BD Phoenix™ PMIC/ID-108 448608Recalls & reported problems
Becton, Dickinson And Company · Panel Phoenix Pmic/Id 108 · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 30382904486083), received between Mar 3, 2021 and Sep 15, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (4 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Dec 8, 2020Terminated
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
Z-0762-2021 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings4
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT WHILE USING BD PHOENIX¿ PMIC/ID-108 THAT THERE WAS A INCORRECT LABEL INFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REVIEWING DOCUMENTATION ON THE BD WEBSITE, TECHNICAL SERVICES IDENTIFIED A DISCREPANCY IN THE IFU RECORD. WHEN SEARCHING FOR THE IFU FOR CAT # 448608, THE IFU FOR NMIC-303 LOADS. NO IMPACT TO PATIENTS OR CUSTOMERS WAS ASSOCIATED WITH THIS DISCREPANCY AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17758408 ↗MalfunctionOct 3, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAD MISIDENTIFICATIONS ON BD PHOENIX¿ PMIC/ID-108. THERE WAS ONE OCCURRENCE IN THIS EVENT WITH PATIENT INVOLVEMENT BUT PATIENT TREATMENT WAS NOT AFFECTED AND THERE WERE NO ADVERSE EVENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: CUSTOMER IS REPORTING MISIDENTIFICATION WITH PANEL 448608, LOT 2048460. PATIENT IDENTIFIER: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15529323 ↗MalfunctionOct 3, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAD MISIDENTIFICATIONS ON BD PHOENIX¿ PMIC/ID-108. THERE WAS ONE OCCURRENCE IN THIS EVENT WITH PATIENT INVOLVEMENT BUT PATIENT TREATMENT WAS NOT AFFECTED AND THERE WERE NO ADVERSE EVENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: "CUSTOMER IS REPORTING MISIDENTIFICATION WITH PANEL 448608, LOT 2048460. PATIENT IDENTIFIER: (B)(6)."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15529199 ↗MalfunctionOct 3, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAD MISIDENTIFICATIONS ON BD PHOENIX¿ PMIC/ID-108. THERE WAS ONE OCCURRENCE IN THIS EVENT WITH PATIENT INVOLVEMENT BUT PATIENT TREATMENT WAS NOT AFFECTED AND THERE WERE NO ADVERSE EVENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: "CUSTOMER IS REPORTING MISIDENTIFICATION WITH PANEL 448608, LOT 2048460. PATIENT (B)(6)."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15529161 ↗MalfunctionMar 3, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE TESTING WITH BD PHOENIX¿ PMIC/ID-108 AN UNSPECIFIED PANEL WAS USED WITH DIFFERENT SWABS AND STAPH IDENTIFICATIONS DID MATCH. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION, THE CUSTOMER HAS NOT RESPONDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11404870 ↗
Other SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION devices
- BD Phoenix™ M50 Automated Microbiology System 443624Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION3 recalls417 reports
- BD Phoenix™ NMIC/ID-307 449289Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION529 reports
- BD Phoenix™ PMIC/ID-107 448607Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION2 recalls169 reports
- BD Phoenix™ NMIC-306 449292Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION1 recall143 reports
- VITEK® 2 AST-GP67Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls83 reports
- BD Phoenix™ AST Broth 246003Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION182 reports
- VITEK® 2 AST-GP75Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls52 reports
- VITEK® 2 AST-GP78Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls41 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page