Skip to content

Shiley 86445Recalls & reported problems

Covidien LP · Hi-Lo Oral/Nasal Tracheal Tube Cuffed,Intermediate, Murphy Eye · Class 2

All models of Shiley (387)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 30884522005719), received between Oct 8, 2019 and Feb 4, 2025 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem5
  • Material Split, Cut or Torn2
  • Appropriate Device Problem Term/Code Not Available1
  • Disconnection1

Patient problems

  • Extubate2
  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • No Patient Involvement1
  • Unintended Extubation1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2025Use of Device Problem; Material Split, Cut or Torn; Use of Device Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT A CIRCUIT DISCONNECTION ALARMED ON THE VENTILATOR AND THE BIOMEDICAL ENGINEER CAME AND PERFORMED THE CORRESPONDING REVISIONS OF THE EQUIPMENT, WALL, OUTLETS, FILTERS AND PATIENT CIRCUIT. ALTHOUGH NO FLAWS WAS DETECTED, THE DECISION WAS MADE TO CHANGED THE VENTILATOR. WHEN THE PATIENT WAS RECONNECTED TO THE VENTILATOR, THE PROBLEM PERSISTED AND THE CONDITION OF THE ENDOTRACHEAL TUBE WAS CHECKED. IT WAS DETECTED THAT IT WAS DEFLATED AND ONCE INFLATED THE EQUIPMENT STOPPED ALARMING. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21309618 ↗
  • MalfunctionAug 22, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    "ACCORDING TO THE REPORTER, DURING USE, A CIRCUIT DISCONNECTION ALARM WAS HEARD ON THE VENTILATOR. CORRESPONDING CHECKS WERE PERFORMED ON THE VENTILATOR, WALL OUTLETS, FILTERS, AND PATIENT CIRCUIT, AND NO FLAWS WERE DETECTED. WHEN THE PATIENT WAS REINSTALLED ON NEW EQUIPMENT, THE PROBLEM PERSISTED. THE CONDITION OF THE ENDOTRACHEAL TUBE WAS CHECKED AND IT WAS FOUND TO BE DEFLATED. THE TUBE WAS INFLATED, AND THE EQUIPMENT STOPPED ALARMING. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20039992 ↗
  • MalfunctionMar 4, 2024Use of Device Problem; Appropriate Device Problem Term/Code Not Available; Use of Device Problem; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE REPORTER, FOR THE 4.5-MM TUBE SIZE, THE BALLOON WAS SMALL ENOUGH TO PASS BACK THROUGH THE VOCAL CORDS AFTER DEFLATING. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 18830962 ↗
  • MalfunctionDec 14, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE PILOT BALLOON FELL OFF AT THE END OF THE ENDOTRACHEAL TUBE AFTER INTUBATION. THE PATIENT WAS EXTUBATED, AND A NEW ENDOTRACHEAL TUBE WAS INSERTED.

    Patient problems: Unintended Extubation

    MAUDE report 18332353 ↗
  • InjuryJul 22, 2020Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, WHEN INTUBATING A PATIENT IT WAS OBSERVED THAT THE DEVICE¿S PIPE TUBE 4.5 HAD INFLATION/DEFLATION ISSUE. IT WAS EXCHANGED FOR ANOTHER AND IT HAPPENED THE SAME. PATIENT WAS REINTUBATED WITH 5.0 TUBE.

    Patient problems: Extubate

    MAUDE report 10310953 ↗
  • InjuryJul 22, 2020Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, WHEN INTUBATING A PATIENT IT WAS OBSERVED THAT THE DEVICE¿S PIPE TUBE 4.5 HAD INFLATION/DEFLATION ISSUE. IT WAS EXCHANGED FOR ANOTHER AND IT HAPPENED THE SAME. PATIENT WAS REINTUBATED WITH 5.0 TUBE.

    Patient problems: Extubate

    MAUDE report 10308223 ↗
  • MalfunctionOct 8, 2019Disconnection; Disconnection

    ACCORDING TO THE REPORTER, THE UNIT'S TUBE PARTS WERE APART. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9165681 ↗

Other Tube, tracheal (w/wo connector) devices

All Tube, tracheal (w/wo connector) devices →