Skip to content

BD Phoenix™ SMIC-101 448803Recalls & reported problems

Becton, Dickinson And Company · BD Phoenix™ SMIC-101 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 30382904488032), received between May 20, 2022 and Feb 13, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (5 reports).

3 other recalls cover devices cleared under the same FDA submission (K032675, K050745, K050747, K050780, K050865, K050881, K050946, K050982, K051109, K051138, K051204, K051266, K051272, K051692, K062206); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD PHOENIX¿ SMIC-101 THE MIC RESULT FOR THE DRUG VANCOMYCIN IS SUPPRESSED AND CLINDAMYCIN IS BEING REPORTED INSTEAD. IT IS TO BE NOTED THE RULE FOR SUPPRESSION IS ACCURATE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24337386 ↗
  • MalfunctionJul 28, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD PHOENIX¿ SMIC-101 PATIENT ISOLATE RESULTS FOR THE DRUGS ERYTHROMYCIN AND CLINDAMYCIN DO NOT MATCH THE D TEST RESULT NOR REPEAT RESULTS ON THE DEVICE. PATIENT ONE HAD A SENSITIVE CLINDAMYCIN AND A RESISTANT ERYTHROMYCIN RESULT. THE D TEST RESULT SHOWS BOTH CLINDAMYCIN AND ERYTHROMYCIN AS RESISTANT. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 1 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22636456 ↗
  • MalfunctionJul 28, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD PHOENIX¿ SMIC-101 PATIENT ISOLATE RESULTS FOR THE DRUGS ERYTHROMYCIN AND CLINDAMYCIN DO NOT MATCH THE D TEST RESULT NOR REPEAT RESULTS ON THE DEVICE. PATIENT THREE HAD A SENSITIVE CLINDAMYCIN AND A RESISTANT ERYTHROMYCIN RESULT. THE D TEST RESULT SHOWS BOTH CLINDAMYCIN AND ERYTHROMYCIN AS RESISTANT. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 3 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22636282 ↗
  • MalfunctionJul 28, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD PHOENIX¿ SMIC-101 PATIENT ISOLATE RESULTS FOR THE DRUGS ERYTHROMYCIN AND CLINDAMYCIN DO NOT MATCH THE D TEST RESULT NOR REPEAT RESULTS ON THE DEVICE. PATIENT TWO HAD A RESISTANT CLINDAMYCIN AND A SENSITIVE ERYTHROMYCIN RESULT. THE D TEST RESULT SHOWS BOTH CLINDAMYCIN AND ERYTHROMYCIN AS RESISTANT. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22636121 ↗
  • MalfunctionMay 20, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT BD PHOENIX¿ SMIC-101 HAS INSUFFICIENT INDICATOR AND NO GROWTH WITH 448803 PANELS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING INSUFFICIENT INDICATOR AND NO GROWTH WITH 448803 PANELS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14455823 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032675, K050745, K050747, K050780, K050865, K050881, K050946, K050982, K051109, K051138, K051204, K051266, K051272, K051692, K062206. One submission often covers many models, so check the product and lot details in each record.

  • Z-0935-2019 · Dec 20, 2018 · Terminated
    Phoenix AP AST Indicator Bag, Catalog Number 246006
  • Z-0933-2019 · Dec 20, 2018 · Terminated
    Phoenix AST-s Indicator, Catalog Number 246009
  • Z-0934-2019 · Dec 20, 2018 · Terminated
    Phoenix AST Indicator, Catalog Number 246004

Other SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION devices

All SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION devices →