Skip to content

Medline SPT-2208CSRecalls & reported problems

Medline Industries, Inc. · GOWN POLY W/BREATH SLEEVES XL · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 40080196029755), received between Sep 25, 2026 and Sep 25, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Nonstandard Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 29, 2023Open, Classified

    Surgical gowns were manufactured with the wrong sleeve.

    Z-0711-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Nonstandard Device

    It was reported that "their technician who is cleaning scopes in the scope room has complained of getting wet through the gown"; "Specifically, it¿s the seam on the sleeves"; "She¿s getting wet on the sleeves"; "She is not getting wet on the trunk."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897371 ↗

Other GOWN, SURGICAL devices

All GOWN, SURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page