Medline MDS89697Recalls & reported problems
Medline Industries, Inc. · RAIL,HALF SZ,CLAMP ON,SM GAP BAR SPACE · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 40080196746027), received between Aug 24, 2026 and Sep 28, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Activation Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Problem2
- Unstable2
- Separation Problem1
- Sharp Edges1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Sharp Edges
It was reported that "the edges of her handrails being sharp, ¿knife railings¿.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917037 ↗MalfunctionSep 22, 2026Unstable
It was reported that the clamp will not tighten.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50857763 ↗MalfunctionAug 27, 2026Unstable
It was reported that side rails wiggle alot.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50476923 ↗MalfunctionAug 27, 2026Activation Problem; Separation Problem
It was reported that "PER PT SHE HAS HOSPITAL BED W/RAILS ON BOTH SIDES. ONE RAIL FELL OFF AND OTHER IS NOT WORKING."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50476520 ↗MalfunctionAug 24, 2026Activation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50415435 ↗
FDA clearance and classification
The FDA regulates product code FNL (bed, ac-powered adjustable hospital) under 21 CFR 880.5100 as a Class II device. 21 CFR 880.5100 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Medline MDS89697
Has the Medline MDS89697 been recalled?
No FDA recall record in our data names this exact device identifier (DI 40080196746027), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Medline MDS89697?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Aug 24, 2026 and Sep 28, 2026. In those reports, the most common specific device problems are “Activation Problem” (2), “Unstable” (2) and “Separation Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Medline MDS89697?
None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Medline MDS89697 FDA approved?
Devices of this type (product code FNL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Medline Industries, Inc.
Who makes the Medline MDS89697, and is it still sold?
The FDA's device database lists Medline Industries, Inc. as the company responsible for it (brand Medline, model MDS89697). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,831 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,320 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,149 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths333 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death330 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths354 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL299 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page