BD Vacutainer® FX 10mg 8mg Blood Collection Tubes 368587Recalls & reported problems
Becton, Dickinson And Company · TUBE GLU PLC 13X75 4.0 PLBL GR · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 30382903685876), received between Aug 23, 2019 and Dec 27, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Short Fill” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K901449); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Short Fill2
- Leak/Splash1
- Volume Accuracy Problem1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 27, 2019Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE BD VACUTAINER® SODIUM FLUORIDE 10MG POTASSIUM OXALATE 8 MG (FX) BLOOD COLLECTION TUBE EXPERIENCED DIFFICULTY MOVING THE STOPPER DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 368587, BATCH NO. 9095806. IT WAS REPORTED THAT A SAMPLE HAD FAILED THE SECOND STOPPER PULLOUT TEST. I¿VE BEEN MANAGING SOME PRODUCT TESTING AND DISCOVERED A SAMPLE THAT HAD FAILED THE SECOND STOPPER PULLOUT TEST. THE SAMPLES ARE REQUIRED TO RECORD A 2ND STOPPER PULLOUT FORCE OF ABOVE 0.7LB, BUT THIS SAMPLE ONLY RECORDED A PULLOUT FORCE OF 0.155LB. TUBE TYPE: SODIUM FLUORIDE 13X75MM, 4ML DRAW VOLUME CATALOG NUMBER: 368587 AND LOT NUMBER: 9095806.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9527829 ↗MalfunctionAug 28, 2019Short Fill; Short Fill
IT WAS REPORTED THAT UNDERFILL OCCURRED DURING USE WITH A BD VACUTAINER® PRECISIONGLIDE¿ MULTIPLE SAMPLE NEEDLES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "TUBE EXPERIENCED UNDER FILLING DURING USE." CAT# 368587 AND LOT# 9065849; 29-JUL-2019, 7-AUG-2019.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8940436 ↗MalfunctionAug 23, 2019Short Fill; Volume Accuracy Problem; Short Fill; Volume Accuracy Problem
IT WAS REPORTED THAT BD VACUTAINER® SODIUM FLUORIDE 10MG POTASSIUM OXALATE 8 MG (FX) TUBES HAD LOW DRAW. THIS OCCURRED ON 5 OCCASIONS DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 368587. BATCH/LOT: 8345628. IT WAS REPORTED THAT "WHEN THEY SLID THE TUBES ONTO THE NEEDLE HOLDERS IT ACTED LIKE THEY WERE NOT IN A GOOD VEIN. GROUP WAS HAVING ISSUES WITH SOME OF THE T03 TUBES. WHEN THEY SLID THE TUBES ONTO THE NEEDLE HOLDERS IT ACTED LIKE THEY WERE NOT IN A GOOD VEIN. THE PROBLEM WITH THAT IS THEY HAD ALREADY DRAWN BLOOD TUBES BEFORE AND AFTER THEY TRIED WITH THE GRAY TUBES AND THEY DIDN¿T HAVE TO MOVE THE NEEDLE AT ALL."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8926868 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K901449. One submission often covers many models, so check the product and lot details in each record.
- Z-2297-2019 · Jun 12, 2019 · TerminatedBD Vacutainer Lithium HeparinN (LH) 56 USP Units Blood Collection Tubes, Catalog Number 366667
- Z-2309-2019 · Jun 12, 2019 · TerminatedBD Vacutainer LH (Lithium Heparin) 68 I.U. Plus Blood Collection Tubes, Catalog Number 368884
- Z-2298-2019 · Jun 12, 2019 · TerminatedBD Vacutainer Lithium HeparinN (LH) 158 USP Units Blood Collection Tubes, Catalog Number 367880
- Z-2299-2019 · Jun 12, 2019 · TerminatedBD Vacutainer Lithium HeparinN 75 USP Units Blood Collection Tubes, Catalog Number 367884
- Z-2310-2019 · Jun 12, 2019 · TerminatedBD Vacutainer LH (Lithium Heparin) 102 I.U. Plus Blood Collection Tubes, Catalog Number 368886
- Z-2296-2019 · Jun 12, 2019 · TerminatedBD Vacutainer Lithium HeparinN (LH) 37 USP Units Blood Collection Tubes, Catalog Number 366664
Other Tubes, vials, systems, serum separators, blood collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.