BD Microtainer® Tubes with LH (Lithium Heparin) 365965Recalls & reported problems
Becton, Dickinson And Company · BD Microtainer® Tubes with LH (Lithium Heparin) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 30382903659655), received between May 1, 2018 and Oct 30, 2024 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (8 reports).
1 FDA recall record name this device's identifier.
8 other recalls cover devices cleared under the same FDA submission (K991702); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 13, 2019Class IITerminated
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
Z-0055-2020 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings8
- Incorrect Or Inadequate Test Results3
- Coagulation in Device or Device Ingredient2
- Leak/Splash2
- Component Missing1
- Device Contamination with Chemical or Other Material1
- Short Fill1
Patient problems
- No Known Impact Or Consequence To Patient10
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
- No Information1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2024Leak/Splash; Device Contamination with Chemical or Other Material; Leak/Splash; Device Contamination with Chemical or Other Material
IT WAS REPORTED PRIOR TO USE OF BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN), 200 MICROTAINERS HAD FOREIGN MATTER(FM) AND MELTED/DEFORMED CAPS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20567764 ↗MalfunctionSep 11, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) 15 SAMPLES WERE CLOTTED. THE CUSTOMER INDICATED TUBES WERE UNDERFILLED AND OVERFILLED. THE CUSTOMER INDICATED THE CLOTTED SAMPLES WERE DUE TO IMPROPER COLLECTION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20197058 ↗MalfunctionJun 28, 2024Incorrect, Inadequate or Imprecise Result or Readings; Component Missing; Incorrect, Inadequate or Imprecise Result or Readings; Component Missing
IT WAS REPORTED WHILE USING BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) 10 TUBES ARE CLOTTING AND ARE MISSING ADDITIVE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19639403 ↗MalfunctionJun 22, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING USE SAMPLES ARE HEMOLYZED WITH A BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: REPORTED EXPERIENCING HEMOLYZED SPECIMENS WHEN USING LOTS 925441N AND 927410N. THIS STARTED OCCURRING WITHIN THE LAST 3 MONTHS, NO SPECIFIC AMOUNT OF SPECIMENS AFFECTED. CUSTOMER STATES THERE ARE NO REPORTS OF ERRONEOUS RESULTS AS THEY ARE THE DRAWING LAB. NO REPORTS FROM TESTING LAB OF ERRONEOUS RESULTS. SHE WILL BE CONDUCTING DRAWS TODAY SO PHOTOS MAY BE AVAILABLE AFTER TODAY. THOUGH SHE DID CONFIRM UNUSED TUBES ARE AVAILABLE FOR INVESTIGATION NOW.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10179775 ↗MalfunctionJun 17, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT 6 BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) WERE BELIEVED TO BE SODIUM LITHIUM TUBES AND USED FOR CYTOGENETIC TESTING, WHERE THEY WERE REJECTED BY THE CLINIC FOR "LITHIUM US CYTOTOXIC". THE ISSUE WAS DISCOVERED AFTER USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: STATED THE 6 GREEN TOP MICROTAINERS HAD BEEN DRAWN AT THE NEXT-DOOR CHILDREN'S HOSPITAL AND SENT TO THE CLINIC FOR GENETIC TESTING, ONLY TO BE TOLD THEY WERE UNACCEPTABLE B/C LITHIUM US CYTOTOXIC. I TOLD HIM SODIUM HEPARIN MICROTAINERS ARE NOT AVAILABLE AT THIS TIME, AND WE HAVE NO DATA SUPPORTING THE USE OF LITHIUM HEPARIN FOR CYTOGENETIC TESTING. I TOLD HIM A 2.0 ML VACUTAINER IS AVAILABLE WITH SODIUM HEPARIN, AND MAY BE USED WHEN VENOUS BLOOD IS DRAWN FROM NEONATES, TYPICALLY FROM CENTRAL LINES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10164857 ↗MalfunctionNov 11, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
MATERIAL NO. 365965 BATCH NO. 910158N, 912391N IT WAS REPORTED THAT DURING USE OF THE BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) THERE IS DECREASED CO2 RESULTS WHEN USING BD MICROTAINER TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER COMPLAINED OF HAVING DECREASED CO2 RESULTS USING BD MICROTAINER GREEN TOP TUBES. THEY ARE STILL INVESTIGATING THE CAUSE AND HAVE CONFIRMED THE ISSUE IS NOT FROM THEIR REAGENTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9306087 ↗MalfunctionJul 11, 2019Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient
IT WAS REPORTED THAT THE SAMPLE COLLECTED FOR BILIRUBIN IN THE BD MICROTAINER® TUBE WITH LH (LITHIUM HEPARIN) CLOTTED DURING USE, BUT WAS RUN THROUGH THE ANALYZER REGARDLESS "WITH NO ISSUES". THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SPOKE WITH CUSTOMER AND THEY HAD 2-3 SAMPLES THAT WERE CLOTTED IN THE MICROTAINER TUBES. THEY WERE COLLECTED FOR BILIRUBIN. HE FEELS IT IS A COLLECTION ISSUE BY THE NURSES, THAT THEY ARE NOT MIXING THEM CORRECTLY. SAMPLES WERE NOT HEMOLYSED. THEY DID RUN SAMPLES WITH NO ISSUES AND HAVE VALIDATED THE BILIRUBIN TEST ON SERUM SAMPLES."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8781572 ↗MalfunctionMay 23, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
MATERIAL NO. 365965. BATCH NO. UNKNOWN. IT WAS REPORTED THAT AFTER USE OF THE BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) THERE WERE ISSUES WITH HEMOLYSIS. THERE WERE THREE OCCURRENCES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE WAS 3 INCIDENTS THAT INVOLVED HEMOLYSIS ON (B)(6) WITH PRODUCT 365965/UNKNOWN LOT. NO PATIENT IDENTIFIERS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8638434 ↗MalfunctionApr 30, 2019Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient
IT WAS REPORTED THAT SAMPLE COLLECTED IN THE BD MICROTAINER® TUBE WITH LH (LITHIUM HEPARIN) CLOTTED. THIS COMPLAINT WAS CREATED TO CAPTURE 1 OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE TUBES WERE CLOTTING."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8565546 ↗MalfunctionApr 30, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) WERE CLOTTING. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 365965, BATCH NO. 830556N. IT WAS REPORTED THAT THE TUBES WERE CLOTTING. THE TUBES WERE CLOTTING. DATE OF EVENT: (B)(6) 2019 (2 OF 2). DATES OF EVENT: (B)(6) 2019: NO PATIENT IDENTIFIERS, UNKNOWN LOT, NO SAMPLES. (B)(6) 2019: FEMALE, 18 YRS OLD, LOT 830556N, SAMPLES AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8565432 ↗MalfunctionJan 11, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT A BD MICROTAINER® TUBES WITH LH DID NOT SEPARATE PROPERLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8242138 ↗MalfunctionJun 14, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
IT WAS REPORTED THAT BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) HAD HEMOLYSIS. NO SERIOUS INJURY OR MEDICAL INTERVENTION. NO REPORT OF BLOOD EXPOSURE TO MUCOUS MEMBRANE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7603559 ↗MalfunctionJun 14, 2018Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) HAD CRACKED CAPS. NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7601242 ↗MalfunctionJun 12, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
IT WAS REPORTED THAT BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) PROVIDED ERRONEOUS RESULTS. NO REPORT OF SERIOUS INJURY OR MEDICAL INTERVENTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7594389 ↗MalfunctionMay 2, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
IT WAS REPORTED THAT BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) CONTAINED CLOTS WHICH COULD CAUSE ERRONEOUS RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7481589 ↗MalfunctionMay 1, 2018Short Fill; Short Fill
IT WAS REPORTED THAT BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN) WERE UNDER FILLING.
Patient problems: No Information
MAUDE report 7475850 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K991702. One submission often covers many models, so check the product and lot details in each record.
- Z-0059-2020 · May 13, 2019 · TerminatedBD Microtainer PSTTubes with LH (Lithium Heparin) Part/Catalog Number:365985
- Z-0058-2020 · May 13, 2019 · TerminatedBD Microtainer SST- Amber Part/Catalog Number: 365978
- Z-0061-2020 · May 13, 2019 · TerminatedBD Microtainer Tubes with FE (Sodium Fluoride / Disodium EDTA) Glycolytic Inhibitor Part/Catalog Number: 365992
- Z-0054-2020 · May 13, 2019 · TerminatedBD Microtainer Z (No Additive Tubes) Part/Catalog No.365963
- Z-0057-2020 · May 13, 2019 · TerminatedBD MicrotainerTubes with K2E (K2EDTA) Catalog Number: 365974
- Z-0056-2020 · May 13, 2019 · TerminatedBD Microtainer SST Part/Catalog No.365967
- Z-0060-2020 · May 13, 2019 · TerminatedBD Microtainer PST Tubewith LH (Lithium Heparin) - Amber Part/Catalog Number: 365987
- Z-1603-05 · Aug 20, 2004 · TerminatedBD Vacutainer Microtainer Serum tubes (with or without gel).
Other Tubes, vials, systems, serum separators, blood collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page