Skip to content

BD Vacutainer® Eclipse™ Blood Collection Needle 368608Recalls & reported problems

Becton, Dickinson And Company · Becton Dickinson and Company, USA 22G x 1 1/4 · Class 2

All models of BD Vacutainer Eclipse Blood Collection Needle (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Vacutainer Eclipse blood collection needle is a needle used to draw blood into collection tubes. Its 489 reports most often describe a safety shield that is hard or slow to activate, and leaks. Some describe needle sticks, and the reports include 41 injuries and no deaths. Reports rose from 59 the year before to 77 in the latest year. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
486
Deaths reported
0
Injuries reported
41
Recalls naming it
0

Summary

The FDA's MAUDE database contains 486 adverse event reports that list this exact device (DI 30382903686088), received between Nov 10, 2017 and Sep 23, 2026 — classified as 41 injuries, 445 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Activation” (122 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 4 reportsJan 2024: 7 reportsFeb 2024: 8 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 5 reportsJul 2024: 4 reportsAug 2024: 3 reportsSep 2024: 9 reportsOct 2024: 5 reportsNov 2024: 3 reportsDec 2024: 7 reportsJan 2025: 9 reportsFeb 2025: 0 reportsMar 2025: 6 reportsApr 2025: 4 reportsMay 2025: 6 reportsJun 2025: 7 reportsJul 2025: 5 reportsAug 2025: 3 reportsSep 2025: 4 reportsOct 2025: 5 reportsNov 2025: 10 reportsDec 2025: 12 reportsJan 2026: 20 reportsFeb 2026: 11 reportsMar 2026: 10 reportsApr 2026: 3 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult or Delayed Activation122
  • Leak/Splash109
  • Product Quality Problem71
  • Device Contamination with Chemical or Other Material31
  • Fluid/Blood Leak29
  • Break26
  • Detachment of Device or Device Component22
  • Separation Problem16

Patient problems

  • No Clinical Signs, Symptoms or Conditions245
  • No Known Impact Or Consequence To Patient148
  • No Consequences Or Impact To Patient41
  • Needle Stick/Puncture40
  • No Patient Involvement8
  • No Information5
  • Exposure to Body Fluids2
  • Blood Loss1

Recent adverse event reports

The 25 most recent of 486 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876310 ↗
  • MalfunctionSep 23, 2026Fluid/Blood Leak

    It was reported while using BD Vacutainer® Eclipse¿ Blood Collection Needle, blood leakage from the rubber sleeve occurred with an unspecified number of devices. No adverse impact was reported regarding the patient or user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870512 ↗
  • MalfunctionSep 23, 2026Difficult or Delayed Activation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863115 ↗
  • MalfunctionSep 8, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50670376 ↗
  • MalfunctionMay 6, 2026Difficult or Delayed Activation

    REPORT 3 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD IS MISALIGNED OR EASILY POPS OFF OF AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25095826 ↗
  • MalfunctionMay 6, 2026Difficult or Delayed Activation

    REPORT 2 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD IS MISALIGNED OR EASILY POPS OFF OF AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25095824 ↗
  • InjuryMay 6, 2026Physical Resistance/Sticking

    REPORT 1 OF 3. IT WAS REPORTED THAT WHEN THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WAS IN PROCESS OF BEING REMOVED, THE SAFETY WAS UNINTENTIONALLY KNOCKED OUT OF PLACE BY THE PATIENT AND STAFF COULD NOT ACTIVATE THE SAFETY AND WAS STUCK BY THE NEEDLE. STAFF MEMBER FOLLOWED STANDARD NEEDLESTICK PROTOCOL PER FACILITY. IT IS UNSPECIFIED IF MEDICAL INTERVENTION OR TREATMENT TOOK PLACE.

    Patient problems: Needle Stick/Puncture

    MAUDE report 25095818 ↗
  • MalfunctionMay 4, 2026Detachment of Device or Device Component

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE NEEDLE AND HOLDER WERE NOT FIXED AND SPUN AROUND. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25071440 ↗
  • MalfunctionApr 13, 2026Difficult or Delayed Activation

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD DETACHED FROM AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24866572 ↗
  • MalfunctionApr 6, 2026Improper Flow or Infusion

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, INSUFFICIENT BLOOD FLOW WAS SEEN WITH AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24795145 ↗
  • MalfunctionApr 3, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO BLOOD DRAW USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE NEEDLE AND SINGLE USE HOLDER COULD NOT BE SECURED, THE CUSTOMER ATTEMPTED TO TURN IT, BUT IT CONTINUED TO ROTATE AND WOULD NOT LOCK INTO PLACE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24780533 ↗
  • MalfunctionMar 31, 2026Difficult or Delayed Activation

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF FROM THREE (3) NEEDLES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24746533 ↗
  • MalfunctionMar 31, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT PRIOR TO USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, FOREIGN MATTER WAS FOUND ON THE TIP OF ONE NEEDLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24746529 ↗
  • MalfunctionMar 25, 2026Premature Separation

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD HAS DETACHED FROM AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24690606 ↗
  • MalfunctionMar 18, 2026Premature Separation

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD OF AN UNSPECIFIED NUMBER OF DEVICES BROKE OFF LEAVING THE NEEDLE EXPOSED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24635140 ↗
  • MalfunctionMar 12, 2026Loss of or Failure to Bond

    IT WAS REPORTED PRIOR TO USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, 6 DEVICES WERE FOUND TO HAVE THE HUB SEPARATED FROM THE COLLAR. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24584157 ↗
  • MalfunctionMar 5, 2026Loss of or Failure to Bond

    REPORT 3 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF THE DEVICE LEADING TO A CLEAN NEEDLESTICK. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: Needle Stick/Puncture

    MAUDE report 24524749 ↗
  • MalfunctionMar 5, 2026Loss of or Failure to Bond

    REPORT 1 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24524742 ↗
  • MalfunctionMar 5, 2026Loss of or Failure to Bond

    REPORT 2 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24524741 ↗
  • MalfunctionMar 5, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, BLOOD FROM ONE DEVICE LEAKED INTO THE HOLDER WHILE TUBES WERE BEING CHANGED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24524738 ↗
  • MalfunctionMar 5, 2026Difficult or Delayed Activation; Premature Separation

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD OF ONE DEVICE BROKE OFF. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520425 ↗
  • MalfunctionFeb 27, 2026Difficult or Delayed Activation

    REPORT 2 OF 3. IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24476493 ↗
  • MalfunctionFeb 27, 2026Difficult or Delayed Activation

    REPORT 3 OF 3. IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24475709 ↗
  • MalfunctionFeb 27, 2026Difficult or Delayed Activation

    REPORT 1 OF 3. IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24475623 ↗
  • MalfunctionFeb 27, 2026Difficult or Delayed Activation

    IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF OF ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24474226 ↗

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page