Skip to content

Shiley 4.0 PEDRecalls & reported problems

Covidien LP · Pediatric Tracheostomy Tube Cuffless · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 30884522026714), received between Jun 28, 2018 and Feb 15, 2024 — classified as 6 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K945513); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem2
  • Defective Component2
  • Material Deformation2
  • Material Split, Cut or Torn2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Device Markings/Labelling Problem1
  • Device Packaging Compromised1

Patient problems

  • Injury4
  • No Patient Involvement3
  • Unintended Extubation3
  • Extubate2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 15, 2024Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE WAS CAUSING THE PATIENT TO COUGH AFTER JUST AN HOUR AND WAS REPLACED WITH A DIFFERENT TYPE OF TRACH.

    Patient problems: Unintended Extubation

    MAUDE report 18717850 ↗
  • MalfunctionFeb 23, 2023Defective Component; Connection Problem; Defective Component; Connection Problem

    ACCORDING TO THE REPORTER, THE CONNECTOR OF THE DEVICE WAS TOO RIGID AND COULD NOT BE CONNECTED TO THE CATHETER MOUNT DURING PLACEMENT WITH NO VISIBLE DAMAGED ON BOTH. THE PATIENT REQUIRED REPLACEMENT OF THE DEVICE.

    Patient problems: Unintended Extubation

    MAUDE report 16427855 ↗
  • MalfunctionJul 14, 2021Connection Problem; Connection Problem

    ACCORDING TO THE REPORTER, THE DEVICE WAS ATTACHED TO AN ARTIFICIAL NOSE (SPONTANEOUS) BUT THE CONNECTION TO THE CONNECTOR WAS STIFF AND HAD A DIFFICULTY TO DISCONNECT. THE PATIENT HAD A REPLACEMENT OF THE TUBE WITHOUT HARM.

    Patient problems: Unintended Extubation

    MAUDE report 12167927 ↗
  • MalfunctionMar 4, 2021Defective Component; Adverse Event Without Identified Device or Use Problem; Defective Component; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE PRODUCT WAS APPROXIMATELY 2MM SHORTER THAN SPECIFICATION. ANOTHER PRODUCT HAD A BENT CANNULA INTERNALLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11410828 ↗
  • MalfunctionSep 28, 2020Material Separation; Material Separation

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE TUBE WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10592324 ↗
  • MalfunctionOct 7, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    ACCORDING TO THE REPORTER, PRIOR TO USE, THERE WAS A STRONG RESISTANCE FELT WHEN REMOVING THE OBTURATOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9161157 ↗
  • InjuryNov 13, 2018Material Deformation; Material Deformation

    ACCORDING TO THE REPORTER, THE CANNULA'S WINGS HAD A SHARP EDGE THAT DAMAGED THE SKIN ON THE PATIENT¿S NECK.

    Patient problems: Injury

    MAUDE report 8065538 ↗
  • InjuryNov 13, 2018Material Deformation; Material Deformation

    ACCORDING TO THE REPORTER, THE CANNULA'S WINGS HAD A SHARP EDGE THAT DAMAGED THE SKIN ON THE PATIENT¿S NECK.

    Patient problems: Injury

    MAUDE report 8065532 ↗
  • InjuryOct 1, 2018Material Split, Cut or Torn; Material Split, Cut or Torn

    ACCORDING TO THE REPORTER, THE DEVICE HAD A DAMAGED FASTENING FLAP. THE CUSTOMER REPORTED THAT THE PATIENT PRESENTED EXTREME BRADYCARDIA.

    Patient problems: Injury

    MAUDE report 7922674 ↗
  • InjuryOct 1, 2018Material Split, Cut or Torn; Material Split, Cut or Torn

    ACCORDING TO THE REPORTER, THE DEVICE HAD A DAMAGED FASTENING FLAP. THE CUSTOMER REPORTED THAT THE PATIENT PRESENTED EXTREME BRADYCARDIA.

    Patient problems: Injury

    MAUDE report 7922620 ↗
  • InjuryAug 29, 2018Break; Break

    ACCORDING TO THE REPORTER, THE DEVICE HAD PHYSICAL DAMAGE AND THE FLANGE WAS BROKEN. IT WAS REPORTED THAT DURING NURSING PROCEDURE, THE PATIENT WAS RESTLESS AND MADE A SUDDEN MOVEMENT AND BROKE THE LEFT FIN OF THE CANNULA. THE CUSTOMER REPORTED THAT THE PATIENT REQUIRED RE-CANNULATION.

    Patient problems: Extubate

    MAUDE report 7828288 ↗
  • MalfunctionJul 6, 2018Material Discolored; Material Discolored

    ACCORDING TO THE REPORTER, PRIOR TO USE, TUBING WAS DISCOLORED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7666733 ↗
  • InjuryJun 28, 2018Device Markings/Labelling Problem; Device Packaging Compromised; Device Markings/Labelling Problem; Device Packaging Compromised

    ACCORDING TO THE REPORTER, THE DEVICE HAD WRONG TRACHEOSTOMY TUBE INSIDE THE BOX, THEY FOUND A SIZE 6 TRACHEOSTOMY TUBE INSIDE THE SIZE 4 TRACHEOSTOMY BOX. IT WAS INDICATED THAT THE UNIT WAS NOT STERILE BECAUSE THE BOX WAS NOT SEALED. IT WAS ALSO REPORTED THAT THE PATIENT HAD A SEVERE DISCOMFORT DUE TO THE LARGER SIZE OF THE TUBE THAN THE STOMA SITE. THE CUSTOMER REPORTED THAT THEY REPLACED THE TUBE FROM SIZE 6 TO SIZE TO SIZE 4.

    Patient problems: Extubate

    MAUDE report 7645839 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K945513. One submission often covers many models, so check the product and lot details in each record.

  • Z-0997-2009 · Jan 14, 2009 · Terminated
    Shiley PED3.0 Cuffless Pediatric Tracheostomy Tube, REF 3.0 PED, Sterile. Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Assembled in Mexico. The pr

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →