EmpowerCTA®+ Injector System 017285 (UDI 47630039300904)Recalls & reported problems
Bracco Injeneering SA · Composed of : 1 x Remote Control (Onyx) GTIN : 37630039300761 // Alternate : 1 x Remote Control GTIN : 37630039300471 1 x EmpowerSync® Power Supply GTIN : 37630039300839 1 x Injector GTIN : 37630039300891 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 2
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 47630039300904), received between Nov 15, 2022 and Dec 28, 2023 — classified as 2 deaths, 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
Patient problems
- Air Embolism3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6), 2023, (B)(4) BECAME AWARE THAT AFTER A COMPUTED TOMOGRAPHY (CT) SCAN OF THE ABDOMEN AND PELVIS WITH AN EMPOWERCTA+ INJECTION SYSTEM, SERIAL NUMBER (B)(6), THE TECHNOLOGIST NOTICED AN UNKNOWN AMOUNT OF ISOVUE 300 CONTRAST ON THE FLOOR FROM A LEAKING CONNECTION BETWEEN THE INJECTOR TUBING AND THE INTRAVENOUS (IV) LINE. AN UNKNOWN AMOUNT OF AIR WAS LATER OBSERVED IN THE IMAGING IN THE PATIENT'S LEFT VENTRICLE. ON (B)(6) 2023, (B)(4) BECAME AWARE THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL DUE TO THE VISUALIZATION OF AIR ON THE CT IMAGES. THE PATIENT WAS REPORTED TO HAVE RECOVERED ON (B)(6) 2023.
Patient problems: Air Embolism
MAUDE report 18414463 ↗DeathJan 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A COMPUTED TOMOGRAPHY (CT) SCAN OF THE CHEST FOR PULMONARY EMBOLISM, APPROXIMATELY 30 CC OF AIR WAS INJECTED INTO THE PATIENT. THE PATIENT EXPERIENCED NAUSEA, VOMITING, CHEST PAIN, BRADYCARDIA, AND A DECLINE IN MENTAL STATUS. AN ECHOCARDIOGRAM (ECG) REVEALED ELEVATED ST SEGMENTS AND THE PATIENT'S BLOOD PRESSURE READINGS WERE AS FOLLOWS: 64/221, 45/121, AND 123/74. THERE WAS EVIDENCE OF MYOCARDIAL INFARCTION, SPECIFICALLY CARDIAC BIOMARKER ELEVATION. THE AIR INJECTION WAS FIRST NOTICED THIRTY (30) MINUTES AFTER THE EXAM WAS COMPLETED. THE PATIENT EXPIRED ON (B)(6) 2022.
Patient problems: Air Embolism
MAUDE report 16178260 ↗DeathNov 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A COMPUTED TOMOGRAPHY (CT) SCAN OF THE CHEST FOR PULMONARY EMBOLISM, APPROXIMATELY 30 CC OF AIR WAS INJECTED INTO THE PATIENT. THE PATIENT EXPERIENCED NAUSEA, VOMITING, CHEST PAIN, BRADYCARDIA, AND A DECLINE IN MENTAL STATUS. AN ECHOCARDIOGRAM (ECG) REVEALED ELEVATED ST SEGMENTS AND THE PATIENT'S BLOOD PRESSURE READINGS WERE AS FOLLOWS: 64/221, 45/121, AND 123/74. THERE WAS EVIDENCE OF MYOCARDIAL INFARCTION, SPECIFICALLY CARDIAC BIOMARKER ELEVATION. THE AIR INJECTION WAS FIRST NOTICED THIRTY (30) MINUTES AFTER THE EXAM WAS COMPLETED. THE PATIENT EXPIRED ON (B)(6) 2022.
Patient problems: Air Embolism
MAUDE report 15798223 ↗
Other Injector, Contrast Medium, Automatic devices
- MEDRAD® Stellant FLEX Syringe Kit FLEXD-150-SPKBayer Medical Care Inc. · Injector, Contrast Medium, Automatic1 recall0 reports
- MEDRAD® Stellant FLEX CT Injection System with Certegra® WorkstationBayer Medical Care Inc. · Injector, Contrast Medium, Automatic19 reports
- MEDRAD® Stellant FLEX CT Injection System with Certegra® Workstation (Upgrade)Bayer Medical Care Inc. · Injector, Contrast Medium, Automatic16 reports
- LF 844005Liebel-Flarsheim Company LLC · Injector, Contrast Medium, Automatic6 reports
- FastLoad™ MR Syringe Pack 017356Bracco Injeneering SA · Injector, Contrast Medium, Automatic4 reports
- LF 844007Liebel-Flarsheim Company LLC · Injector, Contrast Medium, Automatic2 reports
- UL_INJECT XD 8000Ulrich GmbH & Co. KG · Injector, contrast medium, automatic1 death1 report
- EmpowerMR® Injector System without Remote Control 017377Bracco Injeneering SA · Injector, Contrast Medium, Automatic1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page