Skip to content

BD Vacutainer® PST™ Blood Collection Tubes 368035Recalls & reported problems

Becton, Dickinson And Company · TUBE PST PLH 13X100 4.5 PLBL L/GN BC · Class 2

All models of BD Vacutainer PST Blood Collection Tubes (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 12, 2019Class IITerminated

    False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

    Z-2306-2019 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jun 12, 2019Class IITerminated

    False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

    Z-2308-2019 · Root cause: Device Design · Recall details · FDA record ↗

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page