Skip to content

BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450Recalls & reported problems

Becton, Dickinson And Company · TUBE EDTA GC 13X100 7.0 PLBL LAV · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30382903664505), received between May 1, 2018 and Oct 29, 2019 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Leak/Splash2
  • Break1
  • Device Contamination with Chemical or Other Material1
  • Incorrect Or Inadequate Test Results1
  • Short Fill1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2019Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) BLOOD COLLECTION TUBE DID NOT MEET THE MEAN FORCE SPECIFICATION FOR STOPPER REMOVAL/LOOSE CLOSURE BEFORE USE DURING R&D TESTING. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING R&D TESTING, WE OBSERVED SOME TUBES WITH 2ND TIME STOPPER REMOVAL FORCES THAT DID NOT MEET THE MINIMUM FORCE SPECIFICATION. WE ALSO OBSERVED 1 BATCH OF TUBES THAT DID NOT MEET THE MEAN FORCE SPECIFICATION. THE PRODUCT INFORMATION IS LISTED BELOW. CATALOG #366430, BATCH #9095718, TEST DATE: (B)(6) 2019, DATA REVIEW DATE: 14 OCT 2019. OBSERVATION DID NOT MEET MINIMUM FORCE SPECIFICATION. 366668, 9122779, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MINIMUM FORCE SPECIFICATION. 367729, 9184898, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MINIMUM FORCE SPECIFICATION. 366450, 9095722, (B)(6) 2019, 14 OCT 2019. DID NOT MEET MEAN FORCE SPECIFICATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9250805 ↗
  • MalfunctionFeb 13, 2019Break; Break

    IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) BLOOD COLLECTION TUBE BROKE "IN STORAGE AFTER FREEZING".

    Patient problems: No Information

    MAUDE report 8335920 ↗
  • MalfunctionJul 18, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES HAD DARK CLUMPS IN THE TUBE WHICH ARE POSSIBLY FOREIGN MATERIAL. NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO REPORT OF BLOOD EXPOSURE TO MUCOUS MEMBRANE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7698915 ↗
  • MalfunctionMay 9, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD VACUTAINER® VENOUS BLOOD COLLECTION TUBES ARE BREAKING DURING THE CENTRIFUGE PROCESS. NO EXPOSURE TO BODILY FLUID. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7500922 ↗
  • MalfunctionMay 9, 2018Short Fill; Short Fill

    IT WAS REPORTED THAT BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES WERE UNDERFILLED.

    Patient problems: No Information

    MAUDE report 7500568 ↗
  • MalfunctionMay 1, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results

    IT WAS REPORTED THAT THE BD VACUTAINER® K3 EDTA (K3E) 12MG BLOOD COLLECTION TUBES HAD CLOTTING ISSUES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7477716 ↗

FDA clearance and classification

Cleared through 510(k) K213670 on Aug 25, 2023 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “BD Vacutainer K2EDTA Blood Collection Tubes, BD Vacutainer K3EDTA Blood Collection Tubes”. The FDA review took 641 days. FDA record ↗

The FDA regulates product code GIM (tubes, vacuum sample, with anticoagulant) under 21 CFR 862.1675 as a Class II device. 21 CFR 862.1675 ↗

Common questions about the BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450

Has the BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450 been recalled?

No FDA recall record in our data names this exact device identifier (DI 30382903664505), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between May 1, 2018 and Oct 29, 2019. In those reports, the most common specific device problems are “Leak/Splash” (2), “Break” (1) and “Device Contamination with Chemical or Other Material” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450 FDA approved?

It was cleared through 510(k) K213670 on Aug 25, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes 366450, and is it still sold?

The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD Vacutainer® K3 EDTA (K3E) 12 mg Blood Collection Tubes, model 366450). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.