ARROW IPN037427Recalls & reported problems
Teleflex Incorporated · ARROW(R) PICC powered by ARROW(R) VPS TipTracker(TM) Stylet · Class 2 · Implant
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 30801902125379), received between Oct 10, 2023 and Dec 22, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
7 other recalls cover devices cleared under the same FDA submission (K160925); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Difficult or Delayed Separation2
- Fluid/Blood Leak1
- Material Puncture/Hole1
- Material Separation1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT: AFTER THE PEEL AWAY SHEATH WAS PLACED, THE RN WAS UNABLE TO REMOVE THE PEEL AWAY SHEATH WITHOUT RETRACING THE SHEATH A BIT. ONCE RETRACTED , THE GUIDEWIRE WAS REMOVED AND NOTED TO BE BENT. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18381936 ↗MalfunctionDec 19, 2023Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT: WE HAD A PATIENT COME INTO THE ER WITH A LEAKING PICC AND AFTER ASSESSMENT IT WAS NOTED TO HAVE A PIN HOLE IN THE EXTENSION BETWEEN THE HUB OF THE STABILIZATION PART OF THE PICC WHERE THE CLAMP IS. WE HAD TO SCHEDULE THE PATIENT TO HAVE THE LINE REPLACED FOR CONTINUED TREATMENT. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18357975 ↗MalfunctionOct 10, 2023Break; Break
IT WAS REPORTED THE GLIDE-THRU SHEATH WAS DIFFICULT TO REMOVE FROM THE PATIENT. THE PEEL TABS WOULD NOT PEEL APART, AND ONE TAB ENDED UP BREAKING OFF. THERE WAS NO HARM TO THE PATIENT. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17905322 ↗MalfunctionOct 10, 2023Break; Difficult or Delayed Separation; Break; Difficult or Delayed Separation
IT WAS REPORTED THE CLINICIAN HAD DIFFICULTY PEELING THE SHEATH APART. ONE TAB BROKE OFF AFTER PULLING SO HARD. NO PATIENT HARM WAS REPORTED AND THE PICC WAS SUCCESSFULLY PLACED. THE SHEATH WAS REMOVED FROM THE PATIENT IN ITS ENTIRETY. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17904628 ↗MalfunctionOct 10, 2023Break; Material Separation; Difficult or Delayed Separation; Break; Material Separation; Difficult or Delayed Separation
IT WAS REPORTED THE CLINICIAN HAD DIFFICULTY PEELING THE SHEATH APART. ONE TAB BROKE OFF AFTER PULLING SO HARD. NO PATIENT HARM WAS REPORTED AND THE PICC WAS SUCCESSFULLY PLACED. THE SHEATH WAS REMOVED FROM THE PATIENT IN ITS ENTIRETY. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17904579 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160925. One submission often covers many models, so check the product and lot details in each record.
- Z-0004-2024 · Aug 10, 2023 · Open, ClassifiedArrow ErgoPack Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC pre-loaded with Arrow VPS TipTracker Stylet, REF CDC-45563-TTS; catheter, intravascular, therapeutic, lo
- Z-2382-2018 · May 17, 2018 · TerminatedVPS Rhythm ECG Accessory Pack with Johans Adapter, REF RHY-100-ECGJ Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG
- Z-2384-2018 · May 17, 2018 · TerminatedVPS TipTracker Stylet Accessory, REF RHY-177-TTSAP Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wav
- Z-2383-2018 · May 17, 2018 · TerminatedVPS Rhythm ECG Accessory Pack, REF RHY-100-ECGO Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave c
- Z-2027-2018 · Apr 11, 2018 · TerminatedVPS Rhythm ECG Accessory Pack ; Product codes: RHY-100-ECGO
- Z-2028-2018 · Apr 11, 2018 · TerminatedVPS TipTracker Stylet Accessory; Product codes: RHY-177-TTSAP
- Z-2026-2018 · Apr 11, 2018 · TerminatedVPS Rhythm ECG Accessory Pack with Johans Adapter; Product codes: RHY-100-ECGJ
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page