ARROW IPN036739Recalls & reported problems

Teleflex Incorporated · ARROW(R) JACC with Chlorag+ard(R) Technology Jugular Axillo-subclavian Central Catheter · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 50801902123911), received between Mar 22, 2024 and Jan 21, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (1 report).

9 other recalls cover devices cleared under the same FDA submission (K132133); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture1
  • Material Separation1
  • Material Split, Cut or Torn1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2026Material Rupture

    IT WAS REPORTED THAT "WE HAD A PT TODAY THAT WE PLACED A JACC IN AND LINE WORKED FINE UP UNTIL THE POINT IT NEEDED TO BE POWER INJECTED. JACC WAS PLACED AT 0940. PT SENT TO CT APPROXIMATELY 1430. CT REPORTING LINE LEAKING WHEN ATTEMPTING TO POWER INJECT SALINE. PRIOR TO THIS, THE LINE WORKED FINE INCLUDING HAND INJECTING SALINE. THE FLOOR CONTACTED US AT 1514 THAT IJ IS OOZING. UPON ARRIVAL TO ROOM AT 1530 AND LINE ASSESSMENT PERFORMED, DISTAL POWER INJECTABLE LUMEN OF LINE ASPIRATES BLOOD BUT LEAKS SALINE WHEN FLUSHING. LINE DISCOVERED TO BE RUPTURED RIGHT AFTER WINGS". THE LINE WAS REMOVED AND REPLACED AS THE PATIENT WAS LEAKING BLOOD OUT OF THE HOLE. THERE WAS NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "STABLE, DOWNGRADED FROM ICU STATUS".

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24130147 ↗
  • MalfunctionSep 25, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: " SHEATH BROKE OFF HANDLE". POST INSERTION OF A JACC THROUGH THE SHEATH, WHEN BEGINNING TO TEAR AWAY THE SHEATH, THE WHITE "HANDLE" OR TAB TORE AWAY FROM THE REST OF THE SHEATH LEAVING ONLY THE OTHER TAB AND THE SHEATH INTACT. TO REMOVE THE SHEATH, I MANUALLY PUSHED THE CATHETER ALONG THE LINE OF THE SHEATH WHERE IT IS INTENDED TO TEAR AWAY, OPENING ONLY ONE SIDE OF THE SHEATH. THE SHEATH WAS REMOVED COMPLETELY AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20300457 ↗
  • MalfunctionMar 22, 2024Material Twisted/Bent; Material Split, Cut or Torn; Material Twisted/Bent; Material Split, Cut or Torn

    IT WAS REPORTED THAT: CATHETER WAS PLACED (B)(6) 2024, THEN FOUND TO BE KINKED IN THE SOFT TISSUES ON (B)(6) WITH POOR BLOOD RETURN. ONE OF THE NURSES PULLED IT BACK TO SEE IF THE KINK WOULD SMOOTH OUT AND IF THE LINE WOULD THEN WORK. AS SHE PULLED THE CATHETER BACK, SHE NOTED THE SPLIT AND REMOVED THE CATHETER. THE SPLIT IS FOUND AROUND THE 20CM MARK. PATIENT RECEIVED A NEW JACC PLACEMENT AFTER THIS CATHETER WAS REMOVED. ADDITIONAL INFORMATION: THE CATHETER WAS PLACED SATURDAY NIGHT. ON SUNDAY THE PATIENT WAS VERY AGITATED AND MOVING ALOT. NURSES HAD TO HOLD HER WHILE THEY ATTEMPTED TO GIVE IV SEDATION THROUGH THE CATHETER. THEY USED FORCE AND WERE UNSUCCESFUL, PIVS WERE INSERTED AND USED FOR SEDATION. A CHEST XRAY THEN SHOWED THAT THE LINE WAS KINKED. THE PATIENT CONDITION WAS REPORTED AS "FINE" POST-PROCEDURE. MEDICAL INTERVENTION REPORTED STATES: CATHETER NEEDED TO BE REMOVED PRIOR TO THERAPY COMPLETION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18957525 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K132133. One submission often covers many models, so check the product and lot details in each record.

  • Z-2379-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT: 1-L 4.5 FR X 15 CM/CDC-41541-JX1A; AGBA PI JA
  • Z-1815-2019 · Apr 23, 2019 · Terminated
    Arrow JACC with Chlorag+ard Technology Product Code:JR 42563 HPHNM
  • Z-1817-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance JACC Product Code: JRT 43552 HPHNM
  • Z-1806-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance PICC pre loaded with Arrow VPS" Precision Stylet Product Code: CDC 44052 VPS2
  • Z-1816-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance JACC Product Code: JRT 43541 HPHNM
  • Z-1814-2019 · Apr 23, 2019 · Terminated
    Arrow JACC with Chlorag+ard Technology Product Code: JR 42541 HPHNM
  • Z-1812-2019 · Apr 23, 2019 · Terminated
    Arrow JACC with Chlorag+ard Technology Product Code: JR 42041 HPHNM
  • Z-2015-2018 · Apr 11, 2018 · Terminated
    ARROW JACC with Chlorag+ard Technology; Product Code: JR-42063-HPHNM and JR-42563-HPHNM
  • Z-0207-2016 · Oct 16, 2015 · Terminated
    ARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow Pressure Injectable Arrow PICC powered by Arrow VPS

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →