Skip to content

ARROW IPN037304Recalls & reported problems

Teleflex Incorporated · Pressure Injectable Arrowg+ard Blue Advance(R) One-Lumen PICC pre-loaded with Arrow(R) VPS Precision(TM) Stylet · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 60801902120740), received between Nov 2, 2021 and Feb 7, 2025 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (6 reports).

8 other recalls cover devices cleared under the same FDA submission (K153487); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Misassembled During Manufacturing /Shipping6
  • Break2
  • Component Misassembled1
  • Difficult to Advance1
  • Material Separation1
  • Patient Device Interaction Problem1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Anaphylactic Shock1
  • Cardiac Arrest1
  • Hypersensitivity/Allergic reaction1
  • Rash1
  • Vomiting1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 7, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT THE CATHETER WAS PLACED MIDARM IN THE BRACHIAL VEIN AND SECURED WITH A STAT-LOCK. THE NEXT DAY, THE PATIENT WAS SITTING IN HIS CHAIR WHEN THE NURSE CAME IN AND SAW IT BLEEDING. THE LINE COMPLETELY SEPARATED FROM THE PINK LUMEN. NURSE REPORTS THAT IT WAS POSSIBLE THAT IT WAS TUGGED ON/PULLED WHEN THE PATIENT WAS UP IN THE CHAIR BUT DOES NOT BELIEVE IT SHOULD HAVE BEEN ENOUGH FORCE TO BREAK OFF. THE PATIENT IS FINE AND HAD NO HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21332157 ↗
  • InjuryFeb 27, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT: THE PATIENT HAD AN ANAPHYLACTIC REACTION. THEY WERE RESUSCITATED IN THE ICU. ADDITIONAL INFORMATION: CLINICIAN STATES HE WAS CRITICAL, BUT IS NOW FINE. PATIENT WAS EXTUBATED (B)(6) 2023.

    Patient problems: Anaphylactic Shock; Cardiac Arrest; Hypersensitivity/Allergic reaction; Rash; Vomiting

    MAUDE report 18788501 ↗
  • MalfunctionOct 1, 2023Break; Break

    IT WAS REPORTED WHILE ATTEMPTING TO PEEL THE PEEL-AWAY SHEATH, ONE SIDE OF THE WHITE SHEATH HUB BROKE OFF. CLINICIAN USED THE LEVERAGE OF THE PICC AND ONE WHITE WING TO CONTINUE TO PEEL THE SHEATH EVEN THOUGH THE OTHER WHITE HUB BROKE OFF. PROCEDURE COMPLETED WITHOUT NEEDING TO DROP AN ADDITIONAL SHEATH. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17849435 ↗
  • MalfunctionJun 2, 2022Break; Break

    IT WAS REPORTED THE TAB BROKE OFF THE PEEL-AWAY SHEATH DURING REMOVAL. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14577522 ↗
  • MalfunctionDec 28, 2021Component Misassembled; Component Misassembled

    WHEN GOING TO PLACE A SINGLE LUMEN PICC, THE USER FOUND THE KIT HAD A TRIPLE LUMEN CATHETER INSTEAD OF THE SINGLE LUMEN CATHETER. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13093384 ↗
  • MalfunctionDec 21, 2021Difficult to Advance; Physical Resistance/Sticking; Difficult to Advance; Physical Resistance/Sticking

    PICC WOULD NOT PASS THROUGH THE PEEL-AWAY SHEATH. THEY OPENED A SEPARATE 5.5 FR SHEATH, THROUGH WHICH THEY INSERTED THE CATHETER. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13055394 ↗
  • MalfunctionDec 2, 2021Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    PRIOR TO INSERTION RN OPENED SINGLE LUMEN KIT AND FOUND THERE WAS A TRIPLE LUMEN PICC IN KIT. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12920508 ↗
  • MalfunctionNov 18, 2021Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    PRIOR TO INSERTION RN OPENED SINGLE LUMEN KIT AND FOUND THERE WAS A TRIPLE LUMEN PICC IN KIT. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12838470 ↗
  • MalfunctionNov 2, 2021Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    CUSTOMER OPENED THE KITS AND THEY HAD TRIPLE LUMEN PICCS IN THEM, NOT SINGLE LUMENS AS THEY WERE LABELED. THE ISSUE WAS DETECTED PRIOR TO PATIENT USE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12740220 ↗
  • MalfunctionNov 2, 2021Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    CUSTOMER OPENED THE KITS AND THEY HAD TRIPLE LUMEN PICCS IN THEM, NOT SINGLE LUMENS AS THEY WERE LABELED. THE ISSUE WAS DETECTED PRIOR TO PATIENT USE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12740185 ↗
  • MalfunctionNov 2, 2021Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    CUSTOMER OPENED THE KITS AND THEY HAD TRIPLE LUMEN PICCS IN THEM, NOT SINGLE LUMENS AS THEY WERE LABELED. THE ISSUE WAS DETECTED PRIOR TO PATIENT USE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12740070 ↗
  • MalfunctionNov 2, 2021Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    CUSTOMER OPENED THE KITS AND THEY HAD TRIPLE LUMEN PICCS IN THEM, NOT SINGLE LUMENS AS THEY WERE LABELED. THE ISSUE WAS DETECTED PRIOR TO PATIENT USE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12740024 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K153487. One submission often covers many models, so check the product and lot details in each record.

  • Z-0001-2024 · Aug 10, 2023 · Open, Classified
    Arrow ErgoPack, Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC pre-loaded with Arrow VPS Precision Stylet, REF CDC-45552-VPS2; Catheter: Percutaneous Implanted Long-Ter
  • Z-1808-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance" PICC pre loaded with Arrow VPS" Precision Stylet Product Code: CDC 45041 VPS2
  • Z-1810-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance PICC pre loaded with Arrow VPS" Precision Stylet Product Code: CDC 45063 VPS2
  • Z-1807-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance PICC pre loaded with Arrow VPS" Precision Stylet Product Code: CDC 44063 VPS2
  • Z-1366-2019 · Dec 13, 2018 · Terminated
    Arrow PICC powered by Arrow VPS Precision Stylet Product Code:CDC-44052-VPS2 Product Usage: The Arrow Pressure Injectable PICC is indicated for short-term or longterm peripheral a
  • Z-1365-2019 · Dec 13, 2018 · Terminated
    Arrow PICC powered by Arrow VPS Precision Stylet Product Code:CDC-45052-VPS2 Product Usage: The Arrow Pressure Injectable PICC is indicated for short-term or longterm peripheral a
  • Z-3200-2018 · Jul 17, 2018 · Terminated
    5.5 Fr x 40 cm double-lumen PICC Kit, Product Code CDC-44052-VPS2
  • Z-3199-2018 · Jul 17, 2018 · Terminated
    4.5 Fr x 40 cm single-lumen PICC Kit, Product Code CDC-44041-VPS2, 5.5 Fr double-lumen PICC Kit, Product Code CDC-44052-VPS2

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page