Skip to content

BD Vacutainer® Serum 367820Recalls & reported problems

Becton, Dickinson And Company · Tube PLN PLC 16X100mm 10.0ml PLBL RD · Class 2

All models of BD Vacutainer Serum (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This BD Vacutainer serum blood collection tube is used to collect blood samples for serum tests. Its 148 reports describe tubes that are underfilled, have white flakes or residue inside, leak, or lose vacuum, and some mention inaccurate results. The reports include 2 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
148
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 148 adverse event reports that list this exact device (DI 30382903678205), received between Nov 29, 2017 and May 18, 2026 — classified as 2 injuries, 146 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (28 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 6 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings28
  • Device Contamination with Chemical or Other Material26
  • Short Fill25
  • Leak/Splash17
  • Incorrect Or Inadequate Test Results15
  • Volume Accuracy Problem12
  • Device Ingredient or Reagent Problem10
  • Component Missing7

Patient problems

  • No Clinical Signs, Symptoms or Conditions65
  • No Known Impact Or Consequence To Patient65
  • No Consequences Or Impact To Patient9
  • No Information5
  • No Patient Involvement3
  • Exposure to Body Fluids1
  • Hemolysis1

Recent adverse event reports

The 25 most recent of 148 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2026Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM BLOOD COLLECTION TUBES, FOIUR TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25202360 ↗
  • MalfunctionFeb 6, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, WHITE FLAKES AND RESIDUE ARE SEEN IN AN UNSPECIFIED NUMBER OF TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24279202 ↗
  • MalfunctionFeb 4, 2026Device Ingredient or Reagent Problem

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, WHITE FLAKES AND RESIDUE ARE SEEN IN 1,336 TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24261891 ↗
  • MalfunctionDec 18, 2025Short Fill; Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, ONE CASE OF TUBES ARE EXHIBITING DECREASED VACUUM. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23839546 ↗
  • MalfunctionDec 16, 2025Short Fill; Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, SIX TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23816937 ↗
  • MalfunctionDec 10, 2025Short Fill; Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, ONE TUBE UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23773539 ↗
  • MalfunctionOct 28, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT PRIOR TO USING BD VACUTAINER® SERUM, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED TEMPERATURE EXCURSION DURING TRANSPORT. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23408422 ↗
  • MalfunctionJul 11, 2025Short Fill; Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, 20 TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22485398 ↗
  • MalfunctionJul 7, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, THE ADDITIVE IN 15 TUBES APPEARS TO BE CLUMPING. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22414528 ↗
  • MalfunctionJun 12, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, AN UNSPECIFIED NUMBER OF TUBES ARE NOT CLOTTING AFTER COLLECTION. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22195576 ↗
  • MalfunctionMay 22, 2025Component Missing; Component Missing

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, 15 SAMPLES CLOTTED POST CENTRIFUGATION. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069833 ↗
  • MalfunctionApr 21, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, THE LOT NUMBER AND EXPIRATION DATE WERE MISSING FROM AN UNSPECIFIED NUMBER OF TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876773 ↗
  • MalfunctionFeb 5, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, THERE WAS POOR BARRIER SEPARATION SEEN IN AN UNSPECIFIED NUMBER OF TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21313751 ↗
  • MalfunctionDec 13, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD VACUTAINER® SERUM TUBES WERE USED PAST THEIR EXPIRATION DATE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20930750 ↗
  • MalfunctionDec 3, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM THERE WAS SAMPLE LEAKAGE THROUGH THE CAP ON AN UNSPECIFIED NUMBER OF DEVICES. THERE WERE NO HEALTH CONSEQUENCES OR IMPACTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20831394 ↗
  • MalfunctionOct 28, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED BEFORE USING THE BD VACUTAINER® SERUM, THERE WAS WHITE FOREIGN MATTER IN TWO (2) DEVICES. THERE WERE NO HEALTH CONSEQUENCES OR IMPACTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20545797 ↗
  • MalfunctionAug 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED PRIOR TO USING BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THAT ONE HUNDRED (100) TUBES HAD AN ABNORMAL ADDITIVE APPEARANCE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19950361 ↗
  • MalfunctionJun 28, 2024Component Missing; Component Missing

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM TUBE, THE TUBE DID NOT CLOT LEADING TO THE SAMPLE NOT BEING ABLE TO BE USED. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19639105 ↗
  • MalfunctionApr 25, 2024Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THAT THERE WERE RED BLOOD CELLS IN THE SERUM AFTER CENTRIFUGATION RESULTING IN POOR SERUM; INADEQUATE SEPARATION OF THE SAMPLE. THERE WAS NO REPORT OF IMPACT ON THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19181914 ↗
  • MalfunctionApr 25, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THAT THERE WERE RED BLOOD CELLS IN THE SERUM AFTER CENTRIFUGATION RESULTING IN POOR SERUM; INADEQUATE SEPARATION OF THE SAMPLE. THERE WAS NO REPORT OF IMPACT ON THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19181652 ↗
  • MalfunctionApr 25, 2024Short Fill; Short Fill

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES, THE TUBES WERE UNDERFILLING. THERE WAS NO REPORT OF IMPACT TO THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19181508 ↗
  • MalfunctionApr 24, 2024Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THAT THERE WERE RED BLOOD CELLS IN THE SERUM AFTER CENTRIFUGATION RESULTING IN POOR SERUM; INADEQUATE SEPARATION OF THE SAMPLE. THERE WAS NO REPORT OF IMPACT ON THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19177809 ↗
  • MalfunctionApr 24, 2024Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THAT THERE WERE RED BLOOD CELLS IN THE SERUM AFTER CENTRIFUGATION RESULTING IN POOR SERUM; INADEQUATE SEPARATION OF THE SAMPLE. THERE WAS NO REPORT OF IMPACT ON THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19177080 ↗
  • MalfunctionApr 10, 2024Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THAT THERE WERE RED BLOOD CELLS IN THE SERUM AFTER CENTRIFUGATION RESULTING IN POOR SERUM; INADEQUATE SEPARATION OF THE SAMPLE. THERE WAS NO REPORT OF IMPACT ON THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19079011 ↗
  • MalfunctionDec 14, 2023Component Missing; Component Missing

    IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES HAD MISSING STOPPERS. THERE WAS NO REPORT OF IMPACT TO PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18329067 ↗

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page