Skip to content

VITEK® 2 AST-GN95Recalls & reported problems

Biomerieux , Inc. · VITEK® 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems to determine the susceptibility of clinically significant aerobic gramnegative bacilli · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03573026563844), received between Sep 20, 2018 and Sep 20, 2018 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

3 other recalls cover devices cleared under the same FDA submission (K040845, K072038, K103752, K110106, K113200, K152075, K161217, K161227, N50510); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 14, 2024Open, Classified

    Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

    Z-1464-2024 · Root cause: Employee error · FDA record ↗

  2. Nov 9, 2022Open, Classified

    As the temperature and time out of range was exceeded, product performances cannot be guaranteed

    Z-0853-2023 · Root cause: Storage · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Patient Involvement2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 20, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER FROM THE UNITED STATES NOTIFIED BIOMÉRIEUX OF FALSE RESISTANT CEFTRIAXONE RESULTS OBTAINED ON A CAP SURVEY STRAIN (CITROBACTER BRAAKII) WHEN TESTING WITH THE VITEK® 2 AST-GN95 TEST KIT, LOT 6850485403. THE RESISTANT CEFTRIAXONE RESULT WAS OBTAINED WITH INITIAL AND REPEAT TESTING. THE EXPECTED RESULT WAS SUSCEPTIBLE. WHEN THE CUSTOMER RECEIVED THE SURVEY RESULTS, SHE TESTED WITH MICROSCAN AND OBTAINED SUSCEPTIBLE RESULTS. THE CUSTOMER DID REPEAT TESTING ON (B)(6) 2018 AND OBTAINED A SUSCEPTIBLE RESULT. AS THERE IS NO PATIENT ASSOCIATED WITH THIS QUALITY CONTROL STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.

    Patient problems: No Patient Involvement

    MAUDE report 7893636 ↗
  • MalfunctionSep 20, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER FROM THE UNITED STATES NOTIFIED BIOMÉRIEUX OF FALSE RESISTANT RESULTS FOR A CAP SURVEY STRAIN (D-08) WHEN TESTING WITH THE VITEK® 2 AST-GN95 TEST KIT. THE ALCALIGENES FAECALIS STRAIN TESTED FALSE RESISTANT (MIC >=128) FOR PIPERACILLIN/TAZOBACTAM USING THE AST-GN95 CARD, LOT 6850485403. THE EXPECTED RESULT IS SUSCEPTIBLE. THE CUSTOMER DID REPEAT TESTING AND STILL OBTAINED RESISTANT RESULTS (MIC >/=128) WITH THE SAME LOT. AS THERE IS NO PATIENT ASSOCIATED WITH THIS QUALITY CONTROL STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.

    Patient problems: No Patient Involvement

    MAUDE report 7893634 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K040845, K072038, K103752, K110106, K113200, K152075, K161217, K161227, N50510. One submission often covers many models, so check the product and lot details in each record.

  • Z-0833-2023 · Nov 9, 2022 · Open, Classified
    VITEK 2 REAGENT AST-GN69 TEST KIT 20 CARDS, CATALOG 413400
  • Z-0861-2023 · Nov 9, 2022 · Open, Classified
    VITEK 2 REAGENT AST-XN15 TEST KIT 20 CARDS, CATALOG 423829
  • Z-0837-2023 · Nov 9, 2022 · Open, Classified
    VITEK 2 REAGENT AST-GN80 TEST KIT 20 CARDS, CATALOG 413437

Other System, Test, Automated, Antimicrobial Susceptibility, Short Incubation devices

All System, Test, Automated, Antimicrobial Susceptibility, Short Incubation devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.