Cleo 21-7222-24Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Cleo (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This ICU Medical Cleo infusion set is used to deliver fluids or medication through tubing and a catheter. Reports mostly describe catheters that would not stick to the skin, leaks at tubing connections, and kinked catheters, with several mentions of insulin not being delivered and high blood sugar. There are 26 reports that name its identifier, including no deaths and 5 injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 30610586028353), received between Jul 10, 2017 and May 9, 2024 — classified as 5 injuries, 21 malfunctions.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of or Failure to Bond6
  • Infusion or Flow Problem5
  • Fluid/Blood Leak4
  • Mechanical Problem4
  • Defective Component2
  • Leak/Splash2
  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Hyperglycemia8
  • Foreign Body In Patient1
  • Local Reaction1
  • No Consequences Or Impact To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2024Separation Problem; Separation Problem

    IT WAS REPORTED THAT THE CATHETERS DIDN'T STICK TO THE SKIN DURING INSERTION. THIS OCCURRED WITH 3 - 4 CATHETERS. THERE WAS PATIENT INVOLVEMENT AND NO HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19281793 ↗
  • MalfunctionMay 11, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PATIENT OBSERVED A LEAKAGE AT THE ADAPTER BETWEEN THE TUBING AND THE CATHETER, AFTER A FEW HOURS OF INSERTION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16913596 ↗
  • MalfunctionMay 10, 2023Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT THE ADHESIVE BARELY OR DOES NOT ADHERE TO THE PATIENT'S SKIN. CATHETER NEEDS TO BE CHANGED ALMOST 4 TIMES A DAY. NO CLINICAL CONSEQUENCE OBSERVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16903795 ↗
  • MalfunctionMay 9, 2023Fluid/Blood Leak; Insufficient Flow or Under Infusion; Fluid/Blood Leak; Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THE DEVICE WAS LEAKING ON CATHETER CONNECTION AND MEDICATION WAS NOT BEING DELIVERED PROPERLY. NO MEDICAL OR SURGICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16894228 ↗
  • MalfunctionMay 8, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PATIENT OBSERVED A LEAKAGE AT THE ADAPTER BETWEEN THE TUBING AND THE CATHETER, AFTER A FEW HOURS OF INSERTION. THERE WAS NO PATIENT, OR CLINICIAN INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16892591 ↗
  • MalfunctionMay 8, 2023Defective Component; Defective Component

    IT WAS REPORTED THAT THE CAP ON THE KIT IS CROOKED, SO INSULIN LEAKED OUT INSTEAD OF INTO THE SUBCUTANEOUS FAT TISSUE. THIS OCCURRED AT NIGHT AFTER THE USER HAD TO CHANGE A CATHETER DUE TO KINKING RESULTING IN HIGH BLOOD GLUCOSE, SO THIS SEVERE HYPERGLYCEMIA CONTINUED FOR SEVERAL HOURS. THE USER WAS ABLE TO CORRECT IT BUT STATED A LESS EXPERIENCED USER WOULD PROBABLY HAVE HAD TO GO TO A HOSPITAL OR EMERGENCY ROOM. NO MEDICAL OR SURGICAL INTERVENTION WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 16892125 ↗
  • MalfunctionMay 8, 2023Defective Component; Defective Component

    IT WAS REPORTED THAT THE CATHETER WAS KINKING. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16892046 ↗
  • MalfunctionMay 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT CATHETER COULD NOT BE FIXED AND WAS PULLED OUT AGAIN WHEN THE INSERTER WAS DETACHED. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16889939 ↗
  • MalfunctionMay 8, 2023Leak/Splash; Patient Device Interaction Problem; Leak/Splash; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING INSERTION OF CATHETER, THE CANNULA DID NOT GO THROUGH THE SKIN AND THE NURSE DID NOT NOTICE AT THE MOMENT. THIS HAS HAPPENED ON SEVERAL OCCASIONS WITH EITHER LEAKAGE OR THE NEEDLE NOT RETRACTING, AND ANOTHER REPORT OF THE WHITE CAP OF THE CATHETER BEING BROKEN OFF. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16888877 ↗
  • MalfunctionMay 8, 2023Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED, THAT 12 CATHETERS WERE NOT TOLERATED AND HAD TO BE REMOVED PREMATURELY. ALLERGIC REACTION / INFECTION OF THE SETTING SITE. NO MEDICAL OR SURGICAL INTERVENTION WAS REPORTED.

    Patient problems: Local Reaction

    MAUDE report 16888236 ↗
  • MalfunctionApr 14, 2023Device Alarm System; Device Alarm System

    IT WAS REPORTED THAT THE DEVICE ALARMED A DISCONNECTION OF THE INFUSION LINE DURING THE NIGHT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16745882 ↗
  • MalfunctionOct 31, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE WHILE REMOVING THE NEEDLE, SHORT TUBE WAS NO LONGER PALPABLE. IT COULD NOT EVEN BE FELT AT THE PATIENT ANYMORE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15700395 ↗
  • MalfunctionOct 27, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE WHILE REMOVING THE NEEDLE, SHORT TUBE WAS NO LONGER PALPABLE. IT COULD NOT EVEN BE FELT AT THE PATIENT ANYMORE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15680161 ↗
  • MalfunctionOct 27, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE WHILE REMOVING THE NEEDLE, SHORT TUBE WAS NO LONGER PALPABLE. IT COULD NOT EVEN BE FELT AT THE PATIENT ANYMORE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15680102 ↗
  • MalfunctionSep 7, 2022Disconnection; Disconnection

    IT WAS REPORTED THAT THE TUBE BECAME DISCONNECTED FROM THE DISC. THE NURSES DO HOURLY CHECK OF THE TUBING AND SITE SO THE PROBLEM WAS CAUGHT EARLY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15369422 ↗
  • MalfunctionSep 7, 2022Material Perforation; Material Perforation

    IT WAS REPORTED THAT THE CLIENT HAD HIGH BLOOD GLUCOSE LEVELS WITH MULTIPLE CATHETERS AND WHEN THE NEEDLE WAS TRIGGERED, IT PERFORATED THE CATHETER. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Hyperglycemia

    MAUDE report 15369109 ↗
  • MalfunctionSep 5, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THERE WAS A LEAKAGE FROM TUBING OBSERVED DURING INFUSION BELOW THE SCREW THREAD. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15354224 ↗
  • MalfunctionApr 4, 2022Fluid/Blood Leak; Fluid/Blood Leak

    INFORMATION WAS RECEIVED REGARDING A CADD CLEO INFUSION SET BEING USED TO INFUSE APOMORPHINE. IT WAS REPORTED THAT THE PATIENT OBSERVED A LEAKAGE AT THE ADAPTER BETWEEN THE TUBING AND THE CATHETER, AFTER A FEW HOURS OF INSERTION. THERE WAS NO PATIENT, OR CLINICIAN INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14003226 ↗
  • InjuryJun 24, 2020Loss of or Failure to Bond; Infusion or Flow Problem; Loss of or Failure to Bond; Infusion or Flow Problem

    INFORMATION RECEIVED A SMITH MEDICAL CLEO INFUSION SET HAD SEVERAL MALFUNCTIONS WHERE IT CAUSED CONSUMER TO HAVE COMPLICATIONS ASSOCIATED WITH DIABETES. FIRST NOTED THE SPRING DID NOT WORK. SECONDLY, THE GLUE BETWEEN THE PLASTIC TAPE AND SOLID PART OF DEVICE WAS NOT STICKING; AFTER ONE HOUR SEPARATED. THE REPORT STATES LISTED ABOVE MALFUNCTIONS CAUSED HER TO HAVE HYPERGLYCEMIA AND CLOSE MONITORING OF BLOOD GLUCOSE LEVELS,CHECKING FOR KETONES IN URINE, WHICH SUBSEQUENTIALLY CAUSED INTERVENTIONS OF INSULIN SLIDING SCALE INJECTIONS WITH AIDE FROM HOSPITAL NURSE. ONE OTHER OCCASION TH TUBE SEPARATED FROM THE ADHESIVE. THE CONSUMER PROVIDED PICTURES OF THE MALFUNCTIONS. NO OTHER LONG TERM ADVERSE EVENTS REPORTED. HYPERGLYCEMIA CAN CAUSE DIABETIC KETOACIDOSIS. INTERVENTIONS WERE REQUIRED TO PRECLUDE SERIOUS INJURY FOR RISK OF KIDNEY FAILURE, HEART AND MOST COMMON CEREBRAL EDEMA. CRITICAL CARE WOULD NEED TO BE IMPLEMENTED TO STABILIZE THE PATIENT.

    Patient problems: Hyperglycemia

    MAUDE report 10189838 ↗
  • InjuryNov 15, 2019Loss of or Failure to Bond; Infusion or Flow Problem; Loss of or Failure to Bond; Infusion or Flow Problem

    INFORMATION RECEIVED A SMITH MEDICAL CLEO INFUSION SET HAD SEVERAL MALFUNCTIONS WHERE IT CAUSED CONSUMER TO HAVE COMPLICATIONS ASSOCIATED WITH DIABETES. FIRST NOTED THE SPRING DID NOT WORK. SECONDLY ,THE GLUE BETWEEN THE PLASTIC TAPE AND SOLID PART OF DEVICE WAS NOT STICKING; AFTER ONE HOUR SEPARATED. THE REPORT STATES LISTED ABOVE MALFUNCTIONS CAUSED HER TO HAVE HYPERGLYCEMIA AND CLOSE MONITORING OF BLOOD GLUCOSE LEVELS,CHECKING FOR KETONES IN URINE, WHICH SUBSEQUENTIALLY CAUSED INTERVENTIONS OF INSULIN SLIDING SCALE INJECTIONS WITH AIDE FROM HOSPITAL NURSE. ONE OTHER OCCASION TH TUBE SEPARATED FROM THE ADHESIVE. THE CONSUMER PROVIDED PICTURES OF THE MALFUNCTIONS. NO OTHER LONG TERM ADVERSE EVENTS REPORTED. HYPERGLYCEMIA CAN CAUSE DIABETIC KETOACIDOSIS. INTERVENTIONS WERE REQUIRED TO PRECLUDE SERIOUS INJURY FOR RISK OF KIDNEY FAILURE, HEART AND MOST COMMON CEREBRAL EDEMA. CRITICAL CARE WOULD NEED TO BE IMPLEMENTED TO STABILIZE THE PATIENT.

    Patient problems: Hyperglycemia

    MAUDE report 9329667 ↗
  • InjuryNov 14, 2019Loss of or Failure to Bond; Infusion or Flow Problem; Loss of or Failure to Bond; Infusion or Flow Problem

    INFORMATION RECEIVED A SMITH MEDICAL CLEO INFUSION SET HAD SEVERAL MALFUNCTIONS WHERE IT CAUSED CONSUMER TO HAVE COMPLICATIONS ASSOCIATED WITH DIABETES. FIRST NOTED THE SPRING DID NOT WORK. SECONDLY ,THE GLUE BETWEEN THE PLASTIC TAPE AND SOLID PART OF DEVICE WAS NOT STICKING; AFTER ONE HOUR SEPARATED. THE REPORT STATES LISTED ABOVE MALFUNCTIONS CAUSED HER TO HAVE HYPERGLYCEMIA AND CLOSE MONITORING OF BLOOD GLUCOSE LEVELS,CHECKING FOR KETONES IN URINE, WHICH SUBSEQUENTIALLY CAUSED INTERVENTIONS OF INSULIN SLIDING SCALE INJECTIONS WITH AIDE FROM HOSPITAL NURSE. ONE OTHER OCCASION TH TUBE SEPARATED FROM THE ADHESIVE. THE CONSUMER PROVIDED PICTURES OF THE MALFUNCTIONS. NO OTHER LONG TERM ADVERSE EVENTS REPORTED. HYPERGLYCEMIA CAN CAUSE DIABETIC KETOACIDOSIS. INTERVENTIONS WERE REQUIRED TO PRECLUDE SERIOUS INJURY FOR RISK OF KIDNEY FAILURE, HEART AND MOST COMMON CEREBRAL EDEMA. CRITICAL CARE WOULD NEED TO BE IMPLEMENTED TO STABILIZE THE PATIENT.

    Patient problems: Hyperglycemia

    MAUDE report 9325446 ↗
  • InjuryNov 14, 2019Loss of or Failure to Bond; Infusion or Flow Problem; Loss of or Failure to Bond; Infusion or Flow Problem

    INFORMATION RECEIVED A SMITH MEDICAL CLEO INFUSION SET HAD SEVERAL MALFUNCTIONS WHERE IT CAUSED CONSUMER TO HAVE COMPLICATIONS ASSOCIATED WITH DIABETES. FIRST NOTED THE SPRING DID NOT WORK. SECONDLY ,THE GLUE BETWEEN THE PLASTIC TAPE AND SOLID PART OF DEVICE WAS NOT STICKING; AFTER ONE HOUR SEPARATED. THE REPORT STATES LISTED ABOVE MALFUNCTIONS CAUSED HER TO HAVE HYPERGLYCEMIA AND CLOSE MONITORING OF BLOOD GLUCOSE LEVELS,CHECKING FOR KETONES IN URINE, WHICH SUBSEQUENTIALLY CAUSED INTERVENTIONS OF INSULIN SLIDING SCALE INJECTIONS WITH AIDE FROM HOSPITAL NURSE. ONE OTHER OCCASION TH TUBE SEPARATED FROM THE ADHESIVE. THE CONSUMER PROVIDED PICTURES OF THE MALFUNCTIONS. NO OTHER LONG TERM ADVERSE EVENTS REPORTED. HYPERGLYCEMIA CAN CAUSE DIABETIC KETOACIDOSIS. INTERVENTIONS WERE REQUIRED TO PRECLUDE SERIOUS INJURY FOR RISK OF KIDNEY FAILURE, HEART AND MOST COMMON CEREBRAL EDEMA. CRITICAL CARE WOULD NEED TO BE IMPLEMENTED TO STABILIZE THE PATIENT.

    Patient problems: Hyperglycemia

    MAUDE report 9325296 ↗
  • MalfunctionNov 14, 2019Loss of or Failure to Bond; Infusion or Flow Problem; Loss of or Failure to Bond; Infusion or Flow Problem

    INFORMATION RECEIVED A SMITH MEDICAL CLEO INFUSION SET HAD SEVERAL MALFUNCTIONS WHERE IT CAUSED CONSUMER TO HAVE COMPLICATIONS ASSOCIATED WITH DIABETES. FIRST NOTED THE SPRING DID NOT WORK. SECONDLY ,THE GLUE BETWEEN THE PLASTIC TAPE AND SOLID PART OF DEVICE WAS NOT STICKING; AFTER ONE HOUR SEPARATED. THE REPORT STATES LISTED ABOVE MALFUNCTIONS CAUSED HER TO HAVE HYPERGLYCEMIA AND CLOSE MONITORING OF BLOOD GLUCOSE LEVELS,CHECKING FOR KETONES IN URINE, WHICH SUBSEQUENTIALLY CAUSED INTERVENTIONS OF INSULIN SLIDING SCALE INJECTIONS WITH AIDE FROM HOSPITAL NURSE. ONE OTHER OCCASION TH TUBE SEPARATED FROM THE ADHESIVE. THE CONSUMER PROVIDED PICTURES OF THE MALFUNCTIONS. NO OTHER LONG TERM ADVERSE EVENTS REPORTED. HYPERGLYCEMIA CAN CAUSE DIABETIC KETOACIDOSIS. INTERVENTIONS WERE REQURIED TO PRECLUDE SERIOUS INJURY FOR RISK OF KIDNEY FAILURE, HEART AND MOST COMMON CEREBRAL EDEMA. CRITICAL CARE WOULD NEED TO BE IMPLEMETED TO STABALIZE THE PATIENT.

    Patient problems: Hyperglycemia

    MAUDE report 9322658 ↗
  • MalfunctionApr 20, 2018Break; Break

    INFORMATION WAS RECEIVED INDICATING THAT FOUR DAYS PRIOR THE CANNULA TO A SMITHS MEDICAL CLEO® 90 INFUSION SET BROKE OFF UNDER THE PATIENT'S SKIN. THE PATIENT REPORTS SEEKING MEDICAL ATTENTION TWO TIMES AND HAS BEEN PRESCRIBED ANTIBIOTICS. COMPLAINTS OF THE SITE BEING REDDENED AND SCABBED OVER. REPORTED THAT IF THE CANNULA DOES NOT COME OUT ON ITS OWN, THE PATIENT WILL HAVE TO GO TO THE EMERGENCY ROOM TO HAVE IT REMOVED. NO FURTHER ADVERSE EFFECTS ARE REPORTED.

    Patient problems: Foreign Body In Patient; Unspecified Infection

    MAUDE report 7447375 ↗
  • MalfunctionSep 27, 2017Mechanical Problem; Mechanical Problem

    IT WAS REPORTING THAT DURING PLACEMENT OF A CLEO® 90 INFUSION SET, THE NEEDLE DIDN'T RETRACT INTO THE INSERTER, INSTEAD IT REMAINED STUCK IN THE CANNULA. NO INJURY WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6898043 ↗

Other Set, administration, intravascular devices