Nellcor DS100A-1Recalls & reported problems
Covidien LP · Adult SpO2 Sensor Reusable · Class 2
All models of Nellcor (64)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30884522042714), received between Apr 30, 2018 and Nov 11, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “High Readings” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K012891); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High Readings3
- Low Readings3
- Device Damaged Prior to Use1
- Incorrect Measurement1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 11, 2024Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SENSOR WAS NOT READING CORRECTLY. IT WAS TESTED WITH A SENSOR SIMULATOR AND THE MONITOR WAS READING SPO2 CORRECTLY. THE SENSOR WAS REPLACED WITH A NEW ONE TO RESOLVE THE FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20652566 ↗MalfunctionFeb 1, 2024Low Readings; Low Readings
ACCORDING TO THE REPORTER, THE DEVICE HAD LOW READINGS. THE SENSOR WAS REPLACED WITH A NEW ONE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18626739 ↗MalfunctionApr 19, 2021High Readings; High Readings
ACCORDING TO THE REPORTER, THE MONITOR WAS NOT WORKING PROPERLY AND WOULD HAVE HIGH READINGS OF BP AND SPO2 HEART RATE OF 300. THE SENSOR WAS REPLACED AS A PRECAUTIONARY MEASURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11688816 ↗MalfunctionMar 16, 2020Device Damaged Prior to Use; Low Readings; Device Damaged Prior to Use; Low Readings
ACCORDING TO THE REPORTER, THE DEVICE PARAMETER WAS BROKEN, STATING THAT THE MONITOR WAS NOT PUMPING. IT WAS ALSO INDICATED THAT THE SENSOR WAS SHOWING SPO2 READING OF 94% WITH 100% OXYGEN. THE SENSOR WAS REPLACED. THERE WAS NO ALLEGATION OF PATIENT DEATH OR SERIOUS INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9835969 ↗MalfunctionMay 2, 2019High Readings; Low Readings; High Readings; Low Readings
ACCORDING TO THE REPORTER, THE UNIT WAS GIVING HIGH PULSE RATE READINGS OF 200, HOWEVER WHEN THEY USED DIFFERENT DEVICE THE PATIENT'S PULSE RATE WAS 97. IT WAS ALSO REPORTED THE PRODUCT GAVE LOW SPO2 READINGS OF 50'S. NEVERTHELESS, WHEN TRIED WITH ANOTHER UNIT THE SPO2 RESULT WAS NORMAL. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8573725 ↗MalfunctionApr 30, 2018High Readings; High Readings
ACCORDING TO THE REPORTER, THE UNIT GAVE NORMAL PULSE RATE READING EXCEPT FOR ONE STUDENT WHICH GAVE A HIGH PULSE RATE READING OF 285. CUSTOMER INDICATED THERE WAS NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7472526 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K012891. One submission often covers many models, so check the product and lot details in each record.
- Z-0664-05 · Oct 18, 2004 · TerminatedNellcor brand OxiMax Pulse Oximeter, model number N-595 Made in Ireland; Distributed by Tyco Healthcare Group, LP, Nellcor Puritan Bennett, Inc., Pleasanton, CA 94588 U.S.A.
Other Oximeter devices
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy Trilogy 202 Ventilator, United StatesRespironics, Inc. · Oximeter2 deaths1,594 reports
- HAMILTON-G5 159001Hamilton Medical AG · Oximeter18 deaths1,440 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Model EMV+ Vent, Army, MRI 799-EMVP-04Zoll Medical Corporation · Oximeter1 death605 reports
- Trilogy Evo, O2, Japan JP2100X16BRespironics, Inc. · Oximeter4 recalls5 deaths342 reports
- Model EMV+ Vent, Army, COMMERCIAL 799-EMVP-01-01Zoll Medical Corporation · Oximeter1 death497 reports
- HAMILTON-G5 159002Hamilton Medical AG · Oximeter3 deaths451 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page