Skip to content

Shiley 110875Recalls & reported problems

Covidien LP · Evac Oral Tracheal Tube Seal Guard, Murphy Eye · Class 2

All models of Shiley (387)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 40884521068224), received between Jun 17, 2019 and Jun 26, 2024 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K082520); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Leak/Splash2
  • Use of Device Problem2
  • Human-Device Interface Problem1

Patient problems

  • Unintended Extubation3
  • No Clinical Signs, Symptoms or Conditions2
  • No Patient Involvement2
  • Extubate1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2024Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, SHORTLY AFTER INTUBATION, THE NURSE NOTICED FLUID INSIDE THE ENDOTRACHEAL TUBE'S PILOT BALLOON, AND ABOUT 1/4 OF THE VOLUME WAS FILLED WITH FLUID DURING USE. THE CUFF PRESSURE WAS GOOD, AND THERE WAS NO OTHER DAMAGE NOTICED. IT WAS STATED THAT INITIALLY THE CUFF WAS FILLED WITH A NEW AND DRY SYRINGE, BUT IT WAS UNKNOWN IF THERE WAS LIQUID INSIDE THE CUFF TOO. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 19619991 ↗
  • MalfunctionFeb 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE REPORTER, THE ENDOTRACHEAL TUBE'S PILOT BALLOON AND INFLATION LINE HAD ACCUMULATED DROPLETS OF WATER DURING PATIENT USE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 18714656 ↗
  • MalfunctionFeb 1, 2022Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE REPORTER, WHEN THE AIR WAS REMOVED FROM THE CUFF OF THE DEVICE AT THE TIME OF EXTUBATION, A LARGE AMOUNT OF MOISTURE WAS DETECTED IN THE SYRINGE. THE PATIENT HAD A SUCCESSFUL REPLACEMENT OF THE TUBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13417809 ↗
  • MalfunctionNov 24, 2021Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE CUFF WAS UNABLE TO INJECT AIR OR REMOVED AIR SO A HOLE WAS MADE IN THE PILOT BALLOON TO DEFLATE. REPLACEMENT WAS DONE.

    Patient problems: Unintended Extubation

    MAUDE report 12868614 ↗
  • MalfunctionMar 11, 2021Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE REPORTER, THE WATER DROPLETS WERE GENERATED IN THE CUFF AND CUFF PRESSURE COULD NOT BE CONTROLLED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11462048 ↗
  • InjurySep 25, 2019Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, DURING USE, THERE WAS A CONDENSATION ON THE PILOT BALLOON AND 1.4 CC WAS COLLECTED AFTER A WEEK OF INTUBATION. WHEN THEY TRIED TO SUCTION AFTER 5 DAYS, 1 CC WAS ABLE TO BE SUCTION. THEY SUCTIONED 3.2 ML OF WATER IN THE CUFF ON JULY 12, 3 ML ON AUGUST 19, AND 0.5 ML ON AUGUST 22. THEY REPLACED IT WITH A CUSTOM-MADE TUBE.

    Patient problems: Extubate

    MAUDE report 9115591 ↗
  • MalfunctionJun 17, 2019Leak/Splash; Leak/Splash

    ACCORDING TO THE REPORTER, THE UNIT HAD CUFF INFLATION/DEFLATION ISSUE. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8706165 ↗
  • MalfunctionJun 17, 2019Leak/Splash; Leak/Splash

    ACCORDING TO THE REPORTER, THE UNIT HAD CUFF INFLATION/DEFLATION ISSUE. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8706164 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K082520. One submission often covers many models, so check the product and lot details in each record.

  • Z-1625-2016 · Apr 4, 2016 · Terminated
    Mallinckrodt Evac Tracheal Tubes Model #124170-MALLINCKRODT & 124175
  • Z-1626-2016 · Apr 4, 2016 · Terminated
    Launcher Guiding Catheter Model #LA6EBU30. Used in the coronary or peripheral vascular system..

Other Tube, tracheal (w/wo connector) devices

All Tube, tracheal (w/wo connector) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page