Skip to content

BD Vacutainer® Lithium HeparinN (LH) 158 USP Units Blood Collection Tubes 367880Recalls & reported problems

Becton, Dickinson And Company · BD Vacutainer® Lithium Heparin(N) 158 USP Units · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30382903678809), received between May 3, 2018 and May 9, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Short Fill” (2 reports).

1 FDA recall record name this device's identifier.

5 other recalls cover devices cleared under the same FDA submission (K901449); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2019Class IITerminated

    False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

    Z-2298-2019 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Short Fill2
  • Coagulation in Device or Device Ingredient1
  • Device Contamination with Chemical or Other Material1
  • Device Markings/Labelling Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Volume Accuracy Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2024Short Fill; Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 158 USP UNITS BLOOD COLLECTION TUBES AN UNSPECIFIED NUMBER OF TUBES WERE UNDERFILLING. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19280145 ↗
  • MalfunctionMay 21, 2021Short Fill; Volume Accuracy Problem; Short Fill; Volume Accuracy Problem

    IT WAS REPORTED THAT BD VACUTAINER® LITHIUM HEPARINN (LH) 158 USP UNITS BLOOD COLLECTION TUBES UNDERFILLED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " IT IS REPORTED CUSTOMER IS EXPERIENCING UNDER FILLING. VERBIAGE RECEIVED, - "LITHIUM HEPARIN TUBES (LOT 9126505) VACUUM IS ONLY FILLING THE VACUTAINERS WITH 1/3 TO 1/2 THE VOLUME OTHER LITHIUM HEPARIN TUBES FROM OTHER LOTS. I HAVE A TRAY OF THESE TUBES IF YOU WOULD LIKE THEM. IMPACTS INTERNAL BDTI RESEARCH PROJECTS""

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11866317 ↗
  • MalfunctionJun 10, 2019Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT BD VACUTAINER® LITHIUM HEPARINN (LH) 158 USP UNITS BLOOD COLLECTION TUBES HAD INCORRECT LABEL INFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 367880 BATCH NO: UNKNOWN. IT WAS REPORTED THE INFORMATION ON THE UNIT LABEL IS DIFFERENT FROM WHAT IS ONLINE. WE TYPICALLY ORDER THESE BD TUBES FROM VWR AND HAVE NOTICED THAT ALTHOUGH THE PN IS THE SAME FOR EACH, THE UNITS OF LIHEP STATED ON THE PACKAGE LABEL (ORDERED FROM VWR) IS DIFFERENT THAN WHAT IS STATED ON THE BD WEBSITE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8683807 ↗
  • MalfunctionMay 16, 2019Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient

    IT WAS REPORTED THAT BLOOD DRAWN FROM A PIG INTO THE BD VACUTAINER® LITHIUM HEPARIN (LH) 158 USP UNITS BLOOD COLLECTION TUBE CLOTTED AFTER INVERTING THE TUBE "SEVERAL TIMES". THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "(B)(6) 2019 SPOKE WITH THE CUSTOMER. HE WORKS WITH PIGS AND ONE TUBE THAT HE DREW THIS MORNING CLOTTED AFTER HE INVERTED THE TUBE "SEVERAL TIMES." I EXPLAINED THAT WE DO NOT HAVE ANY DATA ON ANIMAL TESTING WITH THE BD VACUTAINERS. HE DISCARDED THE TUBES AND DID NOT TEST THEM."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8615686 ↗
  • MalfunctionJul 26, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT BD VACUTAINER® LITHIUM HEPARINN (LH) 158 USP UNITS BLOOD COLLECTION TUBES EXHIBITED ERRONEOUS RESULTS. NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7723720 ↗
  • MalfunctionMay 3, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THE BD VACUTAINER® LITHIUM HEPARIN (LH) 158 USP UNITS BLOOD COLLECTION TUBES WERE DIRTY.

    Patient problems: No Information

    MAUDE report 7484815 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K901449. One submission often covers many models, so check the product and lot details in each record.

  • Z-2310-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer LH (Lithium Heparin) 102 I.U. Plus Blood Collection Tubes, Catalog Number 368886
  • Z-2309-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer LH (Lithium Heparin) 68 I.U. Plus Blood Collection Tubes, Catalog Number 368884
  • Z-2296-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer Lithium HeparinN (LH) 37 USP Units Blood Collection Tubes, Catalog Number 366664
  • Z-2297-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer Lithium HeparinN (LH) 56 USP Units Blood Collection Tubes, Catalog Number 366667
  • Z-2299-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer Lithium HeparinN 75 USP Units Blood Collection Tubes, Catalog Number 367884

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page