BD Vacutainer® Sodium HeparinN (NH) 75 USP Units Blood Collection Tubes 367871Recalls & reported problems

Becton, Dickinson And Company · BD Vacutainer® Sodium Heparin(N) 75 USP Units · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Vacutainer sodium heparin blood collection tube. Reports mostly describe wrong-tube collection errors causing inaccurate sodium results, plus short fills, coagulation, and leaks. 103 reports mention 1 injury and no deaths; 102 malfunctions. No recalls name this device. There were 31 reports in the last 12 months versus 16 the prior year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
103
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 103 adverse event reports that list this exact device (DI 30382903678717), received between Nov 27, 2017 and Apr 9, 2026 — classified as 1 injury, 102 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (46 reports).

2 other recalls cover devices cleared under the same FDA submission (K252040, K944566); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 13 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 30 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202630
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings46
  • Short Fill23
  • Coagulation in Device or Device Ingredient7
  • Volume Accuracy Problem7
  • Separation Problem5
  • Leak/Splash4
  • Device Contamination with Chemical or Other Material3
  • Device Markings/Labelling Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions69
  • No Known Impact Or Consequence To Patient22
  • No Consequences Or Impact To Patient4
  • No Information3
  • No Patient Involvement3
  • Insufficient Information2
  • No Code Available1

Recent adverse event reports

The 25 most recent of 103 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24837666 ↗
  • MalfunctionApr 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24832418 ↗
  • MalfunctionApr 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831068 ↗
  • MalfunctionApr 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831065 ↗
  • MalfunctionApr 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831060 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823531 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823515 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823514 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823510 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823509 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823508 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823507 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823505 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823504 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823500 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821827 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821826 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821825 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821824 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821823 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821822 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821821 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821820 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821819 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD VACUTAINER® SODIUM HEPARINN (NH) 75 USP UNITS BLOOD COLLECTION TUBES, ERRONEOUS SODIUM RESULTS WERE OBSERVED ON ONE PATIENT AFTER USING THE WRONG GREEN TOP TUBE. SODIUM HEPARIN TUBES WERE COLLECTED AND PROCESSED INSTEAD OF LITHIUM HEPARIN TUBES. THE CUSTOMER APPLIED THEIR PATIENT IDENTIFICATION LABEL OVER THE BD TUBE LABEL AT THE TIME OF COLLECTION. PATIENT WAS NOT TREATED FOR ELEVATED SODIUM LEVELS. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821818 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K252040, K944566. One submission often covers many models, so check the product and lot details in each record.

  • Z-2306-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer PST Blood Collection Tubes, Catalog Number 368035, 368036
  • Z-2307-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer PST Gel and Lithium HeparinN (LH) 56 Units Blood Collection Tubes, Catalog Number 368056

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →