BD Vacutainer® Lithium HeparinN (LH) 95 USP Units Blood Collection Tubes 367886Recalls & reported problems

Becton, Dickinson And Company · TUBE LIHEP PLH 13X100 6.0 PLBL GN · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

These BD Vacutainer lithium heparin tubes are used to collect blood samples for laboratory testing. Of its 279 reports, 277 are malfunction reports, mostly describing underfilled tubes and foreign matter or clotting inside them. Reports fell to 24 in the last 12 months from 58 before. One closed recall, from June 2019, covered falsely high carboxyhemoglobin results when used with one blood gas analyzer.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
280
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 280 adverse event reports that list this exact device (DI 30382903678861), received between Dec 6, 2017 and Sep 23, 2026 — classified as 2 injuries, 278 malfunctions.

The most frequently reported specific device problem is “Short Fill” (154 reports).

1 FDA recall record name this device's identifier.

6 other recalls cover devices cleared under the same FDA submission (K901449); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 7 reportsNov 2024: 1 reportsDec 2024: 4 reportsJan 2025: 7 reportsFeb 2025: 0 reportsMar 2025: 12 reportsApr 2025: 8 reportsMay 2025: 8 reportsJun 2025: 9 reportsJul 2025: 3 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 6 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 4 reportsMay 2026: 7 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2019Class IITerminated

    False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

    Z-2300-2019 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Short Fill154
  • Volume Accuracy Problem96
  • Device Contamination with Chemical or Other Material28
  • Coagulation in Device or Device Ingredient15
  • Material Deformation14
  • Leak/Splash13
  • Difficult to Insert11
  • Incorrect, Inadequate or Imprecise Result or Readings8

Patient problems

  • No Clinical Signs, Symptoms or Conditions175
  • No Consequences Or Impact To Patient73
  • No Known Impact Or Consequence To Patient29
  • No Patient Involvement2
  • Exposure to Body Fluids1
  • Insufficient Information1
  • No Code Available1
  • No Information1

Recent adverse event reports

The 25 most recent of 280 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Short Fill

    It was reported while using BD Vacutainer® Lithium HeparinN (LH) 95 USP Units Blood Collection Tubes, one tube exhibited slow fill. A new tube was used and blood flowed normally. No adverse impact was reported regarding the patient or user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867849 ↗
  • MalfunctionSep 9, 2026Microbial Contamination of Device

    It was reported during performance validation/testing while using BD Vacutainer® Lithium HeparinN (LH) 95 USP Units Blood Collection Tubes, microbial contamination was found inside 10 tubes. There was no health impact or consequences reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50692443 ↗
  • MalfunctionMay 26, 2026Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, ONE TUBE UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25280517 ↗
  • MalfunctionMay 13, 2026Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, THE TUBE OF ONE PATIENT SAMPLE UNDERFILLED RESULTING IN RECOLLECTION. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25166465 ↗
  • MalfunctionMay 6, 2026Short Fill

    REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25094926 ↗
  • MalfunctionMay 6, 2026Short Fill

    REPORT 1 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25094919 ↗
  • MalfunctionMay 4, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED PRIOR TO USING BD VACUTAINER® LITHIUM HEPARIN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, FOREIGN MATTER WAS FOUND IN ONE (1) TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25075597 ↗
  • MalfunctionMay 4, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, FOREIGN MATTER WAS SEEN INSIDE OF ONE TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25072347 ↗
  • MalfunctionMay 1, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT PRIOR TO USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, FOREIGN MATTER WAS SEEN INSIDE OF ONE TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25059409 ↗
  • MalfunctionApr 29, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED PRIOR TO USING BD VACUTAINER® LITHIUM HEPARIN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, FOREIGN MATTER WAS FOUND IN ONE (1) TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25029675 ↗
  • MalfunctionApr 23, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, A FOREIGN OBJECT (A GREEN PLASTIC PIECE) WAS SEEN INSIDE OF ONE TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24971676 ↗
  • MalfunctionApr 21, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARIN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, ONE (1) TUBE LEAKED FROM A CRACK IN THE BOTTOM OF THE TUBE. THE TUBE WAS REPLACED AND DRAW CONTINUED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24946837 ↗
  • MalfunctionApr 8, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED PRIOR TO USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES FOREIGN MATTER WAS SEEN INSIDE OF ONE TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817141 ↗
  • MalfunctionMar 30, 2026Fluid/Blood Leak

    IT WAS REPORTED AFTER USING BD VACUTAINER® LITHIUM HEPARIN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, BLOOD LEAKED FROM THE CAP OF ONE (1) TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24736478 ↗
  • MalfunctionFeb 17, 2026Crack

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, THE BOTTOM OF ONE TUBE WAS CRACKED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24368838 ↗
  • MalfunctionJan 27, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT WHEN USING ONE (1) BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, DEBRIS WAS OBSERVED IN THE TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24192567 ↗
  • MalfunctionDec 22, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT WHILE USING THE BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD, THERE WAS A PROTRUDING PLASTIC SPIKE INSIDE THE TUBE OF 20 DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23871886 ↗
  • MalfunctionDec 19, 2025Short Fill; Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD, ONE (1) TUBE UNDERFILLED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23851347 ↗
  • MalfunctionDec 17, 2025Separation Problem; Separation Problem

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, 2 TUBES HAD COCKED STOPPERS RESULTING IN UNDERFILLED TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23832184 ↗
  • MalfunctionNov 24, 2025Clumping in Device or Device Ingredient; Improper Chemical Reaction; Clumping in Device or Device Ingredient; Improper Chemical Reaction

    IT WAS REPORTED AFTER CENTRIFUGING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, 10 SAMPLES CONTAINED FIBRIN CLUSTERS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23635105 ↗
  • MalfunctionNov 24, 2025Improper Chemical Reaction; Improper Chemical Reaction

    IT WAS REPORTED AFTER CENTRIFUGING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, 30 SAMPLES CONTAINED FIBRIN CLUSTERS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23633405 ↗
  • MalfunctionNov 19, 2025Improper Chemical Reaction; Improper Chemical Reaction

    REPORT 2 OF 3: IT WAS REPORTED AFTER USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, 45% OF SAMPLES ANALYZED CONTAINED FIBRIN. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23600540 ↗
  • MalfunctionNov 19, 2025Improper Chemical Reaction; Improper Chemical Reaction

    REPORT 1 OF 3. IT WAS REPORTED AFTER USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, 59% OF SAMPLES ANALYZED CONTAINED FIBRIN. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23600510 ↗
  • MalfunctionNov 19, 2025Improper Chemical Reaction; Improper Chemical Reaction

    REPORT 3 OF 3: IT WAS REPORTED AFTER USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, 69% OF SAMPLES ANALYZED CONTAINED FIBRIN. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23600499 ↗
  • MalfunctionNov 11, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED WHILE USING BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES, THERE WAS TUBE PUSH OFF SEEN WITH 20 TUBES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23528800 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K901449. One submission often covers many models, so check the product and lot details in each record.

  • Z-2309-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer LH (Lithium Heparin) 68 I.U. Plus Blood Collection Tubes, Catalog Number 368884
  • Z-2298-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer Lithium HeparinN (LH) 158 USP Units Blood Collection Tubes, Catalog Number 367880
  • Z-2310-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer LH (Lithium Heparin) 102 I.U. Plus Blood Collection Tubes, Catalog Number 368886
  • Z-2296-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer Lithium HeparinN (LH) 37 USP Units Blood Collection Tubes, Catalog Number 366664
  • Z-2297-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer Lithium HeparinN (LH) 56 USP Units Blood Collection Tubes, Catalog Number 366667
  • Z-2299-2019 · Jun 12, 2019 · Terminated
    BD Vacutainer Lithium HeparinN 75 USP Units Blood Collection Tubes, Catalog Number 367884

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →