BD Vacutainer® UltraTouch™ Push Button Blood Collection Set with PAH 368685Recalls & reported problems

Becton, Dickinson And Company · UT-PBBCS 23G 7 PAH CE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Vacutainer UltraTouch push button blood collection set is used to draw blood samples. Most of the 82 reports describe blood leaking from the tubing, sleeve, or adapter. The reports include no injuries and no deaths, and reports rose from 3 in the previous 12 months to 79 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
77
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 30382903686859), received between Nov 8, 2024 and Aug 17, 2026 — classified as 77 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (61 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 9 reportsFeb 2026: 20 reportsMar 2026: 16 reportsApr 2026: 25 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202625
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak61
  • Leak/Splash15
  • Material Split, Cut or Torn5
  • Volume Accuracy Problem2
  • Packaging Problem1
  • Retraction Problem1
  • Short Fill1

Patient problems

  • No Clinical Signs, Symptoms or Conditions71
  • Exposure to Body Fluids4
  • Discomfort2
  • Pain2

Recent adverse event reports

The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50278297 ↗
  • MalfunctionApr 29, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD IS LEAKING FROM THE LUER ADAPTER OF AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034818 ↗
  • MalfunctionApr 29, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, THE TUBING OF AN UNSPECIFIED NUMBER OF DEVICES LEAKED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034805 ↗
  • MalfunctionApr 27, 2026Short Fill

    IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, TUBING WAS PINCHED RESULTING IN POOR BLOOD FLOW. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25009426 ↗
  • MalfunctionApr 23, 2026Retraction Problem

    IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, THE NON PATIENT NEEDLE SLEEVE WAS PIERCED BY THE NEEDLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24972267 ↗
  • MalfunctionApr 23, 2026Fluid/Blood Leak; Material Split, Cut or Torn

    REPORT 4 OF 4: IT WAS REPORTED THAT DURING COLLECTION OF ONE PATIENT USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, A HOLE IN THE TUBING RESULTED IN SAMPLE LEAKAGE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968291 ↗
  • MalfunctionApr 23, 2026Fluid/Blood Leak; Material Split, Cut or Torn

    REPORT 2 OF 4: IT WAS REPORTED THAT DURING COLLECTION OF ONE PATIENT USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, A HOLE IN THE TUBING RESULTED IN SAMPLE LEAKAGE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968267 ↗
  • MalfunctionApr 23, 2026Fluid/Blood Leak; Material Split, Cut or Torn

    REPORT 1 OF 4: IT WAS REPORTED THAT DURING COLLECTION OF ONE PATIENT USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, A HOLE IN THE TUBING RESULTED IN SAMPLE LEAKAGE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968239 ↗
  • MalfunctionApr 23, 2026Fluid/Blood Leak; Material Split, Cut or Torn

    REPORT 3 OF 4: IT WAS REPORTED THAT DURING COLLECTION OF ONE PATIENT USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, A HOLE IN THE TUBING RESULTED IN SAMPLE LEAKAGE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968229 ↗
  • MalfunctionApr 20, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING [BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE TUBING OF THE DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24934037 ↗
  • MalfunctionApr 20, 2026Fluid/Blood Leak

    REPORT 1 OF 2 IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE TUBING OF THE DEVICE WITH ONE (1) PATIENT. THE PATIENT HAD TO HAVE AN ADDITIONAL NEEDLESTICK FOR THE DRAW. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: Discomfort; Pain

    MAUDE report 24933480 ↗
  • MalfunctionApr 20, 2026Fluid/Blood Leak

    REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE TUBING OF THE DEVICE WITH ONE (1) PATIENT. THE PATIENT HAD TO HAVE AN ADDITIONAL NEEDLESTICK FOR THE DRAW. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: Discomfort; Pain

    MAUDE report 24933477 ↗
  • MalfunctionApr 17, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, SAMPLE LEAKAGE OCCURS WITH AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24915760 ↗
  • MalfunctionApr 17, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT WHILE USING THE VACUTAINER ULTRATOUCH THAT THE DEVICE SET LEAKED INTO THE HOSE DOWN TO THE VACUTAINER. THIS OCCURRED IN AN UNSPECIFIED AMOUNT OF DEVICES. NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911006 ↗
  • MalfunctionApr 16, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, THERE WAS A HOLE IN THE TUBING OF ONE DEVICE RESULTING IN SAMPLE LEAKAGE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24903881 ↗
  • MalfunctionApr 16, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT WHEN USING THE BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, ONE (1) UNIT EXHIBITED BLOOD LEAKAGE AT LUER. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24900019 ↗
  • MalfunctionApr 8, 2026Fluid/Blood Leak

    REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, THERE IS A HOLE IN THE TUBING OF AN UNSPECIFIED NUMBER OF DEVICES RESULTING IN SAMPLE LEAKAGE AND THE PATIENTS HAVING TO BE RE-PRICKED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821967 ↗
  • MalfunctionApr 8, 2026Fluid/Blood Leak

    REPORT 1 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, THERE IS A HOLE IN THE TUBING OF ONE DEVICE RESULTING IN SAMPLE LEAKAGE AND THE PATIENT HAVING TO BE RE-PRICKED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821803 ↗
  • MalfunctionApr 7, 2026Fluid/Blood Leak

    REPORT 2 OF 2. IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE TUBING OF THE DEVICE. THIS OCCURRED AN UNSPECIFIED NUMBER OF TIMES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24809043 ↗
  • MalfunctionApr 7, 2026Fluid/Blood Leak

    REPORT 1 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE TUBING OF THE DEVICE. THIS OCCURRED AN UNSPECIFIED NUMBER OF TIMES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24809042 ↗
  • MalfunctionApr 3, 2026Fluid/Blood Leak

    IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE NON-PATIENT END OF THE DEVICE AT THE CONNECTION BETWEEN THE HUB AND SLEEVE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24780425 ↗
  • MalfunctionApr 3, 2026Fluid/Blood Leak; Volume Accuracy Problem

    REPORT 2 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE HUB OF THE DEVICE WHEN ATTEMPTING TO DRAW INTO A BLOOD COLLECTION TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24780222 ↗
  • MalfunctionApr 2, 2026Fluid/Blood Leak

    REPORT 1 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE HUB OF THE DEVICE WHEN ATTEMPTING TO DRAW INTO A BLOOD COLLECTION TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24771270 ↗
  • MalfunctionApr 2, 2026Fluid/Blood Leak; Volume Accuracy Problem

    REPORT 3 OF 3. IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD LEAKED FROM THE HUB OF THE DEVICE WHEN ATTEMPTING TO DRAW INTO A BLOOD COLLECTION TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24771252 ↗
  • MalfunctionApr 2, 2026Fluid/Blood Leak

    REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET WITH PAH, BLOOD FROM AN UNSPECIFIED NUMBER OF DEVICES LEAKED BETWEEN THE TUBE AND THE HOLDER. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24770152 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →