Skip to content

BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450Recalls & reported problems

Becton, Dickinson And Company · TUBE EDTA GC 13X100 7.0 PLBL LAV · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30382903664505), received between May 1, 2018 and Oct 29, 2019 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Leak/Splash2
  • Break1
  • Device Contamination with Chemical or Other Material1
  • Incorrect Or Inadequate Test Results1
  • Short Fill1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2019Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) BLOOD COLLECTION TUBE DID NOT MEET THE MEAN FORCE SPECIFICATION FOR STOPPER REMOVAL/LOOSE CLOSURE BEFORE USE DURING R&D TESTING. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING R&D TESTING, WE OBSERVED SOME TUBES WITH 2ND TIME STOPPER REMOVAL FORCES THAT DID NOT MEET THE MINIMUM FORCE SPECIFICATION. WE ALSO OBSERVED 1 BATCH OF TUBES THAT DID NOT MEET THE MEAN FORCE SPECIFICATION. THE PRODUCT INFORMATION IS LISTED BELOW. CATALOG #366430, BATCH #9095718, TEST DATE: (B)(6) 2019, DATA REVIEW DATE: 14 OCT 2019. OBSERVATION DID NOT MEET MINIMUM FORCE SPECIFICATION. 366668, 9122779, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MINIMUM FORCE SPECIFICATION. 367729, 9184898, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MINIMUM FORCE SPECIFICATION. 366450, 9095722, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MEAN FORCE SPECIFICATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9250805 ↗
  • MalfunctionFeb 13, 2019Break; Break

    IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) BLOOD COLLECTION TUBE BROKE "IN STORAGE AFTER FREEZING".

    Patient problems: No Information

    MAUDE report 8335920 ↗
  • MalfunctionJul 18, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES HAD DARK CLUMPS IN THE TUBE WHICH ARE POSSIBLY FOREIGN MATERIAL. NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO REPORT OF BLOOD EXPOSURE TO MUCOUS MEMBRANE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7698915 ↗
  • MalfunctionMay 9, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD VACUTAINER® VENOUS BLOOD COLLECTION TUBES ARE BREAKING DURING THE CENTRIFUGE PROCESS. NO EXPOSURE TO BODILY FLUID. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7500922 ↗
  • MalfunctionMay 9, 2018Short Fill; Short Fill

    IT WAS REPORTED THAT BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES WERE UNDERFILLED.

    Patient problems: No Information

    MAUDE report 7500568 ↗
  • MalfunctionMay 1, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results

    IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES HAD CLOTTING ISSUES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7477716 ↗

Other Tubes, vacuum sample, with anticoagulant devices

All Tubes, vacuum sample, with anticoagulant devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.