Shiley 86445Recalls & reported problems
Covidien LP · Hi-Lo Oral/Nasal Tracheal Tube Cuffed,Intermediate, Murphy Eye · Class 2
All models of Shiley (387)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 30884522005719), received between Oct 8, 2019 and Feb 4, 2025 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem5
- Material Split, Cut or Torn2
- Appropriate Device Problem Term/Code Not Available1
- Disconnection1
Patient problems
- Extubate2
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Patient Involvement1
- Unintended Extubation1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2025Use of Device Problem; Material Split, Cut or Torn; Use of Device Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT A CIRCUIT DISCONNECTION ALARMED ON THE VENTILATOR AND THE BIOMEDICAL ENGINEER CAME AND PERFORMED THE CORRESPONDING REVISIONS OF THE EQUIPMENT, WALL, OUTLETS, FILTERS AND PATIENT CIRCUIT. ALTHOUGH NO FLAWS WAS DETECTED, THE DECISION WAS MADE TO CHANGED THE VENTILATOR. WHEN THE PATIENT WAS RECONNECTED TO THE VENTILATOR, THE PROBLEM PERSISTED AND THE CONDITION OF THE ENDOTRACHEAL TUBE WAS CHECKED. IT WAS DETECTED THAT IT WAS DEFLATED AND ONCE INFLATED THE EQUIPMENT STOPPED ALARMING. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21309618 ↗MalfunctionAug 22, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
"ACCORDING TO THE REPORTER, DURING USE, A CIRCUIT DISCONNECTION ALARM WAS HEARD ON THE VENTILATOR. CORRESPONDING CHECKS WERE PERFORMED ON THE VENTILATOR, WALL OUTLETS, FILTERS, AND PATIENT CIRCUIT, AND NO FLAWS WERE DETECTED. WHEN THE PATIENT WAS REINSTALLED ON NEW EQUIPMENT, THE PROBLEM PERSISTED. THE CONDITION OF THE ENDOTRACHEAL TUBE WAS CHECKED AND IT WAS FOUND TO BE DEFLATED. THE TUBE WAS INFLATED, AND THE EQUIPMENT STOPPED ALARMING. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20039992 ↗MalfunctionMar 4, 2024Use of Device Problem; Appropriate Device Problem Term/Code Not Available; Use of Device Problem; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, FOR THE 4.5-MM TUBE SIZE, THE BALLOON WAS SMALL ENOUGH TO PASS BACK THROUGH THE VOCAL CORDS AFTER DEFLATING. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Insufficient Information
MAUDE report 18830962 ↗MalfunctionDec 14, 2023Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE PILOT BALLOON FELL OFF AT THE END OF THE ENDOTRACHEAL TUBE AFTER INTUBATION. THE PATIENT WAS EXTUBATED, AND A NEW ENDOTRACHEAL TUBE WAS INSERTED.
Patient problems: Unintended Extubation
MAUDE report 18332353 ↗InjuryJul 22, 2020Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, WHEN INTUBATING A PATIENT IT WAS OBSERVED THAT THE DEVICE¿S PIPE TUBE 4.5 HAD INFLATION/DEFLATION ISSUE. IT WAS EXCHANGED FOR ANOTHER AND IT HAPPENED THE SAME. PATIENT WAS REINTUBATED WITH 5.0 TUBE.
Patient problems: Extubate
MAUDE report 10310953 ↗InjuryJul 22, 2020Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, WHEN INTUBATING A PATIENT IT WAS OBSERVED THAT THE DEVICE¿S PIPE TUBE 4.5 HAD INFLATION/DEFLATION ISSUE. IT WAS EXCHANGED FOR ANOTHER AND IT HAPPENED THE SAME. PATIENT WAS REINTUBATED WITH 5.0 TUBE.
Patient problems: Extubate
MAUDE report 10308223 ↗MalfunctionOct 8, 2019Disconnection; Disconnection
ACCORDING TO THE REPORTER, THE UNIT'S TUBE PARTS WERE APART. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9165681 ↗
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.