Skip to content

NA 50-20706Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · Bone Screws, MP, Cross-Pin · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 34546540602566), received between Mar 25, 2020 and Sep 8, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Appears to Trigger Rejection1
  • Loss of Osseointegration1

Patient problems

  • Pain2
  • Abscess1
  • Bacterial Infection1
  • Failure of Implant1
  • Inadequate Osseointegration1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 8, 2020Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT A PATIENT WITH A HISTORY OF SLEEP APNEA AND BLEEDING UNDERWENT A LEFORT 1/BSSO PROCEDURE ON (B)(6) 2020. PATIENT CAME IN FOR FOLLOW UP VISIT COMPLAINING OF SOME PAIN, AND SURGEON NOTICED A CROSS BITE. SURGEON HAD PATIENT SCANNED AND DETERMINED THAT HARDWARE MAY BE LOOSE. SURGEON REMOVED HARDWARE FROM PATIENT¿S RIGHT AND LEFT MANDIBLE, PUT PATIENT IN ARCH BARS TO REESTABLISH OCCLUSION AND FIXATED RIGHT AND LEFT MANDIBLE. PROCEDURE WAS COMPLETED WITH NO ISSUES.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain

    MAUDE report 10504420 ↗
  • InjuryMar 25, 2020Device Appears to Trigger Rejection; Adverse Event Without Identified Device or Use Problem; Device Appears to Trigger Rejection; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY A SURGEON THAT A PATIENT REQUIRED A REVISION SURGERY DUE TO EXPERIENCING PAIN AND HAVING PUS SECRETIONS APPROXIMATELY 4-5 WEEKS AFTER INITIAL IMPLANTATION WITH AN IMPLANT. THE ORIGINAL PROCEDURE WAS REPORTED AS COMPLETED SUCCESSFULLY WITHOUT A DELAY, BUT DURING THE REVISION SURGERY IT WAS FOUND THAT THE SCREWS HAD LOOSENED, CALCIFICATION FORMED NEAR THE PLATE, AND A PATIENT SWAB TESTED AS MRSA POSITIVE. THE PLATE WAS REMOVED.

    Patient problems: Abscess; Bacterial Infection; Pain

    MAUDE report 9877948 ↗

Other PLATE, BONE devices

All PLATE, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page