ARROW IPN040163Recalls & reported problems
Teleflex Incorporated · FlexTip Plus(R) Epidural Catheterization Kit · Class 2
All models of ARROW (1488)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 30801902126420), received between Aug 12, 2022 and Oct 1, 2024 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
9 other recalls cover devices cleared under the same FDA submission (K140110); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Obstruction of Flow2
- Corroded1
- Material Separation1
- Patient-Device Incompatibility1
- Premature Separation1
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- Inadequate Pain Relief1
- Purulent Discharge1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryOct 1, 2024Material Separation; Material Separation
IT WAS REPORTED THAT: "POSTPARTUM PATIENT WHO RECEIVED EVALUATION FROM AN ANESTHESIOLOGIST , WHO PERFORMED THE EPIDURAL CATHETER REMOVAL ON WHICH PRESENTS RUPTURE, LEAVING A FRAGMENT OF THE CATHETER IN THE LEFT FORAMINAL LOCATION IN L3/L4 IN CONTACT WITH THE L3 EMERGING ROOT, ACCORDING TO A LUMBAR SPINE CT SCAN REPORT." AWAITING ADDITIONAL INFORMATION FROM THE CUSTOMER.
Patient problems: Foreign Body In Patient
MAUDE report 20342759 ↗InjuryMar 29, 2024Corroded; Patient-Device Incompatibility; Corroded; Patient-Device Incompatibility
IT WAS REPORTED "THE CLINICIAN OBSERVED A PURULENT DISCHARGE AT THE PUNCTURE SITE. HE THEN REMOVED THE COMPRESSES AND NOTICED THAT THE FILTER WAS CORRODED, ORANGE-STAINED AND POROUS." ADDITIONAL INFORMATION REPORTS THERE WAS NO SYSTEMIC INFECTION AND THE PATIENT WAS NOT SEPTIC PRIOR TO INSERTING THE CATHETER. THE FILTER WAS CHANGED AND ANTIBIOTICS WAS ADMINISTERED. THE PATIENTS CURRENT CONDITION IS REPORTED AS "STABLE".
Patient problems: Purulent Discharge
MAUDE report 19004580 ↗MalfunctionOct 28, 2022Premature Separation; Premature Separation
IT WAS REPORTED THAT ON (B)(6) 2022 IN THE OBSTETRIC BLOCK DURING CHILDBIRTH, THE WINGCLIP SNAPLOCK SEPARATED FROM THE FILTER. THE CONSEQUENCE WAS A LEAK OF THE TREATMENT AND THE NON-ADMINISTRATION OF THE PATIENT'S EPIDURAL ANALGESIA. THIS INTERRUPTION LASTED 1 HOUR WHILST CLEANING AND REPLACING THE CATHETER. IT CAUSE A RISK FOR THE PATIENT. THE PATIENT'S CONDITION WAS REPORTED AS FINE.
Patient problems: Inadequate Pain Relief
MAUDE report 15692043 ↗MalfunctionSep 24, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT: INCIDENT OCCURRED ON (B)(6) 2022 WHILST GIVING BIRTH PROCEDURE. THE CONNECTION OF THE CATHETER WAS CHECKED PRIOR TO INSERT IN THE PATIENT. ALL WAS FINE. DURING INSERTION IN THE PATIENT A LIQUID LEAK WAS OBSERVED. THERE WAS NO CONSEQUENCE FOR THE PATIENT BUT ANOTHER INSERTION OF CATHETER WAS PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15477269 ↗MalfunctionSep 6, 2022Fluid/Blood Leak; Fluid/Blood Leak
WE FACED ISSUES WITH THE PLUNGER OF THE SYRINGES OF THE KIT, THERE WERE SALINE LEAKS.
Patient problems: Insufficient Information
MAUDE report 15362956 ↗MalfunctionAug 12, 2022Obstruction of Flow; Obstruction of Flow
THIS COMPLAINT COVERS EVENTS WITH UNKNOWN DETAILS. IT WAS REPORTED THAT IN ON (B)(6) 2022 A DOCTOR OBSERVED A DIFFICULTY DURING THE INJECTION OF PERIDURAL. SIMILAR EVENTS OF THIS ISSUE OF DIFFICULTIES TO INJECT THE PRODUCT HAVE BEEN REPORTED IN THE DEPARTMENT BUT WITHOUT BEING REPORTED WITH ANY AVAILABLE DETAILS. IT IS UNKNOWN WHAT PART OF THE KIT IS THE CAUSE OF THE INCIDENT. WE CANNOT PROVIDE THIS INFORMATION.
Patient problems: Insufficient Information
MAUDE report 15218770 ↗MalfunctionAug 12, 2022Obstruction of Flow; Obstruction of Flow
ON (B)(6) 2022 BEFORE GIVING BIRTH, THE DOCTOR OBSERVED A DIFFICULTY DURING THE INJECTION OF PERIDURAL. INDEED, THE PERIDURAL ANESTHETIC KIT DOES NOT INJECT THE PRODUCT OR WE HAVE TO APPLY FORCE ON THE PLUNGER. CLINICAL CONSEQUENCES: BECAUSE OF THIS, THE DOCTOR DECIDED TO CHANGE THE ENTIRETY OF THE KIT AND TO PUNCTURE THE PATIENT AGAIN. THERE WERE NO OTHER MEDICAL CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15218693 ↗
FDA clearance and classification
Cleared through 510(k) K140110 on Mar 27, 2015 (decision: Substantially Equivalent). Submitted by Teleflex, Inc. as “Flextip Plus Epidural Catheter;Flexblock Continous Peripheral Nerve Block Catheter Kit/Set”. The FDA review took 436 days. FDA record ↗
The FDA regulates product code CAZ (anesthesia conduction kit) under 21 CFR 868.5140 as a Class II device. 21 CFR 868.5140 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K140110. One submission often covers many models, so check the product and lot details in each record.
- Z-2371-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION K
- Z-1913-2026 · Mar 11, 2026 · Open, ClassifiedFLEXTIP PLUS EPIDURAL CATHETER;FLEXBLOCK CONTINOUS PERIPHERAL NERVE BLOCK CATHETER KIT/SET REF ASK-05502-BSM UDI code: (01)10801902209853(17)270831(11)251210(10)33F25K0409 The Arr
- Z-2288-2020 · May 8, 2020 · Open, ClassifiedArrow FlexTip Plus Epidural Kit with NRFit Connector, catheter system, Product Codes (US) OT19TAGSK OT19TAGSKA OT19TAPSK OT19TCGLK Product Codes (OUS) AT-05501-NRF OT19HBPSS UM-05
- Z-1174-2020 · Jan 15, 2020 · TerminatedFlexTip Plus Epidural Catheterization Kit, Product Code ASK-02220-SRH
- Z-1177-2020 · Jan 15, 2020 · TerminatedFlexTip Plus Epidural Catheterization Kit, Product Code AK-05500
- Z-1172-2020 · Jan 15, 2020 · TerminatedFlexTip Plus Epidural Catheterization Kit, Product Code AK-05501
- Z-1236-2020 · Dec 20, 2019 · TerminatedArrow FlexTip Plus(R) Epidural Catheterization Kit ASK-05400-CA1
- Z-1238-2020 · Dec 20, 2019 · TerminatedArrow FlexTip Plus(R) Epidural Catheterization Kit ASK-05500-CAN
- Z-1237-2020 · Dec 20, 2019 · TerminatedArrow FlexTip Plus(R) Epidural Catheterization Kit ASK-17019-MSC
Common questions about the ARROW IPN040163
Has the ARROW IPN040163 been recalled?
No FDA recall record in our data names this exact device identifier (DI 30801902126420), based on the FDA recall file dated Nov 27, 2026. However, 9 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ARROW IPN040163?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Aug 12, 2022 and Oct 1, 2024. In those reports, the most common specific device problems are “Fluid/Blood Leak” (2), “Obstruction of Flow” (2) and “Corroded” (1); the most common patient problems are “Foreign Body In Patient” (1), “Inadequate Pain Relief” (1) and “Purulent Discharge” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARROW IPN040163?
None of the reports are classified as deaths, 2 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARROW IPN040163 FDA approved?
It was cleared through 510(k) K140110 on Mar 27, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ARROW IPN040163 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the ARROW IPN040163, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN040163). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Anesthesia conduction kit devices
- BD Luer-Lok, BD Glaspak 405671Becton, Dickinson And Company · Anesthesia conduction kit1 recall68 reports
- ARROW IPN922678Teleflex Incorporated · Anesthesia conduction kit2 recalls0 reports
- BD Luer-Lok, BD Glaspak 405672Becton, Dickinson And Company · Anesthesia conduction kit1 recall26 reports
- BD Luer-Lok, BD Glaspak 405673Becton, Dickinson And Company · Anesthesia conduction kit1 recall10 reports
- BD Luer-Lok, BD Glaspak 405637Becton, Dickinson And Company · Anesthesia conduction kit1 recall7 reports
- BD Luer-Lok, BD Glaspak 405707Becton, Dickinson And Company · Anesthesia conduction kit1 recall4 reports
- BD Luer-Lok, BD Glaspak 405609Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
- BD Luer-Lok, BD Glaspak 405735Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page