MARS Treatment Kit, Use only with MARS(R) Monitor 1TC… recallZ-0411-2018 · Baxter Healthcare Corporation · initiated Dec 21, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm initiated their recall on 12/21/2017 by letter. The recall notice stated the following actions to be taken by the customers: "1. Locate and remove all affected product from your facility. The product code and lot number can be found on the individual MARS Treatment Kit box. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you purchased this product directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270- 5457 or scanning and e-mailing it to [email protected], even if you do not have any inventory."

Recall details

Recall number
Z-0411-2018
Status
Terminated
Initiated by the company
Dec 21, 2017
Recalling firm
Baxter Healthcare Corporation
Quantity
186 units
Product code
FLD
FDA submissions
K133807
Distribution
US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.

Product description

MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541

Code information (models, lots, serial numbers)

UDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.