Skip to content

BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder 368659Recalls & reported problems

Becton, Dickinson And Company · BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 30382903686590), received between Nov 28, 2017 and Mar 20, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash2
  • Cut In Material1
  • Device Markings/Labelling Problem1
  • Fluid/Blood Leak1
  • Physical Resistance/Sticking1
  • Product Quality Problem1
  • Retraction Problem1
  • Volume Accuracy Problem1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • No Clinical Signs, Symptoms or Conditions3
  • Needle Stick/Puncture1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 20, 2025Product Quality Problem; Product Quality Problem

    IT WAS REPORTED THAT WHILE USING ONE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER, THE HUB DETACHED EXPOSING THE NEEDLE WHEN THE SAFETY SHIELD WAS ACTIVATED. NO PATIENT OR USER IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21659740 ↗
  • MalfunctionMar 1, 2024Volume Accuracy Problem; Volume Accuracy Problem

    IT WAS REPORTED THAT DURING THE USE OF BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER, THE BLOOD ENTERS THE TUBING REALLY FAST WHEN THE NEEDLE WAS INSERTED INTO THE PATIENT. THEN WHEN THE VACUTAINER TUBE WAS CONNECTED, IT SHOOTS THE BLOOD IN REALLY QUICKLY AND LOSES ALL PRESSURE; IT STOPS WORKING. NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18817530 ↗
  • InjuryMay 22, 2023Retraction Problem; Physical Resistance/Sticking; Retraction Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT WHEN RETRACTING THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER THE NEEDLE WOULD NOT RETRACT AND A NEEDLE STICK INJURY OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT, AND INFORMATION REGARDING INTERVENTION FOR HEALTHCARE WORKER HAS NOT YET BEEN OBTAINED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTING STICKING HERSELF WITH THE NEEDLE WHILE TRYING TO RETRACT IT.

    Patient problems: Needle Stick/Puncture; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16981915 ↗
  • MalfunctionJun 19, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE TUBING OF THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER WAS CUT CAUSING A LEAK. FOUND DURING USE. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7614037 ↗
  • MalfunctionMay 21, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE TUBING OF THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER WAS CUT. FOUND BEFORE USE. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7530246 ↗
  • MalfunctionApr 16, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT THE PACKAGING FOR THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER CONTAINED MIXED PRODUCTS; 25 GAUGE BUTTERFLIES FOUND IN A 23 GAUGE PACKAGE. NO INJURY OR MEDICAL INTERVENTION REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7433306 ↗
  • MalfunctionNov 28, 2017Fluid/Blood Leak; Cut In Material

    IT WAS REPORTED THAT TUBING ON THE BD VACUTAINER PUSH BUTTON BLOOD COLLECTION SET WAS CUT, RESULTING IN BLOOD LEAKAGE. NO SERIOUS INJURY OR MEDICAL INTERVENTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7064375 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page