Skip to content

ARROW IPN037427Recalls & reported problems

Teleflex Incorporated · ARROW(R) PICC powered by ARROW(R) VPS TipTracker(TM) Stylet · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 30801902125379), received between Oct 10, 2023 and Dec 22, 2023 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

7 other recalls cover devices cleared under the same FDA submission (K160925); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Difficult or Delayed Separation2
  • Fluid/Blood Leak1
  • Material Puncture/Hole1
  • Material Separation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT: AFTER THE PEEL AWAY SHEATH WAS PLACED, THE RN WAS UNABLE TO REMOVE THE PEEL AWAY SHEATH WITHOUT RETRACING THE SHEATH A BIT. ONCE RETRACTED , THE GUIDEWIRE WAS REMOVED AND NOTED TO BE BENT. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18381936 ↗
  • MalfunctionDec 19, 2023Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    IT WAS REPORTED THAT: WE HAD A PATIENT COME INTO THE ER WITH A LEAKING PICC AND AFTER ASSESSMENT IT WAS NOTED TO HAVE A PIN HOLE IN THE EXTENSION BETWEEN THE HUB OF THE STABILIZATION PART OF THE PICC WHERE THE CLAMP IS. WE HAD TO SCHEDULE THE PATIENT TO HAVE THE LINE REPLACED FOR CONTINUED TREATMENT. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18357975 ↗
  • MalfunctionOct 10, 2023Break; Break

    IT WAS REPORTED THE GLIDE-THRU SHEATH WAS DIFFICULT TO REMOVE FROM THE PATIENT. THE PEEL TABS WOULD NOT PEEL APART, AND ONE TAB ENDED UP BREAKING OFF. THERE WAS NO HARM TO THE PATIENT. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17905322 ↗
  • MalfunctionOct 10, 2023Break; Difficult or Delayed Separation; Break; Difficult or Delayed Separation

    IT WAS REPORTED THE CLINICIAN HAD DIFFICULTY PEELING THE SHEATH APART. ONE TAB BROKE OFF AFTER PULLING SO HARD. NO PATIENT HARM WAS REPORTED AND THE PICC WAS SUCCESSFULLY PLACED. THE SHEATH WAS REMOVED FROM THE PATIENT IN ITS ENTIRETY. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17904628 ↗
  • MalfunctionOct 10, 2023Break; Material Separation; Difficult or Delayed Separation; Break; Material Separation; Difficult or Delayed Separation

    IT WAS REPORTED THE CLINICIAN HAD DIFFICULTY PEELING THE SHEATH APART. ONE TAB BROKE OFF AFTER PULLING SO HARD. NO PATIENT HARM WAS REPORTED AND THE PICC WAS SUCCESSFULLY PLACED. THE SHEATH WAS REMOVED FROM THE PATIENT IN ITS ENTIRETY. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17904579 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160925. One submission often covers many models, so check the product and lot details in each record.

  • Z-0004-2024 · Aug 10, 2023 · Open, Classified
    Arrow ErgoPack Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC pre-loaded with Arrow VPS TipTracker Stylet, REF CDC-45563-TTS; catheter, intravascular, therapeutic, lo
  • Z-2382-2018 · May 17, 2018 · Terminated
    VPS Rhythm ECG Accessory Pack with Johans Adapter, REF RHY-100-ECGJ Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG
  • Z-2384-2018 · May 17, 2018 · Terminated
    VPS TipTracker Stylet Accessory, REF RHY-177-TTSAP Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wav
  • Z-2383-2018 · May 17, 2018 · Terminated
    VPS Rhythm ECG Accessory Pack, REF RHY-100-ECGO Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave c
  • Z-2027-2018 · Apr 11, 2018 · Terminated
    VPS Rhythm ECG Accessory Pack ; Product codes: RHY-100-ECGO
  • Z-2028-2018 · Apr 11, 2018 · Terminated
    VPS TipTracker Stylet Accessory; Product codes: RHY-177-TTSAP
  • Z-2026-2018 · Apr 11, 2018 · Terminated
    VPS Rhythm ECG Accessory Pack with Johans Adapter; Product codes: RHY-100-ECGJ

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page