EmpowerCTA®+ Injector System 017285 (UDI 47630039300904)Recalls & reported problems

Bracco Injeneering SA · Composed of : 1 x Remote Control (Onyx) GTIN : 37630039300761 // Alternate : 1 x Remote Control GTIN : 37630039300471 1 x EmpowerSync® Power Supply GTIN : 37630039300839 1 x Injector GTIN : 37630039300891 · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
2
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 47630039300904), received between Nov 15, 2022 and Dec 28, 2023 — classified as 2 deaths, 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • Air Embolism3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6), 2023, (B)(4) BECAME AWARE THAT AFTER A COMPUTED TOMOGRAPHY (CT) SCAN OF THE ABDOMEN AND PELVIS WITH AN EMPOWERCTA+ INJECTION SYSTEM, SERIAL NUMBER (B)(6), THE TECHNOLOGIST NOTICED AN UNKNOWN AMOUNT OF ISOVUE 300 CONTRAST ON THE FLOOR FROM A LEAKING CONNECTION BETWEEN THE INJECTOR TUBING AND THE INTRAVENOUS (IV) LINE. AN UNKNOWN AMOUNT OF AIR WAS LATER OBSERVED IN THE IMAGING IN THE PATIENT'S LEFT VENTRICLE. ON (B)(6) 2023, (B)(4) BECAME AWARE THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL DUE TO THE VISUALIZATION OF AIR ON THE CT IMAGES. THE PATIENT WAS REPORTED TO HAVE RECOVERED ON (B)(6) 2023.

    Patient problems: Air Embolism

    MAUDE report 18414463 ↗
  • DeathJan 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A COMPUTED TOMOGRAPHY (CT) SCAN OF THE CHEST FOR PULMONARY EMBOLISM, APPROXIMATELY 30 CC OF AIR WAS INJECTED INTO THE PATIENT. THE PATIENT EXPERIENCED NAUSEA, VOMITING, CHEST PAIN, BRADYCARDIA, AND A DECLINE IN MENTAL STATUS. AN ECHOCARDIOGRAM (ECG) REVEALED ELEVATED ST SEGMENTS AND THE PATIENT'S BLOOD PRESSURE READINGS WERE AS FOLLOWS: 64/221, 45/121, AND 123/74. THERE WAS EVIDENCE OF MYOCARDIAL INFARCTION, SPECIFICALLY CARDIAC BIOMARKER ELEVATION. THE AIR INJECTION WAS FIRST NOTICED THIRTY (30) MINUTES AFTER THE EXAM WAS COMPLETED. THE PATIENT EXPIRED ON (B)(6) 2022.

    Patient problems: Air Embolism

    MAUDE report 16178260 ↗
  • DeathNov 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A COMPUTED TOMOGRAPHY (CT) SCAN OF THE CHEST FOR PULMONARY EMBOLISM, APPROXIMATELY 30 CC OF AIR WAS INJECTED INTO THE PATIENT. THE PATIENT EXPERIENCED NAUSEA, VOMITING, CHEST PAIN, BRADYCARDIA, AND A DECLINE IN MENTAL STATUS. AN ECHOCARDIOGRAM (ECG) REVEALED ELEVATED ST SEGMENTS AND THE PATIENT'S BLOOD PRESSURE READINGS WERE AS FOLLOWS: 64/221, 45/121, AND 123/74. THERE WAS EVIDENCE OF MYOCARDIAL INFARCTION, SPECIFICALLY CARDIAC BIOMARKER ELEVATION. THE AIR INJECTION WAS FIRST NOTICED THIRTY (30) MINUTES AFTER THE EXAM WAS COMPLETED. THE PATIENT EXPIRED ON (B)(6) 2022.

    Patient problems: Air Embolism

    MAUDE report 15798223 ↗

Other Injector, Contrast Medium, Automatic devices

All Injector, Contrast Medium, Automatic devices →