Skip to content

Bair Hugger™ 36000Recalls & reported problems

3m Company · 3M™ Bair Hugger™ Temperature Monitoring Sensor, 25 per case, 36000 · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 50707387775808), received between Nov 2, 2017 and Nov 2, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Computer Software Problem1
  • Device Issue1
  • Device Operational Issue1

Patient problems

  • No Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 2, 2017Computer Software Problem; Device Issue; Device Operational Issue; Computer Software Problem; Device Issue; Device Operational Issue

    A CUSTOMER REPORTED THAT THE SENSOR ON THE PRODUCT USED WAS CONSISTENTLY READING 3 DEGREES LOWER ON THE FOREHEAD (RIGHT SIDE AND LEFT SIDE) COMPARED WITH ORAL TEMPERATURE TAKEN. THE CUSTOMER OBTAINED ANOTHER SENSOR FROM A DIFFERENT LOT AND PLACED IT ON THE RIGHT FOREHEAD AND ITS READING WAS CONSISTENT WITH THE ORAL TEMPERATURE TAKEN. THIS ISSUE HAD BEEN EXPERIENCED FROM VARIOUS CLINICIANS FROM WITHIN THE SAME FACILITY. THE LOT NUMBER WAS PROVIDED APPLICABLE TO THE PRODUCT OF CONCERN. A RECALL HAS BEEN INITIATED WHICH INCLUDES THE SAME LOT NUMBER REGARDING THIS REPORTED EVENT AND NOT PROVIDING A CORRECT TEMPERATURE. THE ROOT CAUSE HAS BEEN DETERMINED AND CORRECTION ACTION HAS BEEN IMPLEMENTED.

    Patient problems: No Information

    MAUDE report 6997558 ↗