN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS STATING THEY NO LONGER CAN USE AND ALSO THAT THE NEEDLES ARE BREAKING. I DON'T KNOW IF THEY REPORTED THE ISSUE BUT THEY WON'T USE AND WE DON'T HAVE ANY OTHER CUSTOMERS THAT USE THIS PRODUCT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8859268 ↗
  • MalfunctionMar 7, 2019Break; Break

    IT WAS REPORTED THAT DURING USE OF THE BD MICRO-FINE¿ IV INSULIN SYRINGE THAT THE NEEDLES BREAK OR "CROOKS" WHEN INJECTING INTO THE SKIN OR WHEN RECAPPING THE NEEDLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8401471 ↗
  • MalfunctionDec 3, 2018Break; Break

    IT WAS REPORTED WITH THE USE OF THE BD MICRO-FINE¿ IV INSULIN SYRINGE THERE WAS AN ISSUE WITH NEEDLES BREAKING DURING INJECTION OR WHEN RECAPPING.

    Patient problems: No Information

    MAUDE report 8126730 ↗
  • MalfunctionFeb 5, 2018Break

    IT WAS REPORTED WITH THE USE OF THE BD MICRO-FINE¿ IV INSULIN SYRINGE THE NEEDLE TIPS WERE BREAKING. NO REPORT OF ADDITIONAL INFORMATION. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7242465 ↗

FDA clearance and classification

The FDA regulates product code FMF (syringe, piston) under 21 CFR 880.5860 as a Class II device. 21 CFR 880.5860 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the BD Micro-Fine 329420

Has the BD Micro-Fine 329420 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1238-2023 (Class II), began on Feb 1, 2023: Nonsterile syringe potentially exposing the patient to pathogenic microorganisms and a risk of local infection. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the BD Micro-Fine 329420?

The FDA's MAUDE database lists 12 adverse event reports for this device, received between Feb 5, 2018 and Sep 9, 2026. In those reports, the most common specific device problems are “Break” (7), “Leak/Splash” (2) and “Material Puncture/Hole” (2); the most common patient problem is “Needle Stick/Puncture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BD Micro-Fine 329420?

None of the reports are classified as deaths, 1 as an injury and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the BD Micro-Fine 329420, and is it still sold?

The FDA's device database lists Becton, Dickinson and Company Inc as the company responsible for it (brand BD Micro-Fine, model 329420). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Syringe, piston devices

All Syringe, piston devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page