Skip to content

ARROW IPN040055Recalls & reported problems

Teleflex Incorporated · Epidural Needle · Class 2

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 50801902126615), received between Nov 13, 2025 and Nov 13, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Too Soft/Flexible” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Too Soft/Flexible1
  • Material Twisted/Bent1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2025Material Twisted/Bent; Material Too Soft/Flexible; Material Twisted/Bent; Material Too Soft/Flexible

    IT WAS REPORTED THAT "OUR LAST BATCH OF NEEDLES ARE "BENDING" WHEN THE PHYSICIAN IS USING THEM, AND THIS DID NOT HAPPEN BEFORE".

    Patient problems: Insufficient Information

    MAUDE report 23540648 ↗

Other Anesthesia conduction kit devices

All Anesthesia conduction kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page