BD Vacutainer® Serum Blood Collection Tubes 368660Recalls & reported problems
Becton, Dickinson And Company · BD Vacutainer® Serum · Class 2
All models of BD Vacutainer Serum Blood Collection Tubes (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 30382903686606), received between Nov 7, 2017 and Oct 22, 2020 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (5 reports).
6 other recalls cover devices cleared under the same FDA submission (BK050036, K960250); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Contamination with Chemical or Other Material5
- Device Markings/Labelling Problem2
- Leak/Splash2
- Short Fill2
- Volume Accuracy Problem2
- Break1
- Detachment Of Device Component1
- Device Operates Differently Than Expected1
Patient problems
- No Known Impact Or Consequence To Patient8
- No Consequences Or Impact To Patient3
- No Patient Involvement3
- No Information1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 22, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED WITH A BD VACUTAINER® SERUM BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE WAS HAIR ON THE TUBE.
Patient problems: No Patient Involvement
MAUDE report 10718614 ↗MalfunctionOct 22, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT HAIR WAS DISCOVERED IN TUBE PRIOR TO USE WITH A BD VACUTAINER® SERUM BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE WAS HAIR ON THE TUBE.
Patient problems: No Patient Involvement
MAUDE report 10718612 ↗MalfunctionOct 21, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED BEFORE USE THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES HAD FOREIGN MATTER IN TUBE; BIOLOGICAL AND NON-BIOLOGICAL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER STATED "THERE IS A PARTICULATE ON THE TUBE."
Patient problems: No Patient Involvement
MAUDE report 10715217 ↗MalfunctionMar 24, 2020Leak/Splash; Short Fill; Volume Accuracy Problem; Leak/Splash; Short Fill; Volume Accuracy Problem
IT WAS REPORTED DURING USE THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES HAD UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD CRACKED TUBE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "HAD A CRACK IN THE TUBE IN THE AREA WHERE THE STOPPER IS. THE TUBE HAD A LOW DRAW VOLUME, LIKELY BECAUSE OF THE TUBE".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9872269 ↗MalfunctionMar 18, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT ERRONEOUS RESULTS AND INCORRECT ADDITIVE WERE FOUND AFTER USE WITH A BD VACUTAINER® SERUM BLOOD COLLECTION TUBE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, (B)(6) 2020: RECEIVED PHONE CALL. SHE DOES NOT WANT TO FILE A COMPLAINT. SHE WOULD LIKE TO KNOW WHAT ADDITIVE IS THIS CLOT ACTIVATOR AS THEIR REFERENCE LAB REJECTED SPECIMEN DUE TO CLOT ACTIVATOR IN VACUTAINER. TUBE WAS SENT OFF TO TEST MICROALBUMIN URINE BUT WAS REJECTED. NO OTHER INFORMATION IS AVAILABLE, NO SAMPLES AVAILABLE, NO ACK LETTER REQUIRED AND NO CLOSURE LETTER NEEDED. (B)(6) 2020: I EXPLAINED TO THE CUSTOMER THAT THE NURSES SHOULD NOT BE USING SERUM TUBES TO COLLECT URINE SPECIMENS. I SENT HER THE LINK TO THE IFU TO EXPLAIN THAT THE PLASTIC SERUM TUBES CONTAIN SILICONE AND MICRONIZED SILICA PARTICLES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9849874 ↗MalfunctionMay 30, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THERE WAS WATER CONDENSATION IN THE TUBES WITH A BD VACUTAINER® SERUM BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 368660, BATCH NO. 8345630 AND 8215776. IT WAS REPORTED THERE IS WATER CONDENSATION IN THE TUBES. INFORMED CUSTOMER THAT THEY ARE MOST LIKELY SEEING THE CLOT ACTIVATOR THAT IS SPRAY COATED ON THE SIDE WALLS OF THE TUBE. IT APPEARS AS A DRIED WHITE FILM. SHE STATES THIS IS WHAT THEY ARE SEEING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8654524 ↗MalfunctionJun 12, 2018Short Fill; Short Fill
IT WAS REPORTED THAT BD VACUTAINER® SERUM BLOOD COLLECTION TUBES WERE UNDERFILLING. NO REPORT OF SERIOUS INJURY OR MEDICAL INTERVENTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7594347 ↗MalfunctionJun 11, 2018Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED THAT BD VACUTAINER® SERUM BLOOD COLLECTION TUBES UNDER FILLED. NO SERIOUS INJURY OR MEDICAL INTERVENTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7589524 ↗MalfunctionJun 11, 2018Leak/Splash; Material Deformation; Leak/Splash; Material Deformation
IT WAS REPORTED THAT THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES HAD COLLAPSED TUBES CAUSED NO SERIOUS INJURY MEDICAL INTERVENTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7589005 ↗MalfunctionJun 11, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
IT WAS REPORTED WITH THE USE OF THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THERE WAS AN ISSUE WITH ERRONEOUS RESULTS. IT WAS STATED ¿THE CUSTOMER REPORTED THAT THE TUBE LOOKS LIKE IT WAS WET INSIDE THAT THEN IT DRIED AND LEFT SPOTS ALL OVER THE TUBE.¿ THERE WAS NO REPORT OF INJURY OR FURTHER MEDICAL INTERVENTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7588326 ↗MalfunctionMay 11, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT BD VACUTAINER® SERUM BLOOD COLLECTION TUBES HAD INCORRECT COLOR.
Patient problems: No Information
MAUDE report 7506858 ↗MalfunctionApr 25, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND INSIDE A BD VACUTAINER® SERUM BLOOD COLLECTION TUBES. THERE WAS NO REPORT OF EXPOSURE, SERIOUS INJURY OR MEDICAL INTERVENTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7462404 ↗MalfunctionMar 26, 2018Break
IT WAS REPORTED THAT A BD VACUTAINER¿ VENOUS BLOOD COLLECTION SERUM SILICONE-COATED PLASTIC TUBE WITH CONVENTIONAL STOPPER WAS FOUND BROKEN AT COLLECTION SITE. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION NEEDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7369818 ↗MalfunctionNov 29, 2017Device Markings/Labelling Problem
IT WAS REPORTED THAT ONE RED BD VACUTAINER® SERUM TUBE IN THE BOX DID NOT HAVE A LABEL. NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7068346 ↗MalfunctionNov 7, 2017Detachment Of Device Component; Device Operates Differently Than Expected
IT WAS REPORTED THAT FOR THE 13 X 100 MM 6.0 ML BD VACUTAINER® PLUS PLASTIC TUBE SERUM TUBE, THE STOPPER WAS MORE DIFFICULT TO PIERCE AND THIS LED TO STOPPER PULL OUT WHEN REMOVING COLLECTION DEVICE. NO SERIOUS INJURY, AND NO MEDICAL INTERVENTION. NO MUCOUS MEMBRANE EXPOSURE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7008129 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under BK050036, K960250. One submission often covers many models, so check the product and lot details in each record.
- Z-1618-2022 · May 25, 2022 · Open, Classifieda. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP GLD 3.5ML PLAS VACUTAINER (100/BX), Model Number: 367983. b. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/
- Z-0968-2018 · Aug 29, 2017 · TerminatedBD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing an
- Z-2434-2015 · Jun 23, 2015 · TerminatedBD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting, transporting, separating, and processing blood in a closed tube.
- Z-2631-2014 · Jun 23, 2014 · TerminatedBD Vacutainer SST Tubes (RED/GREY TOP SST TUBE 3.5ML) Part Number: BD 367981 Part Number: 87-7981-5 Single use tube used to collect, separate, transport or process blood specimens
- Z-2629-2014 · Jun 23, 2014 · TerminatedBD Vacutainer Serum Tubes (RED TOP PLAIN TUBE 6.0 ML)Part Number: BD 367815 FMC Part Number: 87-7815-4 Single use tube used to collect, separate, transport or process blood spec
- Z-0052-05 · Aug 20, 2004 · TerminatedBD Vactuainer SST glass and plastic tubes.
Other Tubes, vials, systems, serum separators, blood collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page