Skip to content

Medline SYR101010Recalls & reported problems

Medline Industries, Inc. · SYRINGE, LUER LOCK, 1ML · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 30884389596313), received between Mar 7, 2024 and Apr 4, 2024 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 7 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2024Class IIOpen, Classified

    It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.

    Z-1937-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem7
  • Contamination2
  • Detachment of Device or Device Component2
  • Connection Problem1
  • Fluid/Blood Leak1
  • Material Too Soft/Flexible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 4, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE SYRINGE PLUNGER WAS NOT "PUSHING THROUGH SMOOTHLY AND THEY SEEM TO GET STUCK."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19047532 ↗
  • MalfunctionApr 4, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE SYRINGE PLUNGER WAS "STICKING" AND "DOES NOT MOVE SMOOTHLY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19046452 ↗
  • MalfunctionApr 4, 2024Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BLACK STOPPER OF THE SYRINGE PLUNGER WAS "LOOSE AND UN-USABLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19040699 ↗
  • MalfunctionApr 1, 2024Material Too Soft/Flexible; Material Too Soft/Flexible

    IT WAS REPORTED THAT THE SYRINGE PLUNGER WAS "FLEXIBLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19016884 ↗
  • MalfunctionApr 1, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT "THE RUBBER STOPPER IS GETTING CAUGHT" WITHIN THE SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19016846 ↗
  • MalfunctionMar 29, 2024Connection Problem; Connection Problem

    IT WAS REPORTED THAT NEEDLES "ARE NOT PROPERLY LOCKING" WITH THE SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19008737 ↗
  • MalfunctionMar 29, 2024Contamination; Contamination

    IT WAS REPORTED THAT "FIBER/HAIR" WAS NOTED WITHIN THE SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19008735 ↗
  • MalfunctionMar 25, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE SYRINGE PLUNGER WAS HARD TO PUSH AND THAT IT "FELT STUCK."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18975566 ↗
  • MalfunctionMar 24, 2024Contamination; Contamination

    IT WAS REPORTED THAT HAIR WAS NOTED WITHIN THE SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18966404 ↗
  • MalfunctionMar 18, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT SYRINGE PLUNGER BECOMES "STUCK" WHEN INJECTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18922277 ↗
  • MalfunctionMar 18, 2024Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SYRINGE PLUNGER "LOCKS UP" AND THAT NEEDLE DETACHMENT OCCURRED WHEN APPLYING TOO MUCH PRESSURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18922155 ↗
  • MalfunctionMar 7, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT "WHILE IN USE, THE PLUNGER WAS STICKING TO THE SYRINGE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18861588 ↗

Other Syringe, piston devices

All Syringe, piston devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page