Skip to content

MARS TREATMENT KIT TYPE 1116/1 - X-MARS US 800541Recalls & reported problems

Gambro AB · The MARS Treatment Kit Type 1116/1 X-MARS is a set of disposable items to be used with the MARS Monitor 1TC in combination with the Prismaflex device (produced by Gambro Lundia AB) for MARS therapy. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 37332414107728), received between Oct 9, 2018 and Oct 9, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 21, 2017Class IITerminated

    Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.

    Z-0411-2018 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 9, 2018Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK OCCURRED ON A MARS KIT, AFTER SET-UP AND DURING PRIMING. THE LOCATION OF THE LEAK WAS NOT IDENTIFIED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 7948494 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page