ARROW IPN037428Recalls & reported problems
Teleflex Incorporated · ARROW(R) PICC powered by ARROW(R) VPS TipTracker(TM) Stylet · Class 2 · Implant
All models of ARROW (1494)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 30801902125386), received between Nov 15, 2023 and Nov 28, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K160925); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove1
- Material Deformation1
- Material Separation1
- Material Twisted/Bent1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 28, 2023Separation Problem; Separation Problem
IT WAS REPORTED THAT: WHEN REMOVING PEEL AWAY SHEATH DURING PICC INSERTION, THE TAB BROKE OFF ONE SIDE OF THE SHEATH. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18215126 ↗MalfunctionNov 20, 2023Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT "RIGHT BASILIC VEIN EASILY ACCESSED, GUIDEWIRE WITH NO NOTED DEFECTS, THREADED WITHOUT DIFFICULTY THROUGH ACCESS NEEDLE. AFTER PLACING PEEL AWAY SHEATH, UNABLE TO REMOVE GUIDEWIRE. AFTER MANIPULATION AND PULLING BACK OF THE INTRODUCER AND REMOVING INNER CANNULA WAS ABLE TO REMOVE GUIDEWIRE AND NOTED LOOPING REMAINED IN THE WIRE". AS A RESULT, THE CATHETER WAS REMOVED AND A 2ND CATHETER WAS INSERTED IN THE SAME INSERTION SITE. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18175733 ↗MalfunctionNov 15, 2023Material Separation; Material Deformation; Material Separation; Material Deformation
IT WAS REPORTED THAT: THE TIP TRACKER STYLET SEPARATED FROM THE STEREOJACK CONNECTOR FOLLOWING UNSUCCESSFUL PICC PLACEMENT. WIRE IS NOTED TO HAVE A DEFORMITY AND WAS VERY DIFFICULT TO REMOVE FROM THE PICC PRIOR TO SEPARATION. ADDITIONAL INFORMATION: THE ISSUE WAS IDENTIFIED DURING USE. NO MEDICAL INTERVENTION WAS NEEDED AND THE PATIENT DID NOT EXPERIENCE ANY SIGNS OR SYMPTOMS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18140412 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160925. One submission often covers many models, so check the product and lot details in each record.
- Z-0004-2024 · Aug 10, 2023 · Open, ClassifiedArrow ErgoPack Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC pre-loaded with Arrow VPS TipTracker Stylet, REF CDC-45563-TTS; catheter, intravascular, therapeutic, lo
- Z-2382-2018 · May 17, 2018 · TerminatedVPS Rhythm ECG Accessory Pack with Johans Adapter, REF RHY-100-ECGJ Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG
- Z-2384-2018 · May 17, 2018 · TerminatedVPS TipTracker Stylet Accessory, REF RHY-177-TTSAP Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wav
- Z-2383-2018 · May 17, 2018 · TerminatedVPS Rhythm ECG Accessory Pack, REF RHY-100-ECGO Product Usage: To provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave c
- Z-2028-2018 · Apr 11, 2018 · TerminatedVPS TipTracker Stylet Accessory; Product codes: RHY-177-TTSAP
- Z-2027-2018 · Apr 11, 2018 · TerminatedVPS Rhythm ECG Accessory Pack ; Product codes: RHY-100-ECGO
- Z-2026-2018 · Apr 11, 2018 · TerminatedVPS Rhythm ECG Accessory Pack with Johans Adapter; Product codes: RHY-100-ECGJ
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page