ARROW IPN046311 (REF JC-05400-B)Recalls & reported problems
Teleflex Incorporated · FlexTip Plus(R) Epidural Catheterization Kit · Class 2
All models of ARROW (1488)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Arrow FlexTip Plus epidural kit is used to place a catheter for epidural anesthesia. The 42 reports mostly describe fluid leaking and syringe and needle parts not fitting together, so the resistance check could not be done, with a few reports of catheters breaking or being hard to advance. The reports mention no deaths and 3 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 40801902029674), received between Jan 28, 2021 and Oct 23, 2025 — classified as 3 injuries, 39 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (23 reports).
4 other recalls cover devices cleared under the same FDA submission (K103658); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak23
- Burst Container or Vessel5
- Physical Resistance/Sticking3
- Incomplete or Inadequate Connection2
- Material Separation2
- Premature Separation2
- Break1
- Component or Accessory Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions37
- Foreign Body In Patient2
- Needle Stick/Puncture2
- Pain2
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 23, 2025Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT: "THE SYRINGE TIP DOES NOT FIT INTO THE PUNCTURE NEEDLE, MAKING IT IMPOSSIBLE TO CHECK THE LOSS OF RESISTANCE. THIS MEANS THAT THE SET CANNOT BE USED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23367119 ↗MalfunctionOct 23, 2025Component or Accessory Incompatibility; Incomplete or Inadequate Connection; Component or Accessory Incompatibility; Incomplete or Inadequate Connection
IT WAS REPORTED THAT "THE SYRINGE TIP DOES NOT FIT INTO THE PUNCTURE NEEDLE, MAKING IT IMPOSSIBLE TO CHECK THE LOSS OF RESISTANCE. WE HAD TO REMOVE THE NEEDLE AND PUNCTURE AGAIN. THIS MEANS THAT THE SET CANNOT BE USED. THE PATIENT WAS REPORTED AS FINE POST THE PROCEDURE."
Patient problems: Needle Stick/Puncture
MAUDE report 23366200 ↗MalfunctionOct 15, 2025Material Twisted/Bent; Material Too Soft/Flexible; Material Twisted/Bent; Material Too Soft/Flexible
IT WAS REPORTED THAT: "WE HAVE RECEIVED FEEDBACK FROM OUR ANAESTHETISTS REGARDING THE ARMOURED EPIDURAL SETS CONCERNING THE DIAMETER OF THE NEEDLE, WHICH SLIDES TOO EASILY (SMALLER DIAMETER?) AND BENDS MORE EASILY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23299094 ↗MalfunctionApr 9, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT " NO LOSS OF RESISTANCE WITH THE SYRINGE PLUNGER. I CAN CONFIRM THE OCCURRENCE OF A BREACH WITH CLINICAL MANIFESTATIONS SUCH AS HEADACHE AND NECK PAIN. PATIENT DISCHARGED ON (B)(6) 2025. REQUIRED 3 BLOOD PATCHES ON (B)(6)."
Patient problems: Pain
MAUDE report 21805123 ↗MalfunctionMar 3, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT: "THERE WAS NO LOSS OF RESISTANCE ON THE LOR WHEN REACHING 10 ML WHILE SEARCHING FOR EPIDURAL SPACE CAUSING DURA MATTER BREACH. AFTER CHECKING THE LOR ON STOCK, IT WAS OBSERVED THAT THERE WAS NO LOSS OF RESISTANCE ON THE LOR WHEN REACHING THE 10 ML. THIS CAUSED DELAY IN TREATMENT AND INCREASE COST DUE TO REPLACEMENT." ASSOCIATED COMPLAINTS 3006425876-2025-00180, 3006425876-2025-00181.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21504860 ↗MalfunctionMar 3, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT: "THERE WAS NO LOSS OF RESISTANCE ON THE LOR WHEN REACHING 10 ML WHILE SEARCHING FOR EPIDURAL SPACE CAUSING DURA MATTER BREACH. AFTER CHECKING THE LOR ON STOCK, IT WAS OBSERVED THAT THERE WAS NO LOSS OF RESISTANCE ON THE LOR WHEN REACHING THE 10 ML. THIS CAUSED DELAY IN TREATMENT AND INCREASE COST DUE TO REPLACEMENT." ASSOCIATED COMPLAINTS 3006425876-2025-00180, 3006425876-2025-00181 AND 3006425876-2025-00179.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21504783 ↗InjurySep 18, 2024Material Separation; Material Separation
IT WAS REPORTED THAT: "INCIDENT OCCURRED ON (B)(6 ) 2024. IN THE DELIVERY ROOM, PATIENT: FEMALE 35-YEAR-OLD. INITIALLY INSERTION WITHOUT DIFFICULTY: LOSS OF RESISTANCE AT L2-L3 7 CM FROM THE SKIN. CATHETER INSERTED INTO THE EPIDURAL SPACE WITHOUT DIFFICULTY. WHEN THE CATHETER WAS MOVED IN ORDER TO ATTACH TO THE SKIN, A SUDDEN AND SPONTANEOUS CUT/BREAK OF THE PLASTIC SURROUNDING THE METAL FRAME OF THE CATHETER OCCURRED. THERE WAS A 13 CM SECTION OF PLASTIC LEFT IN THE PATIENT'S EPIDURAL SPACE. PATIENT HAS HAD AN MRI TO CONFIRM, THE SECTION OF CATHETER REMAINS IN THE PATIENT. ACCORDING TO NEURO-SURGEON 'NO SENSORIMOTOR DEFICIT, NO ABSOLUTE URGENCY TO REMOVE THE PORTION OF THE EPIDURAL CATHETER. A CT SCAN OR MRI TO BE PERFORMED AND AWAITING NEUROSURGICAL OPINION."
Patient problems: Foreign Body In Patient
MAUDE report 20244645 ↗MalfunctionAug 22, 2024Difficult to Advance; Difficult to Advance
THERE WAS A PROBLEM ADVANCING THE CATHETER IN THE EPIDURAL SPACE. DEVICE WAS CHANGED. SENSATION OF RESISTANCE AGAIN, THEN SUDDENLY THE CATHETER ADVANCED, CREATING A BREACH / PUNCTURE IN THE DURA MATER. CONSEQUENCE: NEW EPIDURAL CATHETER INSERTED AND BLOOD PATCH APPLIED ON 47 HOURS.
Patient problems: Needle Stick/Puncture
MAUDE report 20040940 ↗MalfunctionApr 14, 2024Difficult to Remove; Stretched; Device Dislodged or Dislocated; Difficult to Remove; Stretched; Device Dislodged or Dislocated
IT WAS REPORTED "THE NURSE WAS UNABLE TO REMOVE THE CATHETER. THE NURSE CALLED THE ANESTHETIST. THE ANESTHETIST OBSERVED THE DIFFICULTY ENCOUNTERED ON THE LAST FEW CENTIMETERS, WHICH HE EVALUATED AS DIFFICULT SINCE THE DEVICE IS FLEXIBLE." ADDITIONAL INFORMATION REPORTED "THE TIP OF THE CATHETER WAS SEVERELY STRETCHED OVER THE LAST THREE TO FOUR CENTIMETRES. NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPROTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19103105 ↗MalfunctionApr 10, 2024Material Separation; Material Separation
IT WAS REPORTED "DURING REMOVAL FROM THE PATIENT, THE CATHETER BROKE. THE REMAINING CATHETER COULD BE COMPLETELY REMOVED. THE PATIENT WAS NOT HARMED." NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19077985 ↗MalfunctionApr 1, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED " THE EPIDURAL CATHETER WAS FOUND WITH A LEAK OF FLUID AT THE LEVEL OF THE PUNCTURE SITE.THE EPIDURAL THREAD WAS REMOVED AND IT WAS FOUND THAT THE FLUID WAS LEAKING FROM THE TIP OF THE CATHETER (WHICH IS NORMAL) BUT ALSO LEAKING ABOUT 10CM FROM THE TIP, WHICH IS NOT NORMAL." NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITON IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19013791 ↗MalfunctionFeb 29, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "LEAK AT LEVEL OF THE CATHETER TUBING WHEN TESTING THE DEVICE ONCE INSERTED". ADDITIONAL INFORMATION RECEIVED STATES "THERE WAS NO CONSEQUENCE FOR THE PATIENT". NO MEDICAL INTERVENTION WAS NECESSARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18810972 ↗InjuryFeb 29, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT "DURING THE REMOVAL OF THE CATHETER, THE CATHETER BODY BROKE, AND A PIECE WAS LEFT IN THE FEMALE PATIENT. CLINICAL CONSEQUENCES: BACK PAIN AND THEN MORE IMPORTANT PAIN IN THE RIGHT LEG. NO MOTOR DYSFUNCTION OR SENSITIVE DEFICIT. IMAGING DONE". ADDITIONAL INFORMATION STATES, "A CT SCAN WAS PERFORMED. REPORT: PRESENCE OF MATERIAL IN POSTERIOR EPIDURAL AND POSTEROLATERAL LEFT L2-L3. ON 01 FEB, THE PATIENT WAS TRANSFERRED TO THE MATERNITY UNIT FOR POST- PARTUM CARE ON DAY 1 FOR REMOVAL OF AN EPIDURAL ANALGESIA CATHETER BROKEN IN THE EPIDURAL SPACE. AS THE CLINICAL EXAMINATION WAS REASSURING, THE REMOVAL OF THE CATHETER UNDER GA WAS NON-URGENT SURGERY AND GA WAS CONTRAINDICATED DURING THE FIRST 5 DAYS POSTPARTUM. PATIENT WAS ADVISED TO RESCHEDULE AN APPOINTMENT FROM DAY 6 POSTPARTUM, IF SHE SO WISHED. THE PATIENT'S CONDITION AT DISCHARGE WAS NORMAL".
Patient problems: Foreign Body In Patient; Pain
MAUDE report 18810822 ↗MalfunctionNov 14, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "THE DEVICE WAS NOT USED BECAUSE THE SYRINGE PLUNGER WAS LEAKING." ADDITIONAL INFORMATION STATES THAT THIS OCCURRED "PRIOR TO USE ON A PATIENT DURING INSPECTION/FUNCTIONAL TESTING".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15790155 ↗MalfunctionDec 29, 2021Premature Separation; Premature Separation
THE PLUNGER OF THE LOR SYRINGE SLIDES OUT OF THE SYRINGE BARREL WITHOUT ANY FORCE BEING APPLIED. THIS ISSUE WAS ALREADY NOTED ON APPROXIMATELY 20 DEVICES IN A PERIOD FROM OCTOBER TO DECEMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13112908 ↗MalfunctionDec 29, 2021Premature Separation; Premature Separation
THE PLUNGER OF THE LOR SYRINGE SLIDES OUT OF THE SYRINGE BARREL WITHOUT ANY FORCE BEING APPLIED. THIS ISSUE WAS ALREADY NOTED ON APPROXIMATELY 20 DEVICES IN A PERIOD FROM OCTOBER TO DECEMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13112893 ↗MalfunctionDec 21, 2021Material Fragmentation; Material Fragmentation
THE OBSTETRICS DEPARTMENT HAD TO DEAL WITH AN INCIDENT WITH AN EPIDURAL CATHETER LAST NIGHT (B)(6) 2021. WHEN REMOVING THE CATHETER AFTER CHILDBIRTH, A PART OF IT, APPROXIMATELY 5 OR 6 CM REMAINED IN THE PATIENT'S BODY. DOCTOR IS WONDERING IF IT IS POSSIBLE TO KEEP THIS PART IN THE BODY OR IF A NEUROSURGERY INTERVENTION BE PLANNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13052869 ↗MalfunctionOct 4, 2021Fluid/Blood Leak; Fluid/Blood Leak
THE PLUNGER OF THE SYRINGE USED TO SEARCH FOR THE EPIDURAL SPACE WAS LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12569072 ↗MalfunctionOct 4, 2021Fluid/Blood Leak; Fluid/Blood Leak
THE PLUNGER OF THE SYRINGE USED TO SEARCH FOR THE EPIDURAL SPACE WAS LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12569058 ↗MalfunctionSep 1, 2021Fluid/Blood Leak; Fluid/Blood Leak
WE FACED A SALINE LEAK ISSUE WITH THE LOR SYRINGE. THE PLUNGER WAS LEAKING. THE DEVICE IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12407281 ↗MalfunctionAug 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
WE STILL OBSERVED LEAKS AT THE LEVEL OF THE PLUNGER OF THE LOR SYRINGE. THE LEAKS OCCURRED OVER THE LEVEL OF 6 ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12353058 ↗MalfunctionAug 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
WE STILL OBSERVED LEAKS AT THE LEVEL OF THE PLUNGER OF THE LOR SYRINGE. THE LEAKS OCCURRED OVER THE LEVEL OF 6 ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12353024 ↗MalfunctionAug 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
WE STILL OBSERVED LEAKS AT THE LEVEL OF THE PLUNGER OF THE LOR SYRINGE. THE LEAKS OCCURRED OVER THE LEVEL OF 6 ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12353019 ↗MalfunctionAug 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
WE STILL OBSERVED LEAKS AT THE LEVEL OF THE PLUNGER OF THE LOR SYRINGE. THE LEAKS OCCURRED OVER THE LEVEL OF 6 ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12353015 ↗MalfunctionAug 23, 2021Fluid/Blood Leak; Fluid/Blood Leak
WE STILL OBSERVED LEAKS AT THE LEVEL OF THE PLUNGER OF THE LOR SYRINGE. THE LEAKS OCCURED OVER THE LEVEL OF 6 ML. IT HAPPENED WITH SEVERAL SYRINGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12352978 ↗
FDA clearance and classification
Cleared through 510(k) K103658 on May 16, 2012 (decision: Substantially Equivalent). Submitted by Teleflex, Inc. as “Flextip Plus Closed Tip Epidural Catherter”. The FDA review took 518 days. FDA record ↗
The FDA regulates product code CAZ (anesthesia conduction kit) under 21 CFR 868.5140 as a Class II device. 21 CFR 868.5140 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K103658. One submission often covers many models, so check the product and lot details in each record.
- Z-2364-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS;
- Z-1179-2020 · Jan 15, 2020 · TerminatedFlexTip Plus Epidural Catheterization Kit, Product Code ASK-05400-CA1
- Z-1962-2018 · Apr 11, 2018 · TerminatedSnapLock" Catheter/Syringe Adapter; Product Code: AS-05500
- Z-2462-2015 · Jul 30, 2015 · TerminatedPICC Kits with Peel-Away Sheath Component; catheter, intravascular, therapeutic, ultrasound Product Usage: catheter, intravascular, therapeutic, ultrasound
Common questions about the ARROW IPN046311
Has the ARROW IPN046311 been recalled?
No FDA recall record in our data names this exact device identifier (DI 40801902029674), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ARROW IPN046311?
The FDA's MAUDE database lists 42 adverse event reports for this device, received between Jan 28, 2021 and Oct 23, 2025. In those reports, the most common specific device problems are “Fluid/Blood Leak” (23), “Burst Container or Vessel” (5) and “Physical Resistance/Sticking” (3); the most common patient problems are “Foreign Body In Patient” (2), “Needle Stick/Puncture” (2) and “Pain” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARROW IPN046311?
None of the reports are classified as deaths, 3 as injuries and 39 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARROW IPN046311 FDA approved?
It was cleared through 510(k) K103658 on May 16, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ARROW IPN046311 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the ARROW IPN046311, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN046311). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Anesthesia conduction kit devices
- BD Luer-Lok, BD Glaspak 405671Becton, Dickinson And Company · Anesthesia conduction kit1 recall68 reports
- ARROW IPN922678Teleflex Incorporated · Anesthesia conduction kit2 recalls0 reports
- BD Luer-Lok, BD Glaspak 405672Becton, Dickinson And Company · Anesthesia conduction kit1 recall26 reports
- BD Luer-Lok, BD Glaspak 405673Becton, Dickinson And Company · Anesthesia conduction kit1 recall10 reports
- BD Luer-Lok, BD Glaspak 405637Becton, Dickinson And Company · Anesthesia conduction kit1 recall7 reports
- BD Luer-Lok, BD Glaspak 405707Becton, Dickinson And Company · Anesthesia conduction kit1 recall4 reports
- BD Luer-Lok, BD Glaspak 405609Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
- BD Luer-Lok, BD Glaspak 405735Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page