BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450Recalls & reported problems
Becton, Dickinson And Company · TUBE EDTA GC 13X100 7.0 PLBL LAV · Class 2
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30382903664505), received between May 1, 2018 and Oct 29, 2019 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Break1
- Device Contamination with Chemical or Other Material1
- Incorrect Or Inadequate Test Results1
- Short Fill1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Information2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2019Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) BLOOD COLLECTION TUBE DID NOT MEET THE MEAN FORCE SPECIFICATION FOR STOPPER REMOVAL/LOOSE CLOSURE BEFORE USE DURING R&D TESTING. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING R&D TESTING, WE OBSERVED SOME TUBES WITH 2ND TIME STOPPER REMOVAL FORCES THAT DID NOT MEET THE MINIMUM FORCE SPECIFICATION. WE ALSO OBSERVED 1 BATCH OF TUBES THAT DID NOT MEET THE MEAN FORCE SPECIFICATION. THE PRODUCT INFORMATION IS LISTED BELOW. CATALOG #366430, BATCH #9095718, TEST DATE: (B)(6) 2019, DATA REVIEW DATE: 14 OCT 2019. OBSERVATION DID NOT MEET MINIMUM FORCE SPECIFICATION. 366668, 9122779, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MINIMUM FORCE SPECIFICATION. 367729, 9184898, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MINIMUM FORCE SPECIFICATION. 366450, 9095722, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MEAN FORCE SPECIFICATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9250805 ↗MalfunctionFeb 13, 2019Break; Break
IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) BLOOD COLLECTION TUBE BROKE "IN STORAGE AFTER FREEZING".
Patient problems: No Information
MAUDE report 8335920 ↗MalfunctionJul 18, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES HAD DARK CLUMPS IN THE TUBE WHICH ARE POSSIBLY FOREIGN MATERIAL. NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO REPORT OF BLOOD EXPOSURE TO MUCOUS MEMBRANE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7698915 ↗MalfunctionMay 9, 2018Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE BD VACUTAINER® VENOUS BLOOD COLLECTION TUBES ARE BREAKING DURING THE CENTRIFUGE PROCESS. NO EXPOSURE TO BODILY FLUID. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7500922 ↗MalfunctionMay 9, 2018Short Fill; Short Fill
IT WAS REPORTED THAT BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES WERE UNDERFILLED.
Patient problems: No Information
MAUDE report 7500568 ↗MalfunctionMay 1, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES HAD CLOTTING ISSUES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7477716 ↗
Other Tubes, vacuum sample, with anticoagulant devices
- Vacuette 454322Greiner Bio-One GmbH · Tubes, vacuum sample, with anticoagulant3 recalls0 reports
- BD MiniDraw™ H&H Capillary Blood Collection Tube 366603Becton, Dickinson And Company · Tubes, vacuum sample, with anticoagulant141 reports
- BD Vacutainer® K2 EDTA (K2E) Plus Blood Collection Tubes 366643Becton, Dickinson U.K. Limited · Tubes, vacuum sample, with anticoagulant124 reports
- Cyto-Chex BCT 218980Streck, Inc. · Tubes, Vacuum Sample, With Anticoagulant1 recall0 reports
- Cyto-Chex BCT 213559Streck, Inc. · Tubes, Vacuum Sample, With Anticoagulant1 recall0 reports
- Vacuette 454334Greiner Bio-One GmbH · Tubes, vacuum sample, with anticoagulant1 recall0 reports
- Vacuette 454332Greiner Bio-One GmbH · Tubes, vacuum sample, with anticoagulant1 recall0 reports
- BD MiniDraw™ Finger Sleeve, Extra-Large 366616Becton, Dickinson And Company · Tubes, vacuum sample, with anticoagulant25 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.