Microtainer® K2E Blood Collection Tubes 365975Recalls & reported problems
Becton, Dickinson And Company · TUBE MICRO W/MICROGARD EDTA LAV · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 30382903659754), received between Sep 10, 2024 and Jan 15, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Ingredient or Reagent Problem” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K940905); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Ingredient or Reagent Problem4
- Coagulation in Device or Device Ingredient1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2026Device Ingredient or Reagent Problem
REPORT 1 OF 4, IT WAS REPORTED WHILE USING MICROTAINER® K2E BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED MORPHOLOGY IN RED CELLS PRESENTING AS BURR CELLS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24087026 ↗MalfunctionJan 15, 2026Device Ingredient or Reagent Problem
REPORT 4 OF 4: IT WAS REPORTED WHILE USING MICROTAINER® K2E BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED MORPHOLOGY IN RED CELLS PRESENTING AS BURR CELLS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24087024 ↗MalfunctionJan 15, 2026Device Ingredient or Reagent Problem
REPORT 3 OF 4: IT WAS REPORTED WHILE USING MICROTAINER® K2E BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED MORPHOLOGY IN RED CELLS PRESENTING AS BURR CELLS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24087023 ↗MalfunctionJan 15, 2026Device Ingredient or Reagent Problem
REPORT 2 OF 4, IT WAS REPORTED WHILE USING MICROTAINER® K2E BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED MORPHOLOGY IN RED CELLS PRESENTING AS BURR CELLS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24087021 ↗MalfunctionSep 10, 2024Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient
IT WAS REPORTED WHILE ANALYZING BLOOD SAMPLES COLLECTED IN BD MICROTAINER® K2E TUBES, THE BLOOD SAMPLES COAGULATED. THIS OCCURRED TEN (10) TIMES. THE BLOOD SAMPLES WERE RECOLLECTED. THERE WAS NO REPORT OF ADVERSE IMPACT TO PATIENTS OR USERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20180396 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K940905. One submission often covers many models, so check the product and lot details in each record.
Other Tubes, vials, systems, serum separators, blood collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page